Caution!
C-6 Receiving and Installation
.
B
C
1. Back of hood
1. Back of hood
2. Holes for screws
2. Holes for screws
3. Tab for hood cover
3. Tab for hood cover
15. Set the speed range for the treadmill using the following
steps.
a. Enter the service mode by simultaneously pressing
Stop Belt
,
Slower
b. Choose the appropriate configuration number from
the following:
Speed Range
0.6 to 7.2 MPH
1.0 to 12.0 MPH
c. Hold down
Stop Belt
correct configuration number appears in the center
display.
The configuration must be set to CP1 or CP3. Do not
attempt to use any other configuration.
d. Press
to store the configuration.
Select
e. Exit the service mode by simultaneously pressing
Stop Belt, Slower,
16. Test all operational functions, including speed, grade,
timer, and parameters using the following section.
A. Hood cover
A. Hood cover
B. Phillips #8 screws
B. Phillips #8 screws
C. Holes for mounting
C. Holes for mounting
A
hood cover
hood cover
B
B
C
C
3
2
Installing the hood
, and
Faster
.
Configuration Setting
CP1
CP3
and press the
and
.
Faster
MedTrack Rehabilitation Treadmills User Guide
2
1
or
key until the
+
-

Hide quick links: