Trouble Shooting - Prestige medical Anima User Manual

Vacuum & non-vacuum autoclave
Table of Contents

Advertisement

10
Section 6:

Trouble Shooting

User Message
Cause
UOD1 and UOD2
Contamination.
Low water LED (E2)
Insufficient water to run a cycle.
"door" illuminates
Cycle start button pressed whilst door
on display (L2)
is open or door micro-switch requires
adjustment.
01
Power failure during a cycle.
b02
Sterilizing Temperature out of range.
d02
Probe difference out of range.
t02
Clock error.
p02
Sterilizing Pressure out of range.
03
Air bleed has not been successful.
04
Vacuum failure.
07
Sensor fault – Boiler thermistor.
08
Sensor fault – Pressure transducer.
10
Water in the boiler.
UOD3
Vacuum failure in drying stage.
13
Water fill time out.
14
Sensor fault – Chamber probe.
15
System Leak.
Should a fault occur a visual and audible indication would be given. The nature of the fault can be determined by reference
to the fault guide below. The recovery sequence allows access to any instruments within the autoclave and is the first step
when rectifying a fault mode.
As non-warranty related calls can be expensive it is advisable to ensure that all consumable items have been replaced or
cleaned as appropriate, and that the water quality is as described in Section 13 before contacting Prestige Medical Ltd.
304278B
Remedy.
Conduct routine maintenance procedures.
Replace gasket.
If fault persists Engineer call-out required.
Press button (A) twice. Top up with water to mark (S3).
If door closed but door LED (C) not illuminated - push on door to see if
LED can be illuminated.
If door open, close door and try again.
If fault occurs during a cycle – follow Recovery sequence "i"
Recovery sequence "i"
Check power supply – repeat the cycle.
Recovery sequence "i".
Clean gasket and chamber face then repeat the cycle.
If fault persists Engineer call-out required.
Recovery sequence "i".
Conduct routine maintenance procedures.
Clean gasket and chamber face then repeat the cycle.
If fault persists Engineer call-out required.
Recovery sequence "i".
Check & rest the clock – repeat the cycle.
If fault persists Engineer call-out required.
Recovery sequence "i".
Clean gasket and chamber face then repeat the cycle.
If fault persists Engineer call-out required.
Recovery sequence "i".
Clean gasket and chamber face then repeat the cycle.
Recovery sequence "i".
Clean gasket and chamber face then repeat the cycle.
Recovery Sequence "ii" – Repeat cycle.
If fault persists Engineer call-out required.
Recovery Sequence "ii" – Repeat cycle.
If fault persists Engineer call-out required.
Recovery Sequence "iii" – Repeat cycle.
Recovery sequence "i".
Clean or change gasket.
If fault persists Engineer call-out required.
Recovery sequence "i".
Drain water out of the fresh water tank and refill with de-ionised, distilled
or sterile water. Repeat the cycle.
Recovery Sequence "ii" – Repeat cycle.
If fault persists Engineer call-out required.
Recovery Sequence "ii" – Clean gasket and chamber face then
repeat the cycle.
If fault persists Engineer call-out required.