About the Product
About the Pr
oduct
Measure intensity, absorbance, transmittance across the visible and
UV spectrum. After specifying a target wavelength, students can study
concentrations (Beer's Law), rates of reactions, or investigate
emission spectra using the optional fiber optic cable.
What's Included
What'
s Included
• UV-VIS Spectrometer
• 2× Cuvette Rack
• 10× Semi-Micro Cuvette with Cap
• AC Adapter
• USB Cable
• Spectrometry Software (online download)
Recommended Items
Recommended Items
• Fiber Optic Cable Kit (SE-7182)
Fiber Optic Cable Kit (SE-7182)
Required for analyzing emission spectra.
©2021 PASCO scientific
UV
UV-VIS Spectr
-VIS Spectrometer
SE-3607
Safety Instructions
Saf
• This instrument contains an internal high-voltage power source
• Only use the power supply supplied with the product.
• Do not remove or modify any of the components of this device.
• Do not block the exhaust of the ventilator fan. After turning off
• This is a precision measuring instrument. Avoid dropping or
• Do not use this instrument if it is damaged in any way. Contact
• PASCO products are designed for educational use only. This
ometer
ety Instructions
CAUTION:
This product contains a high intensity ultraviolet light
source. Always wear proper eye protection (UV
blocking safety glasses) and skin protection when
using this device.
and a replaceable UV lamp assembly. Do not attempt to open or
remove the case at any time except for servicing by qualified
personnel only. Always make sure the unit is unplugged before
servicing.
Doing so will create an unsafe operating condition and will void
the product warranty.
the device, do not disconnect power until the fan stops running.
Failure to follow these directions will damage the equipment.
Notice:
rough handling which can damage the unit or affect the
measuring accuracy.
PASCO Technical Support for assistance.
product should not be used for clinical or professional diagnosis
or industrial control.
Product Guide
012-16942A
www.pasco.com
Need help?
Do you have a question about the SE-3607 and is the answer not in the manual?
Questions and answers
Hola, quisiera saber cual es la incertidumbre de error qué puede tener el equipo
The error uncertainty for the PASCO SE-3607 equipment is ±5% photometric accuracy and a wavelength accuracy of 1.0 nm.
This answer is automatically generated