Page 1
BM3 User’s Manual User’s Manual Patient Monitor Rev. 3.0 2019.11.22 Warning To ensure proper use of this medical equipment, you must read and comply with this user manual.
Page 2
Reproduction in any manner, in whole or in part, except for brief excerpts in reviews and scientific papers, is prohibited without prior written permission of Bionet, Co., Ltd. Before using Bionet devices, read all the manual that are provided with your device carefully. Patient monitoring equipment, regardless of the complexity of the equipment, should never be used as a substitute for the patient care, attention, and critical judgment that only trained health care professionals can provide.
This equipment is connected via BM central. Note All Bionet hardwares and screenshots in this user guide are for illustration purposes only. Actual products or screens may vary slightly. General Description...
BM3 User’s Manual Patient Classification BM3 monitors are designed for use by adults, pediatrics and neonates. At this time, cardiac output, ST segment analysis and arrhythmia should be used for adults and pediatrics only. Functional safety The essential performance of the patient monitor is to provide the clinician with meaningful parameter values and to sound an alarm when the established parameter value is exceeded or the function that provides the value is not working properly.
BM3 User’s Manual Warning, Caution, Note The following terms are defined in the User Guide to emphasize the agreement as follows: The user must follow all warnings and precautions. The specifications and functions shown in this manual are subject to change without prior notice.
BM3 User’s Manual Define groups The define groups for this product are users, service personnel, and experts. Define groups should read the user manual before using the product and be trained in the use, installation, reprocessing, maintenance and repair of the product.
BM3 User’s Manual General precaution on environment Do not keep or operate the equipment in the environment listed below. Avoid placing in an area exposed to moist. Avoid exposure to Do not touch the direct sunlight equipment with wet hand.
The monitor complies with the following standards pertaining to EMI emissions and susceptibility: EN60601-1-2. To reduce possible problems caused by electromagnetic interference, we recommend the following: Use only Bionet approved accessories. Ensure that other products used in areas where patient monitoring and life ...
Please check the following before using the product. 1. Be sure that AC power supply line is appropriate to use. (AC100 - 240V) 2. Be sure that the power source is the one supplied from Bionet.(DC18V, 2.8A, BPM050 Made in BridgePower Co., Ltd.) 3.
Page 16
Note BM3 is classified as follows: - BM3 classifies as Class II, BF & CF concerning electric shock. It is not proper to operate this Equipment around combustible anesthetic or dissolvent. - Noise level is A class regarding IEC/EN 60601-1 and the subject of Nose is A level concerning IEC/EN60601-1-2.
Warning In order to avoid electrical shock, do not open the cover. Disassembling of the equipment should be done only by the service personnel authorized by Bionet Warning Users must pay attention on connection any auxiliary device via LAN port or nurse calling.
BM3 User’s Manual Basic Unit Front view Alarm lamp handle Home key Alarm control key Rotary knob key Printer key Power ON/OFF Key Blood-pressure measurement key Battery status indicator --------------------------------------------------------------------------------------------------------------------------------------- Rev. 3.0...
Page 20
BM3 User’s Manual Right side view ECG connector Blood pressure Hose connector SpO2 connector Temperature connector EtCO2 connector --------------------------------------------------------------------------------------------------------------------------------------- Rev. 3.0...
Page 21
BM3 User’s Manual Left side view Printer --------------------------------------------------------------------------------------------------------------------------------------- Rev. 3.0...
Page 22
BM3 User’s Manual Back side view Potential equivalent NURSE CALL connector USB connector (USB 2.0 5Vdc / Max. 500mA), DC input Service port connector Network connector HDMI output --------------------------------------------------------------------------------------------------------------------------------------- Rev. 3.0...
Page 23
BM3 User’s Manual Warning USB Compatible The BM3 is compatible with external USB memory drives up to 64GB. We recommend brands products listed in the manual (Sandisk, PNY, Transcend, Samsung). When using a product with high power consumption, such as an external hard ...
BM3 User’s Manual Device Markings Caution : Ground terminal Consult accompanying documents TYPE CF APPLIED PART TYPE BF APPLIED PART Printer Auxiliary Port LAN port HDMI HDMI external port DC Input Indicator USB port Battery Operation indicator DC input connector...
