Problem
Walking belt too far left or
right.
Walking belt slips, but
front roller turns.
Walking belt hesitates;
adjusting walking belt
tension is ineffective
Treadmill will not change
grade.
Treadmill will not reach
maximum speed.
Treadmill will not reach
maximum time.
Circuit breaker trips
during normal operation.
Error messages on
controller.
MedTrack Rehabilitation Treadmills User Guide
Possible Cause
Improper belt tracking.
Improper belt tension.
Internal drive belt slipping.
Excess weight on treadmill.
Internal problem.
Speed limit control is
activated.
Time limit control is activated.
Power fault.
Internal problem.
Remedy
Adjust tracking (see "Belt
Tracking" on page 3-4). If problem
persists, contact an authorized
service rep.
Adjust belt tension (see "Belt
Tension" on page 3-3). If problem
persists, contact an authorized
service rep.
Service required. Contact an
authorized service rep.
See Appendix B for maximum
load.
Service required. Contact an
authorized service rep.
See "Speed Limit Control" on
page 2-5.
See "Time Limit Control" on
page 2-4.
Service required. Contact an
authorized service rep.
Service required. Contact an
authorized service rep.
Maintenance/Troubleshooting 3-9