Maintenance Schedule; Obtain Parts And Supplies - Data I/O PSV7000 Manual

Ionizer kit
Hide thumbs Also See for PSV7000:
Table of Contents

Advertisement

3. Carefully inspect the electrode needle/probe for any damage or contaminants (such as dust or dirt).
Inspect electrode needle/probe for contaminants.

Maintenance Schedule

Use the following maintenance schedule to ensure proper ionizer operation (your specific
environmental conditions may vary).
At the start of each job (with the loading of new reels/devices), ensure that the Maintenance LED
status indicator is GREEN.
On a weekly basis, check the nozzle(s) to ensure proper ion discharge (ionizer is not damaged and
its emitter points are clear/unblocked); also check and monitor electrostatic field magnitude with
meter to ensure proper operation.
On a monthly basis, check the electrode needle to ensure there is no corrosion or condensation (the
needle should be clean and clear of contaminants).

Obtain Parts and Supplies

To order consumable ionizer parts from Data I/O, please refer to the following part numbers.
Part Name
Electrode Cleaner
Electrode Cartridge (De-
ionizing, Tungsten)
PSV7000 Ionizer
Self-Service Kit
PSV7000 Ionizer Kit
Description
Tip cleaner for electrode needle/probe
1 cartridge for the Ionizer
(with no electrode needle/probe)
6 cartridges and 1 electrode needle cleaner
18 |
P a g e
Data I/O
Part Number
565-0223-001
420-0011-001
952-0533-001