BM3 User’s Manual Power The BM3 monitor uses a DC adapter (100-240 VAC / 18VDC 2.8A). In the event of a power outage or cable shortage, the monitor automatically switches to battery power to continue patient monitoring without data loss. The built-in battery is intended for back- up use only during power-off.
Page 26
BM3 User’s Manual Operation 1. Battery Power LED is lighted on when the machine is in use. 2. Battery is automatically charged when the machine is connected to DC Power Supply. The charging status is displayed at the top right of the screen 3.
Page 27
To maximize the charge for transport, keep the monitor connected until you are ready to transport the patient. Reconnect the monitor immediately after transport. Bionet recommends replacing the lithium ion battery after 24 months of use. Battery life depends on usage. If battery life continues, battery life will ...
Page 28
Warning Be careful of the polarity when replacing the battery. We strongly recommend that you use the battery supplied by Bionet. Using unauthorized batteries may damage the equipment 5. Presence of battery: When the battery is disconnected from the equipment and it malfunctions, it shows 'X' as shown below.
BM3 User’s Manual How to replace the battery Please assemble and replace as shown below. The Impact of Lithium-Ion Battery Technology on the Battery The following are the key points you should know about Lithium-Ion battery technology: The battery will discharge on its own, even when it is not installed in a monitor. This discharge is the result of the Lithium-Ion cells and the bias current required for the integrated electronics.
Page 30
12 months. Bionet recommends that you remove the battery and store it near the monitor until it is needed for transport.
BM3 User’s Manual Getting Started Starting the monitor: Press the power key (O) at the bottom right of the monitor front panel. The power light on the monitor lights up, the alarm bar lights up, the power is turned off, the screen lights up, the main screen is displayed after running the self-test.
Page 32
BM3 User’s Manual The parameter box displays values, alarm limits and icons for the selected parameter. You can set the parameters and their associated waveforms so that they are easy to distinguish. The message appears at the top of the screen. The patient name bed label is displayed in the upper left corner of the screen.
Page 33
BM3 User’s Manual Function key On the right side of the monitor's front panel, the touch screen icon on the touch screen allows you to perform frequently-used functions. Fixed key Description Fixed key Description Opens a table where you This is an alarm mode key, so...
BM3 User’s Manual 2. SETUP Overview This chapter describes how to configure your monitor and how to upgrade your software. Monitor configuration Setup Menu tree Parameter menu tree --------------------------------------------------------------------------------------------------------------------------------------- Rev. 3.0...
BM3 User’s Manual Main menu setup The Setup menu allows the user to access submenus, display screens, and perform specific monitor setup functions. 1. To display the Settings menu, click the Settings icon to open the submenu. 2. Click the desired setting to access the submenu that performs the desired function or goes one step further down.
Page 36
BM3 User’s Manual A. SETUP menu MENU Description Available settings A-1. PARAMETER SETUP measurement on the monitor PARAMETER enable Parameter selection and color setting ON/OFF menu:ECG,SPO2,RESP,NIBP,TEMP, PARAMETER COLOR ETCO2 setup A-2. PARAMETER UNITS Unit setting menu used for monitor measurement A-2-1.
Page 37
BM3 User’s Manual A-3-2. BED # Set device number 1~300 A-3-3. KEY Sound Set Key activation ON / OFF A-3-4. KEY Volume Set Key sound OFF ~ 100% A-3-5. AC FILTER Power filter settings OFF, 50Hz, 60Hz A-3-6. SCREEN BRIGHTNESS...
Page 38
BM3 User’s Manual A-6-2. CENTRAL Remote Communication menu ON/OFF A-6-3. Server IP Remote PC IP address setting XXX.XXX.XXX.XXX A-7. HL7 Network message settings A-7-1. Communication version A-7-2. Server Remote PC IP address setup XXX.XXX.XXX.XXX A-7-3. PORT Remote PC PORT address XXXX A-7-4.
Page 39
BM3 User’s Manual A-10-2. Address 1 Address information 1 A-10-3. Address 2 Address information 2 A-10-4. Postal Code Set postal Code B. BIOSIGNAL CALIBRATION Set calibration menu B-1. ECG & RESP B-1-1. ECG Calibration ECG calibration menu 10mm/mV input calibration display B-1-2.
