Pa 5840 Front View - Ametek 5000 Series Quick Start Manual And Safety Instructions

Automotive immunity test generators
Hide thumbs Also See for 5000 Series:
Table of Contents

Advertisement

AMETEK CTS
4.3.

PA 5840 front view

1
Voltage display
2
SYSTEM STOP switch
3
Current display
4
CURRENT LIMIT
5
ERRORS indicators
6
COMPENSATION MODES
7
CURRENT LIMITATION
8
OUTPUT RANGE
1
Voltage display
4 digit / seven segment display to show generated dc voltage (RMS)
2
SYSTEM STOP switch
Emergency stop to interrupt a running test. Test output is disconnected.
After activation the PA 5840 needs to re-power up (switch off/on mains switch at the back) to properly
operate again.
3
Current display
4 digit / seven segment display to show the current output (RMS).
4
CURRENT LIMIT
Pushing the button shows the set current limit value (selectable in the software.
Red LED indication shows when the current limiter kicks in.
5
ERRORS indicator
Shows if an error occurs in the system. Different error indications are SAFETY FAIL, POWER FAIL and
OVERTEMP.
6
COMPENSATION MODES
Depending in connected DUT and generated signal, changing to a different compendation mode might be
helpful for testing. STANDARD (up to 40kHz) / CAPACITIVE (up to 3kHz) / HIGH FREQ (>40kHz).
Automotive Immunity Test Generators
9
10
11
12
13
14
15
16
V 1.00
POWER switch
GAIN
Cooling
Sense lines
+ / - output
BNC I INPUT
CONTROL
BNC U INPUT
17 / 27