Lighting Uv Lamp For 24 Hours - Panasonic KM-CC17R2 Series Operating Instructions Manual

Co2

Advertisement

UV LAMP PARAMETER

Lighting UV lamp for 24 hours

If the chamber has been contaminated by dirt or by spilling the medium, use the following procedure to
decontaminate the chamber by lighting the UV lamp for 24 hours.
1. Remove all attachments from the chamber, including the trays, the fan cover, the duct, the fan, the
humidifying pan, and the humidifying pan cover. Clean all the attachments in an autoclave or with alcohol
for disinfection.
2. Clean and wipe off inside the chamber with alcohol for disinfection.
o
3. Set the CO
density to 0 % and set the high limit alarm temperature to at least 10
C higher than the
2
chamber set temperature. Refer to page 31~32.
Note: The UV 24-hour mode may increase the chamber temperature.
4. Press Menu key to lead the Menu screen.
5. Press Tools #1 key to lead the Tools #1 screen.
6. Press UV Setting key to lead the UV Setting
screen.
58