9−1

IEC Classification

9−2

Earth Leakage Current

9−3

Applicable IEC Symbols

In accordance with International Safety Standard IEC 601, the AMX−4+ Mobile
X−ray Unit is classified as Type B Equipment, Class I Equipment, Internally Powered
Equipment, Ordinary Equipment and continuous operation.
Type B Equipment provides an adequate degree of protection against electric shock,
in particular regarding:
D
allowable leakage current,
D
reliability of the protective earth connection.
In accordance with International Safety Standard IEC 601, the manufacturer is not
responsible for any consequences caused by unauthorized modification of this equip-
ment.
This equipment is not suitable for use in the presence of flammable gas.
There are no user replaceable parts in this unit. This would include the cord reel, as
well as any other mains or secondary part.
The AMX−4+ Mobile X−ray Unit complies with (Class I − Type B) standards.
Protective earth (ground). Identifies any terminal which is intended for connection of
an external protective conductor to protect against electrical shock in case of a fault.
This symbol on the equipment means that the accompanying operating instructions
should be consulted to assure safe operation.
Dangerous voltage. Indicates an avoidable dangerous high voltage hazard.
Functional Earth (ground) Terminal. Terminal directly connected to a point of a
measuring supply or control circuit or to a screening part which is intended to be
earthed for functional purposes.