This device is intended for use in the electromagnetic environment specified below. The user of
this device should make sure it is used in such an environment.
i
MMUnity
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical fast
Transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
A-2
M
'
d
anUfaCtURER
s
t
iEC 60601 t
Est
l
EvEl
±6 kV contact
±8 kV air
±2 kV for power sup-
ply lines
±1 kV for input-out-
put lines
±1 kV differential
mode
±2 kV common mode
<5% U
T
(>95% dip in U
0.5 cycle
40% U
T
(60% dip in U
5 cycles
70% U
(30% dip in
T
U
) for 25 cycles
T
<5% U
(>95% dip in
T
U
) for 5 sec
T
- E
EClaRation
lECtRoMagnEtiC
C
Est
oMplianCE
±6 kV contact
±8 kV air
±2 kV for supply
mains
±1 kV for input/out-
put lines
±1 kV differential
mode
±2 kV for common
mode
<5% U
T
) for
(>95% dip in U
T
0.5 cycle
40% U
T
) for
(60% dip in U
T
cycles
70% U
(30% dip in
T
U
) for 25 cycles
T
<5% U
(>95% dip in
T
U
) for 5 sec
T
REM
i
MMUnity
l
E
EvEl
lECtRoMagnEtiC
E
nviRonMEnt
g
UidanCE
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humid-
ity should be at least
30%.
Mains power quality
should be that of a
typical home or hos-
pital environment.
Mains power quality
should be that of a
typical home or hos-
pital environment.
Mains power qual-
) for
ity should be that
T
of a typical home
or hospital environ-
) for 5
ment. If the user of
T
the device requires
continued operation
during power mains
interruptions, it is rec-
ommended that the
device be powered
from an uninterrupt-
ible power supply or
a battery.
M s
U
M
staR
ERiEs
sER
-
anUal