6. Post Use

6.1. Disconnect device after use

A. Turn off gas supply to the ventilator.
B. Disconnect gas supply hose.
C. Disconnect patient circuit from the output connector.
D. Unplug the power cable from mains if no charging is required.
E. Clean and disinfect according to section 8.1 in this manual.

6.2. Storage

Store ventilator within the following environmental range:
- 40
C to +60
o
Note: The ventilator operates within 5 minutes after being brought
back from minimum storage temperature to room temperature;
The ventilator operates within 2 minutes after being brought back
from maximum storage temperature to room temperature.

7. Alarms and Indicators

7.1. Ventilation Alarms

Visual and Audible alarms continue until the cause of the alarm is
resolved.
During alarm activation the user may press the Alarm Silence key
(G) (Figure 01) which will silence the audible alarm for 2 minutes
but the visual alarm will continue to flash until the cause of the
problem is resolved.
During "alarm silence", should a new alarm develop, the Alarm
Silence function will continue and only the new visual alarm will be
shown flashing on the screen.
Alarms will be visible in section 4 or section 7 of the display (Figure
04).