Welch Allyn BP100SBP Directions For Use Manual page 30

Remote monitoring blood pressure device rpm
Table of Contents

Advertisement

26 Appendix
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency
(50Hz/60Hz) magnetic
field IEC 61000-4-8
Note: U
is the AC mains voltage prior to application of the test level.
T
The Welch Allyn Home Blood Pressure Monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of the Welch AllynHome Blood Pressure Monitor should assure that it is used in such an
environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Welch Allyn Remote Monitoring Blood Pressure Device RPM-BP100SBP
Electromagnetic immunity
<5% U
<5% U
T
T
(>95% dip in U
) for 0.5
(>95% dip in U
T
cycle
cycle
40% U
40% U
T
T
(60% dip in U
) for 5
(60% dip in U
T
cycles
cycles
70% U
70% U
T
T
(30% dip in U
) for 25
(30% dip in U
T
cycles
cycles
<5% U
<5% U
T
T
(>95% dip in U
) for 5
(>95% dip in U
T
seconds
seconds
3A/m & 30A/m
3A/m & 30A/m
Electromagnetic immunity
IEC 60601 test level
Compliance level
3 Vrms & 6Vms
3 Vrms & 6Vms
150 kHz to 80 MHz
3 V/m 80 MHz to 2.5
3 V/m
GHz
10 V/m
80 MHz to 2,6 GHz
10 V/m
Mains power quality should be that of a typical
commercial or hospital environment. If the user
) for 0.5
of the Welch Allyn Home Blood Pressure
T
Monitor requires continued operation during
power mains interruptions, it is recommended
that the Welch Allyn Home Blood Pressure
) for 5
Monitor be powered from an uninterruptible
T
power supply or a battery.
) for 25
T
) for 5
T
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any part
of the Welch Allyn Home Blood Pressure
Monitor, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Recommended separation distance
d= 1,2
d= 1,2
80 to 800 MHz
d= 1,2
800 MHz to 2,5 GHz
d= 2,3
800 MHz to 2,6 GHz
whereP is the maximum output power rating of
the transmitter in watts (W) andd is the
recommended separation distance in meters
(m).