9. Technical Data

9.1. Specifications

CLASSIFICATION
PER IEC60601-1
POWER SOURCE (PNEUMATIC) Compressed Oxygen, 45 to 87 PSI (3-6 Bar)
POWER SOURCE (ELECTRIC)
TIDAL VOLUME /RATE (ML/BPM)
MANUALLY TRIGGERED VENTILATION Yes, set flow rate or pressure will be delivered
MAXIMUM INSPIRATORY HOLD TIME 6 sec.
TRIGGER SENSITIVITY (L/MIN) 3 L/min (± 10%) or 2 cmH
APNEA BACK UP TIME (SEC.) 10-60 (± 0.5s)
BATTERY OPERATING TIME AT ROOM
ALTITUDE COMPENSATION up to 4000 m (13000 feet)
BUILT-IN BATTERY CHARGER Yes
DAY/NIGHT DISPLAY MODE Yes
NOISE LEVEL IN NORMAL USE Less than 65 dBA
ALARMS
(VISUAL AND AUDIBLE)
AUDIBLE SILENCE
DIMENSIONS (MM)
WEIGHT (KG)
46
DEVICE CLASS PER MDD II b
Protection against
electric shock
Protection against water IP X4
VENTILATION MODES SIMV, CPAP, Mask CPR and Intubated CPR
I:E RATIO 1:2 (± 20%)
PEEP (cmH
CPAP (cmH
FiO
PMAX (cmH
PMIN (cmH
TEMPERATURE (HRS.)
BATTERY HOT SWAP No
AC/DC POWER SUPPLY 100-240 VAC/ 19 VDC, 4.74 A
PATIENT CIRCUIT O-Two Electronic Ventilator Circuit
MOUNTING BRACKET Smart Mount multi-configuration frame
DISPLAY 4.3" Color TFT
LIVE MONITORING
REAL TIME WAVEFORM Pressure or Flow
PARAMETER SETTINGS Control Selection Knob
LOCK KEY FUNCTION Yes
PAUSE FUNCTION Yes
Class II, Type BF
AC/DC power supply,
Rechargeable Lithium Battery
100/20, 150/20, 200/20, 250/20, 300/10, 350/10
400/10, 450/10, 500/10, 600/10, 650/10
± (4ml + 15%) BTPS */ (± 10% or ± 1BPM)
during I time then Inspiratory hold
O) 0, 4, 5, 6, 7, 8, 9, 10, 12, 15, 18, 20
(± 10% or ± 2 cm H
2
O) 4, 5, 6, 7, 8, 9, 10, 11, 12, 15, 18, 20
(± 10% or ± 2 cm H
2
(%) 60 or 100 (± 15%)
2
10, 15, 20, 25, 30, 35, 40, 45, 50, 55, 60, 80
O)
(± 10% or ± 2 cm H
2
O) 0, 2, 3, 4, 5, 6, 7, 8, 10, 12, 15, 20
(± 10% or ± 2 cm H
2
CPAP mode only
> 16 hrs for default settings (Data obtained using
fully charged new battery)
Mve, Vte, Paw(AV), PAW(Peak), Rate (bpm),
Battery level
Gas Supply Pressure, Airway Pressure limits,
Battery status, APNEA, Breathing Circuit Integrity,
and Leakage
Yes, 120 second max
250 x 200 x 155
2.4 (w/ Battery), 1.77 (w/o Battery )
O)
2
O)
2
O)
2
O)
2
O below baseline in
2