Section 7:
Recovery Sequence
Recovery Sequence (allows instruments to be removed from the unit).
Recovery Sequence "i"
Press button (A)
Stabilize (no flashing or bleeping)
Press button (A) "Recover"
The Recovery Sequence (depending on where the fault occurs) will flush the water from the boiler and eventually
complete the cycle with a continuous bleeping to indicate recovery is completed.
Important.
Before restarting a cycle, check that the mains plug is fully inserted into the mains outlet socket and the outlet is of the
earthed/grounded type.
Should all power be lost the door cannot be opened until power is restored and pressure returned to ambient.
Should an internal power failure occur, the door cannot be opened until: -
1.
The unit has cooled down to atmospheric pressure. Access can then be made using the supplied tool.
2.
A service has been carried out.
In the event of failure of an indication device a service will be required to correct the condition.
Should a safety feature operate, unplug the unit and call for a service – do not attempt to correct the fault.
Primary safety features:
Two primary features have been fitted – a pressure release valve and a boiler over temperature safety cut out.
IMPORTANT NOTE ON EMERGENCY DOOR OPENING PROCEDURES.
If the door interlock lever does not fully release during door opening and the door led has gone out, light pressure on the
door will re-illuminate the door open led. It will then be possible to open the door in the normal manner using the door
open button.
Door opening – Power loss.
If the machine has suffered a loss of power and it is essential to open the machine to remove instruments, the following
procedure should be followed. This should only be performed by the nominated responsible person: -
1.
Disconnect the unit from the mains supply.
2.
Allow the machine to cool down to room temperature.
3.
Remove the protective white protective cap on the top of the machine.
4.
Using the hook on the special tool, release any pressure/vacuum inside the vessel by
hooking the ring on the pressure relief valve and pulling this towards you. As shown
opposite.
5.
Insert the straight end of the special tool into the interlock lever release guide hole and
push inwards. If the tool starts to buckle at the preformed strain relief repeat from 3
above.
See Photograph opposite.
DO NOT ATTEMPT TO OPEN THE VESSEL IF PRESSURISED.
ONLY USE THE TOOL PROVIDED AND FOLLOW THE INSTRUCTIONS ABOVE
TO GAIN ACCESS TO THE MACHINE.
(allows instruments to be removed from the unit).
Recovery Sequence "ii"
Press button (A)
Stabilize (no flashing or bleeping)
NB: Cannot proceed. Switch off at mains supply.
Service required.
Recovery Sequence "iii"
Press button (A)
Stabilize (no flashing or bleeping)
Press button (A) to enter "Ready Mode"
Door
release
tool hole
11
304278B