Page 40
BM3 User’s Manual Parameter color Parameter Basic color Selectable colors Green, light blue, yellow, purple, blue, sky blue, orange, gray, light green, pink, white, red, light yellow ECG (ST) Green SpO2 Blue RESP Yellow NIBP Purple TEMP Green ETCO2 Yellow --------------------------------------------------------------------------------------------------------------------------------------- Rev.
BM3 User’s Manual 3. Network Overview When you connect the monitor to the network, you can access patient information from another monitor or central station connected to the network. These devices provide main screen information for remote viewing from each other.
Upgrading or updating devices connected to the network Warning Recommendations for wireless connections BM3 has a change in the number of equipment connections depending on wireless AP (Access Point) performance. When using a general AP, it is recommended to connect 8 units to the same network.
Page 43
BM3 User’s Manual MENU Description Available Settings B2B VIEW MONITOR LIST sub menu REFRESH Menu to update monitor list connected to network Move to upper list DOWN Move to lower list MONITOR LIST List of monitor information connected to the network...
Page 45
BM3 User’s Manual MENU Description Available settings A-1 SELECT WAVE Waveform selection menu to view ECG, SPO2, RESP,ETCO2, The waveform selection menu for A-1-1. TRACE I TRACE I in the B2B View window The waveform selection menu for ECG, SPO2, RESP,ETCO2 A-1-2.
BM3 User’s Manual 4. Admission and Discharge Overview The Patient admit menu allows you to enter and edit a patient’s personal data (name, ID, Birthday, Height, Weight). If your monitor is operating in a network monitoring, you can also review or change the monitor’s care unit and bed label assignments.
BM3 User’s Manual Patient discharge The patient should be discharged before the other patient is admitted. Otherwise The monitor attaches the existing data to the patient in the back of the hospital. How to discharge a patient: 1. Press the Discharge. fixed key.
Page 48
BM3 User’s Manual Main menu Sub menu A.Admit / Discharge B.Patient Information B-1. Patient Information C.Default Setting D.User Drug Change E.Drug Calculation E-1. Setting E-2. Titration Table MENU Description Available Settings A. Admit / Discharge Admission and discharge setting B. Patient Information B-1.
Page 49
BM3 User’s Manual E-1-1. DRUG TYPE Choose Drug Type in the List(21) AMINOPHYLLINE BRETYLIUM LIDOCAINE PROCAINAMIDE EPINEPHRINE LEVOPHED ISOPROTERENOL DOPAMINE DOBUTAMINE NITROGLYCERINE NITROPRUSSIDE INOCOR HEPARIN INSULIN STREPTOKINASE DRUG-1~5 E-1-2. Drug Quantity Set Drug Quantity E-1-3 Solution Volume E-1-4. Dose Quantity Set Dose Quantity E-1-5.
Page 50
BM3 User’s Manual Refer drug list E-2-5. Inflation Rate bellow. Refer drug list E-2-6. WEIGHT bellow. Refer drug list E-2-7. DOSE STEP bellow. The table shows the formula for calculating the dosage of drugs below. Unit NameUnit Equation Dose ( mg/hr ) × SolutionVolume(ml)
Page 51
BM3 User’s Manual Shows each unit setting range. Dose Drug QTY Unit Proper Range Unit Proper Range mg/hr mg/min mg/kg/hr mg/kg/min 0.01 to 500 0.01 to 2000 mcg/hr mcg/min mcg/kg/hr mcg/kg/min units/hr 10 to 15000 Units 100 to 150000 IU/hr...
BM3 User’s Manual Registration of patient ID using barcode This product can input the PATIENT ID in barcode format to the device using USB barcode scanner. First, connect the barcode scanner to the USB HOST connector on the left as shown in the figure below.
BM3 User’s Manual 5. Alarm Overview The monitor displays the alarm limit (parameter threshold) and can be configured by the user to raise an alarm if exceeded. Limits are displayed both in the alarm limits table and in the parameter box. If this limit is exceeded, a visual or audible alarm will occur.
Page 54
BM3 User’s Manual Alarm priority Alarm sound Alarm Color Alarm printer Alarm lamp HIGH 2.0 Times/Sec Blinking MEDIUM 0.5 Times/Sec Blinking Non Blinking MESSAGE : Alarm sounds : Red Lamp is turned on. : Waves are printed out : Red color alarm indicator on the screen is blinked...
