The external power supply and battery are component parts of
the medical electrical equipment system.
The ventilator is considered as a high flow device as its
maximum flow output at pressure of 40.6 PSI is around 100
L/min. It should only be connected to a pipeline system that
allows for the indicated high flow to avoid interfering with the
operation of adjacent equipment.
Always ensure that all components are assembled correctly
and ready for use.
When selecting very small tidal volumes during ventilation of
infants, take into consideration the dead space in the patient
circuit.

2. Intended Use

The e500 is a time-cycled and volume-constant emergency and
transport ventilator designed for use in the pre-hospital, intra-
hospital, inter-hospital and transport settings. It is intended for use
with Adult, child, infant patients with a tidal volume from 100 ml
upwards who are in respiratory and/or cardiac arrest or respiratory
distress and who require the ventilatory support.