Download Print this page

EASTMAN Kodak X-Omat 3000 RA PROCESSOR Preventive Maintenance

Hide thumbs Also See for Kodak X-Omat 3000 RA PROCESSOR:

Advertisement

Quick Links

Publication No. 5B6746
November 1995
PREVENTIVE MAINTENANCE
for the
Kodak X-Omat 3000 RA PROCESSOR
H150_0035DA
© E a s t m a n K o d a k C o m p a n y

Advertisement

loading
Need help?

Need help?

Do you have a question about the Kodak X-Omat 3000 RA PROCESSOR and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for EASTMAN Kodak X-Omat 3000 RA PROCESSOR

  • Page 1 Publication No. 5B6746 November 1995 PREVENTIVE MAINTENANCE for the Kodak X-Omat 3000 RA PROCESSOR H150_0035DA © E a s t m a n K o d a k C o m p a n y...
  • Page 2: Table Of Contents

    Eastman Kodak Company prior to publication. No patent license is granted by this information. Eastman Kodak Company reserves the right to change this information without notice, and makes no warranty, express or implied, with respect to this information. Kodak shall not be liable for any loss or damage, including consequential or special damages, resulting from any use of this information, even if loss or damage is caused by Kodak’s negligence or other fault.
  • Page 3: Weekly Preventive Maintenance

    Weekly Preventive Maintenance Section 1: Weekly Preventive Maintenance Important The operator of the PROCESSOR should weekly perform the preventive maintenance procedures in this section. Figure 1 WET SECTION COVER 2 EVAPORATION COVERS WATER RESERVOIR EXIT RACK DETECTOR CROSSOVER DEVELOPER RACK WATER DEVELOPER/FIXER DISCONNECT...
  • Page 4 PERIODIC/PREVENTIVE MAINTENANCE [12] Install the following parts. Check that each assembly seats firmly. a. WASH RACK b. all CROSSOVERS [13] Connect the WATER RESERVOIR to the CROSSOVERS. [14] Check that the WATER DISCONNECT is positioned correctly. [15] Install the following parts. Check that each assembly seats firmly. a.
  • Page 5: Monthly Preventive Maintenance

    Monthly Preventive Maintenance Section 2: Monthly Preventive Maintenance Important Completion of the monthly preventive maintenance procedures is recommended to ensure the optimum reliability of the PROCESSOR. [1] De-energize the PROCESSOR. [2] Lift the TOP COVER. [3] Remove the following parts: a.
  • Page 6 PERIODIC/PREVENTIVE MAINTENANCE [13] Wipe any deposits from the processing section of the PROCESSOR. [14] Change the DEVELOPER FILTER. If necessary, see the procedure in the Adjustments and Replacements Manual, Publication Number 5B6331. [15] Allow the DETECTOR CROSSOVER to air dry before you install it in the PROCESSOR. [16] Clean the WATER RESERVOIR.
  • Page 7: Month Preventive Maintenance

    3 Month Preventive Maintenance Section 3: 3 Month Preventive Maintenance [1] De-energize the PROCESSOR. [2] Lift the TOP COVER. [3] Open the FIXER and DEVELOPER DRAIN VALVES to drain the solutions from the PROCESSOR. [4] Remove the following parts: a. WET SECTION COVER b.
  • Page 8 PERIODIC/PREVENTIVE MAINTENANCE [11] Rinse the following parts with warm water, 44 C (110 F) or less. Wipe the ROLLERS and the GUIDE SHOES with a soft, synthetic sponge. a. CROSSOVERS b. EVAPORATION COVERS c. WET SECTION COVER Note If necessary, you may use DEVELOPER SYSTEM CLEANER AND NEUTRALIZER on the DEVELOPER/FIXER CROSSOVER and FIXER/WASH SYSTEM CLEANER on the FIXER/WASH CROSSOVER.
  • Page 9 3 Month Preventive Maintenance Caution • Any obstruction such as biological growth, small bends, or an upward sloping WASH DRAIN HOSE can cause wash water to return to the WASH TANK causing an overflow of water and damage to the ELECTRICAL BOX. •...
  • Page 10 PERIODIC/PREVENTIVE MAINTENANCE Figure 2 SPRING SPADE H104_0383BCB H104_0383BA [29] Check the SPRING SPADES on the LEVEL SENSOR PROBES for corrosion. If necessary, clean or install new SPRING SPADES. [30] Check the operation of the DETECTOR SWITCHES. If necessary, adjust the DETECTOR SWITCHES. See the procedures in the Adjustments and Replacements Manual, Publication Number 5B6331.
  • Page 11: Preventive Maintenance Schedule

    Preventive Maintenance Schedule Section 4: Preventive Maintenance Schedule Monthly†Ä†† Part or Assembly Weekly* †Ä†ë† 3 Month Schedule‡ DETECTOR CROSSOVER DEVELOPER/FIXER CROSSOVER FIXER/WASH CROSSOVER WASH RACK C(2) DEVELOPER RACK C(1) FIXER RACK C(2) DRYER RACK TOP EXIT RACK RACK BEARINGS, CHAINS, and DRIVE GEARS Squareness of RACKS DRYER AIR TUBES DEVELOPER FILTER...
  • Page 12 PERIODIC/PREVENTIVE MAINTENANCE November 1995 – 5B6746...
  • Page 13: Publication History

    Publication History Section 5: Publication History Affected Print Date Pub. No. ECO No. Pages File Name Description October 1995 5B6746 2650-030 All Pages pm3434_1_030.doc 1st Printing of Manual November 1995 5B6746 2650-039 Front and Back pm3434_1_039.doc Graphic Unification Printing Covers January 1998 5B6746 2650-039...