BM3 User’s Manual Alarm management You can use the lock key on the front of the monitor to hold the alarm. To change Alarm Mode: A short press of the alarm control key circulates through the Normal / Audio_Paused / Alarm_Paused alarm modes. Press and hold the key for more than 3 seconds to...
BM3 User’s Manual Note Audio Paused and Audio Off modes only stop the audible alarm sound and touch or key sound is activated always. To adjust the Touch or Key Sound, use the Key Sound menu in Setup. ...
BM3 User’s Manual to 10% to 100%. A-4-2. Alarm Pause Time No sound for 5minutes, Release on 1,2,3,5,10,15min alarm again A-5. Nurse call User Settings menu. NURSE CALL function ON / OFF; After ON/OFF A-5-1. Nurse call on Alarm setting ON, check if relay sound is heard in ALARM situation.
BM3 User’s Manual 6. TREND Overview The monitor stores trend data for all connected signals. Users can request trend recording and can also print the screen of trends displayed. Triggered alarm events are displayed in red inverted triangles on the Event List and...
Page 59
BM3 User’s Manual MENU Description Available settings A. Trend Setup menu A-1. Popup Trend A-1-1. Time Period Show time interval setting menu 30min, 60min, 90min, 3hour, 6hour A-1-2. Configure Parameters B. Graphic Trend menu B-1. Graphic Trend B-1-1. Event List Saved data can be viewed graphically B-1-2.
Page 60
BM3 User’s Manual C-1-1. Event List C-1-2. Time Period Time period setting 1min, 5min, 10min, 15min, 30min, 1hour, 2hour C-1-3. Display Group C-1-4. Print Tabular trend print output C-2. Graphic Trend C-2-1. Event List C-2-2. Time Period Time period setting...
BM3 User’s Manual Graphical trend Trend graph shows saved trend data as individual graph type for each parameter. These graphs show that the displayed parameters are active over a significant period of time Shows five channels at a time. Confirmation color and scale Meter labels and numbers are displayed on the left side of the trend channel.
BM3 User’s Manual Parameter selection menu to show ⑨ Printer menu ⑩ Parameter window selection menu ⑪ Tabular trend The Trends table displays the trend data in an easy-to-read table format. Up to six are displayed, updated every minute. The time stamp above each column indicates the interval at which the data in that column was trended.
BM3 User’s Manual Patient ID ⑤ Numeric Parameter window ⑥ Selection Navigation window ⑦ Trend interval setting menu ⑧ Parameter selection menu ⑨ Printer menu ⑩ Parameter select window menu ⑪ File export The file extract function can transfer trend to a file using USB memory.
Page 64
BM3 User’s Manual Warning USB Compatible The BM3 is compatible with external USB memory drives up to 64GB. We recommend brands products listed in the manual (Sandisk, PNY, Transcend, Samsung). When using a product with high power consumption, such as an external hard drive, be ...
BM3 User’s Manual Popup trend The user can continue to monitor the main screen waveform and parameter box while displaying trend data for up to 5 parameters for up to 6 hours. The pop-up trend graph follows the display order indicated by each parameter in the trend setup and is updated with new trend data every 60seconds.
Page 66
BM3 User’s Manual Popup ST window Popup enlarge trend window You can change the size of the popup menu by pressing and releasing the center of the popup menu for at least 1second. --------------------------------------------------------------------------------------------------------------------------------------- Rev. 3.0...
BM3 User’s Manual 7. ECG Overview The monitor can calculate heart rate, detect arrhythmia (adult and pediatric patients), and display ECG data. The electrocardiogram screen provides 1 channel display. It calculates the heart rate by detecting the electrocardiogram signal of the patient and alarms according to the set upper and lower limit of alarm.
BM3 User’s Manual 5. Connect the lead of the electrode and the wire of the monitor 6. Fix the electrode to the skin, and secure the cable with the remaining length between the instrument and the electrode with surgical tape. This fixation prevents the electrode from moving.
Page 69
BM3 User’s Manual do not come into contact with other grounded, conductive parts when connected to the isolated patient input of the device. Such contact would bridge the patient's isolation and cancel the protection provided by the isolated input. In particular, there must be no contact of the neutral electrode and ground.
