NI FD-11605 Calibration Procedure

8-channel, ±60 v voltage input device for fielddaq

Advertisement

Quick Links

CALIBRATION PROCEDURE
FD-11605
8-Channel, ±60 V Voltage Input Device for FieldDAQ
This document contains information for calibrating the FD-11605. For more information about
calibration, visit ni.com/calibration.
Contents
Software Requirements............................................................................................................. 1
Documentation.......................................................................................................................... 2
Test Equipment..........................................................................................................................2
Test Conditions..........................................................................................................................3
Calibration Procedure............................................................................................................... 3
Initial Setup....................................................................................................................... 4
Verification........................................................................................................................4
Adjustment........................................................................................................................ 7
EEPROM Update.............................................................................................................. 8
Reverification....................................................................................................................9
Accuracy Under Calibration Conditions...................................................................................9
Signal Connector Pinout........................................................................................................... 9
Revision History..................................................................................................................... 10
NI Services.............................................................................................................................. 10

Software Requirements

Calibrating the FD-11605 requires the installation of NI-DAQmx on the calibration system. NI
recommends using the latest NI-DAQmx driver. Earliest driver support version for calibrating
the FD-11605 is listed in the following table.
Driver
NI-DAQmx
You can download NI-DAQmx from ni.com/downloads. NI-DAQmx supports many
programming languages, including LabVIEW, LabWindows
Basic .NET. When you install NI-DAQmx, you only need to install support for the application
software that you intend to use.
Table 1. FD-11605 Driver Support
Earliest Version Support for Device Calibration
19.0
/CVI
, C/C++, C#, and Visual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the FD-11605 and is the answer not in the manual?

Questions and answers

Summary of Contents for NI FD-11605

  • Page 1: Table Of Contents

    You can download NI-DAQmx from ni.com/downloads. NI-DAQmx supports many ™ ™ programming languages, including LabVIEW, LabWindows /CVI , C/C++, C#, and Visual Basic .NET. When you install NI-DAQmx, you only need to install support for the application software that you intend to use.
  • Page 2: Documentation

    NI-DAQmx C Reference Help—Provides reference information for NI-DAQmx C functions and NI-DAQmx C properties. Test Equipment The following table lists the equipment required for calibrating your FD-11605. If you do not have the recommended instruments, use the minimum requirements to select substitute equipment.
  • Page 3: Test Conditions

    22 AWG signal wires Test Conditions The following setup and environmental conditions are required to ensure the FD-11605 meets published specifications: • Keep connections to the device as short as possible. Long cables and wires act as antennas, picking up extra noise that can affect measurements.
  • Page 4: Initial Setup

    Initial Setup Complete the following steps to set up the FieldDAQ device. Install the software and NI-DAQmx driver as outlined in the FD-11605 Quick Start. Note You must install NI-DAQmx 19.0 or later for device calibration support. Set up the FieldDAQ device as outlined in the FD-11605 Quick Start.
  • Page 5 Test Point 1-Year Limits Range Location Value (V) Lower Limit (V) Upper Limit (V) 16 V Negative Full Scale -15.0 -15.0154 -14.9846 -0.0010 0.0010 Positive Full Scale 15.0 14.9846 15.0154 FD-11605 Calibration Procedure | © National Instruments Corporation | 5...
  • Page 6 Repeat Steps 3 through 6 for each test point in the 60 V range. Set the calibrator to Standby mode (STBY). 10. Disconnect the calibrator from the FieldDAQ device. 11. Connect the calibrator to Bank 2 of the FD-11605, as shown in the following figure. Refer to Signal Connector Pinout for signal connection information.
  • Page 7: Adjustment

    FieldDAQ function. Call the Get FD-11605 Calibration Adjustment Points function. Set the calibrator to a reference value determined by the array of adjustment points. Set the calibrator to Operate mode (OPR). FD-11605 Calibration Procedure | © National Instruments Corporation | 7...
  • Page 8: Eeprom Update

    Repeat steps d through e for each calibration voltage in the array. Set the calibrator to Standby mode (STBY). Disconnect the calibrator from the FieldDAQ device. Connect the calibrator to Bank 2 of the FD-11605, as shown in the following figure. Refer to Signal Connector Pinout for signal connection information.
  • Page 9: Reverification

    FD-11605. For operational specifications, refer to the most recent FD-11605 Specifications at ni.com/manuals. Signal Connector Pinout Figure 6.
  • Page 10: Revision History

    . You can find patents.txt ni.com/patents information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at for the NI global trade compliance policy and how ni.com/legal/export-compliance...

Table of Contents