Section 11:

Technical Specification

Chamber component materials.
Vessel:
Stainless Steel – 304 –S15
Boiler:
Aluminium – LM25
Lid:
Aluminium – ASME SB26 356.0 T6
Door Plate:
Stainless Steel – 304 –S15
Chamber Capacities
Overall product width
Overall product height
Overall product length
Unpacked weight
Chamber Diameter
Chamber lengths
Max. Instrument length
Max. load non-vacuum
Max. load vacuum
* Pack size - 3 x standard vet
pack @ 500g each
Sterilizing temp/time.
Operating pressure
(Minimum)
Voltage/Wattage
Frequency
Safety Shutdown:
Refer to "Temperature cut out". Only a qualified person
should reset this.
Environmental conditions:
Indoor use at an altitude of up to 2,000m.
Ambient temperature range +10ºC to +40ºC
Maximum relative humidity 80% for temperatures up to
30ºC, decreasing linearly to 50% at 40ºC.
Mains supply voltage range 207 to 254volts.
Drying performance maybe affected by local environmental
conditions.
Maximum Water usage:
Vacuum cycles = 600mls per cycle (typical).
Non-Vacuum cycles = 450mls per cycle (typical).
Disconnect the autoclave from the mains power supply
before changing fuses.
Only qualified personnel should replace fuses.
304278B
22litre
480mm.
410mm.
610mm.
43Kg. (Max)
250mm.
430mm.
420mm.
6 kg
6 kg (un-pouched) 2 kg (pouched)
1.5 kg (porous)*
134ºC/3½mins.
121ºC/15mins.
2.05bar.(gauge)
230v. /2200W
50 – 60 Hz.
WARNING
Mains plug top fuse
(user replaceable)
F13A to BS1362 UK only.
Maximum single fault temperature:
143ºC determined by the pressure release valve.
Pressure Release Valve:
Operates at 2.9bar. Accumulation is <10%.
Storage:
When leaving the unit standing idle for any length of time,
switch mains off & drain the water tank especially if there
is a possibility of the room temperature dropping below
freezing point.
Packaging:
Packing materials used have been selected for ease of
recycling. Please ensure you use the correct disposal
system for disposal of packing materials.
Maximum power used:
1.18kwh
LEQ Sound level
= 60db.
NB. The overall cycle time will increase as the mains
supply voltage decreases.
Rating:
All products are rated for intermittent use,
continuously.
Heaters:
Cast into the boiler.
Internal heater.
Temperature cut out:
Bi-metallic type rated at 180ºC. with automatic reset.
Over Voltage category:
Pollution Degree:
Insulation:
The machine is fitted with an automatic thermal reset
mechanism. In the event of the boiler overheating
(for example due to accidental overload), the unit will
power down. After approx. 10 minutes, power will
automatically be restored and the display will show
Error 01. The load will need to be re-processed, ie
start the cycle again.
Group II
Group II
Class I.