Section 5:

Routine Care and Maintenance

WARNING!
Disconnect the autoclave from the mains power
supply before cleaning.
Daily Maintenance
Gasket
THE GASKET MUST BE CLEANED ON A DAILY BASIS,
BEFORE USING THE AUTOCLAVE.
Wipe exposed surface of the gasket and
the surface of the vessel with warm soapy
water using a lint free damp cloth.
Wipe both the gasket and the vessel again
with water using a lint free damp cloth to
remove any residual soap.
EVERY 250 CYCLES.
Clean the autoclave using Autoclave cleaning kit
(Part No. 289138A).
WARNING
Failure to perform these procedures may result in the unit
(Part No. 289138).
showing UOD1 on the display (UOD1 indicates that the unit may
have been contaminated and could fail to operate correctly).
Change the air filter situated at the
front of the autoclave.
Gently unscrew the filter, and fit the
replacement.
Exterior surfaces (as required).
Exterior surfaces should be cleaned with warm soapy
water using a damp cloth.
For persistent marks, use a gentle cream cleaner.
Gasket Replacement
Fresh water tank / Waste water container.
On a monthly basis, fully drain the water tank and leave
overnight (use Autoclave cleaning kit Part No. 289138).
Drain, then refill with fresh water. Repeat this operation
twice more to remove any residue.
Always use de-ionised, distilled or sterile water as
recommended. NEVER USE TAP WATER.
Routine Maintenance.
Annually replace the water filters.
Should the gasket develop a persistent leak it should be
removed, cleaned thoroughly in warm soapy water and
shaken dry. Wiping with a lint-free cloth is acceptable
(Other materials may lead to contamination of the gasket
with fibres).
The door gasket plate must also be cleaned.
If the leak persists you should obtain and fit a new gasket.
To remove the gasket, undo the two thumb-nuts (309070)
in the centre of the gasket plate.
Remove the plate/gasket assembly and then remove the
gasket from the plate.
CAUTION: Ensure the gasket is located the right way
around on the plate, the lip should face outwards.
During re-assembly, place new black sealing
washers (supplied with the gasket) in the machined
recesses in the door casting.
Ensure that the entry port plug aligns with the hole in the
cast lid.
Prior to tightening the two thumb-nuts ensure that the
gasket is correctly seated on the gasket plate and in the
door recess.
DO NOT over-tighten the thumb-nuts as this may damage
the thread.
9
Thumb-Nuts
Sealing Lip - ensure
the gasket is
orientated correctly
Entry port plug
304278B