Page 70
BM3 User’s Manual be complied with. Electro surgery Unit Electrosurgical unit (ESU) emits a lot of RF interference. If the monitor is used with an ESU, RF interference may affect the monitor operation. Locate the monitor as far as possible from the ESU. Locate them on opposite sides of ...
There are a variety of reusable and disposable electrodes available. Choose the electrode that best fits your monitoring situation. Bionet recommends Ag / AgCl disposable electrodes. If you are using an electrode with a gel beforehand, make sure that the electrode is sufficiently gelled.
BM3 User’s Manual ECG lead 5 LEAD electrode placement 3 LEAD electrode placement How to attach neonate electrodes --------------------------------------------------------------------------------------------------------------------------------------- Rev. 3.0...
BM3 User’s Manual Cable color and size AHA:American Heart Association (U.S.A. standard) IEC:International Electro technical Commission (Europe standard) 3LEAD / 5LEAD Lead wire Color code Label Color code Label Right arm White Left arm Black Yellow Right leg Green Black...
BM3 User’s Manual change accordingly. If you select Basic, you can display three basic arrhythmias called ASYS, VFIB, and VTAC. If Full is selected, a separate ARR parameter box will be displayed next to the HR parameter box (for details on selecting the arrhythmia mode, refer to the Arrhythmia setting chapter).
BM3 User’s Manual Display ① ② ③ ④ ⑤ Heart rate detector: It detects heart rate and flickers simultaneously. ① Pace maker: Pace maker signal is detected and flashes simultaneously. ② HR Alarm limits: Heart rate threshold is display. ③...
Page 76
BM3 User’s Manual A. ECG menu MENU Description Available settings A-1. Alarm ECG alarm setting menu A-1-1. PARAMETER ALARM HR, ST, PVC parameter alarm limits, LIMIT level, activation setup menu. A-1-2. TECHNICAL ALARM ECG-LEADFAULT CONDITION ECG-CHECKELECTRODE ECG-HR-SEARCH A-2. QRS VOLUME OFF, 0%~100% QRS detection volume setting menu.
Page 77
BM3 User’s Manual waveform. The ECG channel is selectable from I to I, II, III, aVR, aVL, aVF, V1,V2,V3,V4,V5,V6 3 When using the lead cable selection, A-3-6. TRACE 1 only TRACE I can select I, II, III. 5 lead cable selection I, II, III, aVR, aVL, aVF, V can be selected.
Page 79
BM3 User’s Manual Problem: Inaccurate pacemaker detection Solution: Use pacemaker processing: 1. Select ECG parameter label. 2. Display the lead of ECG with the greatest amplitude in the top waveform position. 3. Select Pace Maker. 4. SELECT PACE MAKER ON.
BM3 User’s Manual 8. Arrhythmia Monitoring(*) Overview On supported models only(*), Arrhythmia monitoring is available for adult and pediatric patients. The selected mode (Lethal or OFF) determines which events are processed. Arrhythmia monitoring is not available for newborns. Arrhythmia monitoring is available for adult and pediatric patients only.
BM3 User’s Manual Arrhythmia Settings A. ECG menu MENU Description Available Settings A-1. Arrhythmia ARRHYTHMIA parameter alarm, level, Activation setup menu. A-1-1. Arrhythmia Type Sets up ON/OFF to indicate OFF, LETHAL detection of diagnosis (Asys, VTAC/VFIB and VTAC). OFF: Do not perform arrhythmia diagnosis.
Page 82
BM3 User’s Manual Warning VENTRICULAR ARRHYTHMISAS The arrhythmia analysis program is intended to detect ventricular arrhythmia. This program is not designed to detect trial or supra ventricular arrhythmias. In some cases, it may not be possible to distinguish the presence or absence of arrhythmias.
If skin conditions change, move the sensor to another location. Change the application site every 4 hours at least. Use only Bionet-designated sensors. Other sensors may not provide adequate protection against defibrillation or may put the patient at risk.
If you use your fingers as a monitoring site, remove the nail polish. Cut the patient's fingernail if needed to improve placement of the sensor. Only use sensors provided by Bionet and apply them according to manufacturer's recommendations on a per-sensor basis.
BM3 User’s Manual SpO2 measurement Cable Note The signal input is a high-insulation port and it is defibrillator proof ( The insulated input ensures patient safety and protects the device during defibrillation and electro surgery. Display ② ① % S pO2 100 / ③...
