Konica Minolta REGIUS SIGMA2 Operation Manual page 50

Direct digitizer
Table of Contents

Advertisement

5 .1 Troubleshooting
Error Information
Error
Code
7400
Initialization failed . Reader error was detected
- 7409
during initialization .
7410
Erasure error was detected .
7411
7412
Reader error was detected .
- 7414
Initialization failed . Reader error was detected
7415
during initialization .
7416
Reader error was detected .
- 7419
Initialization failed . Reader error was detected
7420
during initialization .
7421
Reader error was detected .
7422
7650
Reader error was detected .
7651
7700
Initialization failed . Reader error was detected
7701
during initialization .
7710
Detected Memory error .
- 7746
7750
Detected Data error .
- 7759
7800
Parameters which cannot be processed in the
7801
Reader Settings are specified
7802
Reader error was detected .
Parameters which cannot be processed in the
7803
Reader Settings are specified
7810
Top door open
7811
Back door open
7812
Left side door open
Program data of service call reader corrupted .
7999
Need to recover the program .
Message
Appropriate action
Open and close the top door . The device will be re-initial-
ized, so check to see if the error occurs again .
An abnormality was detected in the image-erasing device .
Now that image reading is complete, wait for the image to
be transmitted. After confirming that transmission is com-
plete, refer to "5 .3 Forcibly Removing the Cassette"(p . 61)
and remove the cassette .
Then contact Konica Minolta technical representatives .
Refer to "5 .2 Dealing with Plate Jams"(p . 54) and remove
the plate and cassette from this device .
Then contact Konica Minolta technical representatives .
Open and close the top door . The device will be re-initial-
ized, so check to see if the error occurs again .
Refer to "5 .2 Dealing with Plate Jams"(p . 54) and remove
the plate and cassette from this device .
Then contact Konica Minolta technical representatives .
Open and close the top door . The device will be re-initial-
ized, so check to see if the error occurs again .
Refer to "5 .2 Dealing with Plate Jams"(p . 54) and remove
the plate and cassette from this device .
Then contact Konica Minolta technical representatives .
Contact Konica Minolta technical representatives .
Open and close the top door . The device will be re-initial-
ized, so check to see if the error occurs again .
Once this error occurs, it cannot be solved by turning the
device power OFF and ON again . Contact Konica Minolta
technical representatives .
An abnormality was detected in the transmission data .
Confirm transmission data and communication cable. Then
turn the device power OFF and ON .
An abnormality was detected in the cassette size param-
eters .
Refer to "5 .2 Dealing with Plate Jams"(p . 54) and remove
the plate and cassette from this device .
Then contact Konica Minolta technical representatives .
Refer to "5 .2 Dealing with Plate Jams"(p . 54) and remove
the plate and cassette from this device .
Then contact Konica Minolta technical representatives .
Examination requirement(s) have been made which can-
not be fulfilled by this device. Review current examination
requiment setting .
Close the top door .
Close the rear door .
Close the left door .
Contact Konica Minolta technical representatives .
50