Film sizes

Automatic image quality and processing

Configuring and monitoring the system (using the Web
Portal)
1-4
The laser imager supports four sizes:
• 35 x 43 cm (14 x 17 in.)
• 28 x 35 cm (11 x 14 in.)
• 25 x 30 cm (10 x 12 in.)
• 20 x 25 cm (8 x 10 in.)
For the specific films that are supported, see the Publications Cover
Page.
An internal densitometer enables the laser imager to automatically adjust
image processing parameters (Automatic Image Quality Control, or
AIQC) to produce an optimal image. The laser imager adjusts these
parameters each time it prints a calibration film.
A calibration film is printed when:
• The film tray is inserted in the laser imager with film of a new lot
number.
• You request a calibration film at the display screen or the Web
Portal. For more information, see
the installed film" on page
• A film cartridge is inserted into the laser imager for which a current
calibration is not stored.
The Web Portal is your interface to additional features. In addition to
the installation and setup of your system, you can view and manage the
laser imager's connections over the network, configure features, view
error messages, and access general status information at the Web Portal.
You can also check film count, film size, and film type.
For more information, see
9G3886_en
"Calibrating the laser imager for
2-9.
"Accessing the Web Portal" on page
2-10.