Operation - Prestige medical Anima User Manual

Vacuum & non-vacuum autoclave
Table of Contents

Advertisement

Cycle controls.
Press G to select the vacuum cycles these are as follows:
P1:
134°C/3min.20sec Vacuum cycle
for porous loads, wrapped, pouched,
solid / hollow instruments, with 25
minutes drying.
P2:
134ºC /3min.20sec. Vacuum cycle
for porous loads, wrapped, pouched,
solid / hollow instruments with
20 minutes drying.
P3:
134ºC /3min.20sec. Vacuum
cycle for unwrapped, solid / hollow
instruments without drying. Also
Process challenge device cycle.
P4:
121ºC /15min. Vacuum cycle for
porous loads, wrapped, pouched,
solid / hollow instruments with
25 minutes drying.
Press H to select non-vacuum cycles.
P5:
134ºC /3min.20sec. Non Vacuum
cycle for unwrapped, solid instruments
without drying.
AFTER EVERY CYCLE THE UNIT RETURNS TO THE
DEFAULT CYCLE P1.
Display Options. (Figures shown are for a vacuum unit)
L1
L2
L3
The following show the cycle status: (Vacuum unit)
L5
L7
L4
L8
L6
L1
Display for "Time" "Date"
& "Cycle counter".
The normal display mode
is time. The date or cycle
counter can be selected
by pressing the button D
adjacent to the display.
The selection is valid until
a cycle is completed or
until the unit is returned to
standby mode, at which
point the display will revert
to the current time.
L2
Display for "Temperature"
L3
Display for "Pressure".
L4
Cycle started.
L5
Heating and air bleed.
L6
Sterilizing.
L7
Depressurisation / drying.
L8
Cycle complete.
Section 3:

Operation

Please take time to read these instructions
before using the autoclave. It is essential that the
operator is correctly trained and a "Responsible
Person" has been assigned for the management of
the autoclave. By following these simple step-by-step
instructions you can ensure your instruments are correctly
sterilized every time.
1. Installation.
Ensure the unit
is placed on a
strong, flat and level
surface.
To check if the unit
is level, pour half a
cup of water into
the chamber. The
water should flow
towards the rear of
the chamber, not
out of the front.
WARNING!
The autoclave is heavy; at least two people will be needed
2. Connection.
Plug the unit into the mains outlet socket of the correct
type & rating. Refer to the duplicate rating plate located
inside the door on the bezel.
After a few seconds the LED next to button (A) illuminates.
WARNING!
The mains outlet MUST BE EARTHED (GROUNDED).
The mains plug should always be easily accessible as it is to be
relied upon as "the means of disconnection".
3. Single Use Water System.
This unit has a "single use" water
system designed to prevent the
recycling of any contaminants, which
may be present on the instruments
being sterilized.
Ensure the "waste water container"
is checked regularly to prevent
overflowing. The container should be
emptied when the wastewater reaches
the "Max" line.
To empty the container, undo the screw cap and carefully
remove the steam-condensing coil, which passes through
the cap.
to lift it.
304278B
5