Page 86
BM3 User’s Manual SpO2 Alarm limits display ③ SpO2 strength indicator ④ %SpO2 Value display ⑤ The current SPO2 value and the derived pulse rate (RATE) are displayed. The block sets indicate the strength of the signal (twenty block bars indicate the strongest signal). The SPO2 measurements are averaged over a 6-second period of time.
BM3 User’s Manual Quality of SPO2 Waveform Under normal conditions, the SPO2 waveform corresponds to (but is not proportional to) the arterial pressure waveform. The typical SPO2 waveform indicates not only a good waveform, but helps the user find a probe placement with the least noise spikes present. The figure below represents an SPO2 waveform of good quality.
BM3 User’s Manual WARNING In the monitoring of patients the coincidence of adverse conditions may lead to a disturbed signal going unnoticed. In this situation artifacts are capable of simulating a plausible parameter reading, so that the monitor fails to sound an alarm. In order to ensure reliable patient monitoring, the proper application of the probe and the signal quality must be checked at regular intervals.
BM3 User’s Manual Status messages Below is a list of system status alarm messages which may be displayed in the SPO2 parameter window during monitoring. CHECK PROBE Reusable finger probe is off the patient. Check the probe. The factory default for this alarm is MESSAGE ALARM.
Impedance breath monitoring should not be considered the only way to detect breathing stops. Bionet recommends monitoring of additional parameters, such as EtCO2 and SpO2, that indicate the patient's oxygen supply status. • If you use an ESU block or cable, the impedance breath monitor may not work and the ...
BM3 User’s Manual Patient Preparation Skin preparation and electrode placement must be properly and carefully monitored in impedance breath monitoring. You can produce reliable results. Follow the same recommendations as ECG monitoring Please. In general, the electrodes should be placed as clean as possible with the 60Hz noise minimized Make it possible to generate a signal.
Page 92
BM3 User’s Manual Respiration connector and measurement cable Respiration connector Connector Respiration cable Note Respiration Rate measures the cable and connector will be used as the ECG and common. --------------------------------------------------------------------------------------------------------------------------------------- Rev. 3.0...
BM3 User’s Manual Display Breathe indicator: indicates the detected breath ① Breathing number : displays the number of respiration per minute ② Respiration alarm limit: indicates respiration limits ③ Apnea limit Setting: Apnea limit sign ④ --------------------------------------------------------------------------------------------------------------------------------------- Rev. 3.0...
BM3 User’s Manual RESP Settings A. RESP menu Menu Description Available Settings A-1. Alarm RESP Alarm setting menu A-1-1. PARAMETER ALARM RR, APNEA Parameter alarm , LIMIT level ,Activation setup menu RESP-CABLE OFF A-1-2. TECHNICAL ALARM CONDITION RESP-LEAD FAULT RESP-CHECK ELETRODE This is for changing the reference LEAD A-2.
Page 95
BM3 User’s Manual 1. Set patient type to Neonatal 2. Connect SpO2 sensor, HR lead and breath or etCO2 lead. 3. Set the apnea time in the RESP menu. 4. Press the TREND icon key. 5. Click OxyCRG to display the OxyCRG screen.
Page 96
BM3 User’s Manual Setup menu OxyCRG MENU Description Available Settings ECG AUTO SCALE ECG RANGE AUTO SCALE setup ON/OFF ECG MANUAL SCALE ECG RANGE MANUAL SCALE setup PARAMETER TYPE RESP, ETCO2 parameter setup menu RESP ETCO2 --------------------------------------------------------------------------------------------------------------------------------------- Rev. 3.0...
BM3 User’s Manual 11. NIBP Overview The monitor can acquire and process non-invasive blood pressure (NIBP) signals and display the output. Blood pressure measurements are determined by the oscillometric method and are equivalent to those obtained by intra-arterial methods, within the limits prescribed by the Association for Advancement of Medical Instrumentation, Electronic Automated Sphygmomanometers (AAMI/ANSI SP-10).
Page 98
BM3 User’s Manual Adult Cuff Optional accessory list Big Adult NIBP Cuff Thigh Adult Cuff Size : 458 * 143 Arm circumference : 45 to 56.5Cm Option Big Adult NIBP Cuff Big Adult Cuff Size : 458 * 143 Arm circumference : 35.5 to 46 Cm Option...
