5.1.5.1 Sensor Trigger Function
The Sensor Trigger function is used to reset the IPSO min/max values as well as controlling the Calibration process.
7
0
15
0
Setting the Reset Min/Max bit to 1 will reset the Min/Max values recorded by the IPSO process.
No User Calibration process is supported on the Digital inputs and all configuration bits should be written as 0.
5.2
Digital Input / Output Descriptor
The DIO Interface provides 2 digital inputs which are hardwired to the Digital outputs. These may be used to detect the
state of external switches (output off) or to monitor the state of the outputs. The DIO Input descriptor is at base addresses
0x0068.
Offset
0x00
0x01
0x02
0x03
0x04..0x08
5.2.1 DIO Sensor Type
The interface provides a bit mapped input of the 2 digital signal lines.
5.2.2 DIO Data Type/Format
7
Smart Sensor
0
Note:
Please refer to the Smart Sensor Interface Technical Guide for more information regarding this descriptor.
5.2.2.1 Data Type
The 4-bit Data Type field determines the type of date of the specific sensor.
5.2.2.2 Factory Calibrate
The Factory Calibrate bit is not used for DIO types.
6
5
0
0
14
13
0
Calibration
Reset
Name
Sensor Type
Data Type/Format
Configuration
Sensor Device
UOMR
Sensor Type
SI Derived Units
0x18
6
5
Sensor
Factory
Writable
Calibrate
0
0
Sensor Trigger Function
4
3
0
0
12
11
Calibration
Calibration
Status
Mode
Value
0x18
Digital Type (Bit mapped)
0x46
Configurable, Float type
0x23
Scaling applied, Bits 0 and 1 enabled
0x0f
DIN bits enabled / inverted
"DIN"
Units of measure
Measurement
DIN
Bit mapped digital inputs
DIO Data Type
4
3
Reserved
0
2
1
0
0
10
9
Capture
Capture
High
Low
Description
2
1
Data Type
6 == Floating point
0
Reset
Min/Max
8
Calibration
Start
0
17 |
P a g e