8. Fixing the tare weight in position
Fix the tare weight in position using the
knurled nut. Objects on the tray such as
pads etc. weighing up to 3000 g / 6.6 lbs
can be tared using the tare weight.
9. Fine adjustment
The hand and counter-hand can be set
exactly by turning the fine taring screw. To
do so, the fine taring screw is turned to the
right or left until the hand and counter-hand
line up exactly.

Operation

Once the scale has been prepared as
described in the section entitled "Set up
and align scale", you can start weighing.
Place the infant in the tray and first slide the
large sliding weight to approximately the
right weight.
Now use the small sliding weight to
determine the exact weight. To do so,
slide the small sliding weight along until
hand and counter-hand line up.
Read the weight value off the scale on the
sliding weight bar.
Cleaning
Clean the housing and the tray as required
using a domestic cleaning agent or
commercially-available disinfectant.
Follow the manufacturer's instructions.