GE MEDICAL SYSTEMS
REV 10
5−4
Tube Unit Rotation
ILLUSTRATION 5−3
TUBE ROTATION
Because the tube assembly is counterbalanced, no active lock is required. And little ef-
fort is needed to override friction and rotate the tube.
5. Use the collimator handles to position the tube to the desired angle.
D
You may rotate the tube in two directions:
a.
clockwise and counterclockwise +180 degrees (detents at every 90 de-
grees), or
b. backwards −10 degrees and forwards +100 degrees (detent at 0 de-
gree). See Illustration 5−3.
COLLIMATOR
ROTATION
KNOB
110_
(FORWARD)
5−3
(MODEL 2169360, 2236420 & 2275938 SERIES)
TUBE UNIT
...extend the arm and
0_
rotate the tube to the
DETENT
desired angle.
−10_
(BACKWARD)
AMX−4+ OPERATION
om 2166913−100
RETRACTABLE
ARM