Nivotester FTR525
Endress+Hauser
10�4�2 Previous diagnostics
Diagnostics
Previous diagnostics
→
1000
1200
This function is used to display the previous error (code with short message).
10�4�3 Erase previous diagnostics
Diagnostics
Erase previous diagnostics
→
1000
This function is used to erase the previous logged error (code with short message).
Options: "Keep", "Erase"
10.5 Overview of diagnostic information
Code
Help text
0000
No error
F041
Connection FDR
F042
Connection FQR
F043
Communication FDR
F044
Communication FQR
F261
Flash FTR
F262
EEPROM FTR
F263
EEPROM FQR
F264
EEPROM FDR
F401
FTR: T < Min / T > Max
F402
FDR: T < Min / T > Max
F403
FQR: T < Min / T > Max
C482
Simulation
10�6
Reset device
Reset functions of the FTR525
→ 40
1300
Remedial action
−
• Check electrical connection to the FDR57
• Check function of the FDR57, if necessary
• In case of defect: contact service
• Check electrical connection to the FQR57
• Check function of the FQR57, if necessary
• In case of defect: contact service
• Check function of the FDR57
• Reset device, if necessary
• In case of defect: contact service
• Check function of the FQR57
• Reset device, if necessary
• In case of defect: contact service
• Perform reset of the FTR525
• If the defect occurs repetitively: contact service
• Perform reset of the FQR57
• If the defect occurs repetitively: contact service
• Perform reset of the FDR57
• If the defect occurs repetitively: contact service
• Minimum temperature on the device undercut or
maximum temperature exceeded (Technical Data
→ 57)
• Ensure operating temperature range (Installation
conditions → 13)
• Only for information, no error
• Deactivate simulation (→ 41)
Diagnostics and troubleshooting
49