Welch Allyn BP100SBP Directions For Use Manual page 24

Remote monitoring blood pressure device rpm
Table of Contents

Advertisement

20 Maintenance and troubleshooting
Problem
Low batteries
Error messages
Hardware error
Physiological alarm indications
Symptom
Welch Allyn Remote Monitoring Blood Pressure Device RPM-BP100SBP
Symptom
The display indicates the
"BAT-LO" message, pauses for
3 seconds. The battery icon
shows empty (does not flash).
E 01 shows
E 03 shows
E 10 or E 11 shows
EE and E 19 shows
Root cause
Out of range. Either SYS >260mmHg or DIA
>220mmHg. The symbol may appear in
either the SYS or DIA area of the screen.
Out of range. Either SYS <50mmHg or DIA
<25mmHg. The symbol may appear in either
the SYS or DIA area of the screen.
Root cause
Batteries are inserted
incorrectly.
AC adapter is inserted
incorrectly.
Batteries are low.
The cuff is not secure.
There is too much pressure in
the cuff.
The monitor detected motion
while measuring.
Malfunctioning hardware
Solution
Press and hold the
reestablish Bluetooth connectivity.
Measure again. If the problem persists,
contact your physician.
Alarm priority = Low
Press and hold the
reestablish Bluetooth connectivity.
Measure again. If the problem persists,
contact your physician.
Alarm priority = Low
Solution
Insert the batteries correctly
Insert the AC adapter tightly
Replace with new batteries
Readjust the cuff, relax for a
moment, and then measure
again
Refasten the cuff and then
measure again
Readjust the cuff, relax for a
moment, and then measure
again
Measure again. If the problem
persists, contact the retailer or
Welch Allyn customer service
at
www.welchallynhome.com
for further assistance.
Power
button to
Power
button to