BM3 User’s Manual NIBP Disposable Cuff Neonate 2 (4.2~7.1cm) Option NIBP Disposable Cuff Neonate 3 (5.0~10.5cm) Option NIBP Disposable Cuff Neonate 4 (6.9~11.7cm) Option Note The NIBP should be set in the menu because the measured value differs depending on the patient's age and gender.
BM3 User’s Manual Systolic Alarm limit: Indicates alarm limit of blood pressure. ⑥ Interval Time: indicates interval time when measures the blood pressure ⑦ periodically. Systolic blood pressure: Indicates the maximum limit of blood pressure. ⑧ Diastolic blood pressure: Indicates the maximum limit of blood pressure.
Page 101
BM3 User’s Manual turned on and pressurized to the 80 – 170mmHg maximum pressure range at the initial NEO : pressurization. 60 – 140mmHg Default Settings value: ADT : 170 mmHg PED : 140mmHg NEO : 120mmHg A-4. SETTING TIME Once, How to apply pressure value setting.
Page 102
BM3 User’s Manual Warning Check periodically to see if the circulation from the cuff to the distal part of the patient's arm is good. 1minute and 2minute intervals When using automatic measurement, check the patient's condition frequently. It is not recommended for measuring blood pressure for a long time after the measurement time period is set to 10minutes or less.
BM3 User’s Manual It recommended PATIENT position in NORMAL measurement, as below; 1) Comfortably seated 2) Legs uncrossed 3) Feet flat on the floor 4) Back and arm supported 5) Middle of the CUFF at the level of the right atrium of the heart...
Page 104
BM3 User’s Manual marked on the cuff. Measure the circumference of your patient’s limb. Use only Bionet cuffs with your monitor. Warning Non-invasive blood pressure monitoring is not recommended for patients with hypotension, hypertension, arrhythmias or extremely high or low heart rate. The software algorithm cannot accurately compute NIBP or patients with these conditions.
BM3 User’s Manual Status Messages If the cuff hose is not connected properly INFLATION FAILURE CHECK CUFF When the cuff pressure is excessive OVER PRESSURE When the cuff breaks and cannot exhaust DEFLATION FAILURE When the cuff pressure exceeds the set time ...
BM3 User’s Manual 12. EtCO2(*) Overview On supported models only(*), the BM Series monitor measures concentrations of end-tidal CO2 (EtCO2) when this option is enabled and the EtCO2 module is connected to your monitor. The EtCO2 module can perform mainstream measurements in all monitoring modes and sidestream measurements in the adult and pediatric monitoring modes.
Page 107
BM3 User’s Manual Sidestream EtCO2 Accessories Intubation Sidestream accessories PART FIGURE Description type 3468ADU-00 Nasal CO Sampling Cannula Adult 3468PED-00 Nasal CO Sampling Cannula Child 3468INF-00 Nasal CO Sampling Cannula Neonate Oral/Nasal CO Sampling Cannula 3470ADU-00 Adult 3470PED-00 Oral/Nasal CO...
Page 108
BM3 User’s Manual Nasal CO Sampling Cannula w/ O 3469INF-00 Neonate Delivery Oral/Nasal CO Sampling Cannula w/ O 3471ADU-00 Adult Delivery Oral/Nasal CO Sampling Cannula w/ O 3471PED-00 Child Delivery Intubation accessories Airway Adapter Kit w/ Dehumidification Adult /chid 3473ADU-00 Tubing (ET Tube Size >4.0 mm)
Page 109
BM3 User’s Manual EtCO2 Placements and Accessories (Mainstream, Respironics) EtCO2 connector CAPNOSTAT 5 mainstream CO2 sensor and connector Mainstream Sensor Mainstream Sensor Connector --------------------------------------------------------------------------------------------------------------------------------------- Rev. 3.0...
BM3 User’s Manual Precaution Warning The safety and efficacy of breath measurement methods for apnea detection, especially apnea of premature babies and apnea of infants, have not yet been established. Patient monitors that measure CO2, anesthetics, and / or respiratory mechanics ...
BM3 User’s Manual Sampling method Connecting the CAPNOSTAT® 5 CO2 Sensor to the Host System 1. Insert the CAPNOSTAT 5 CO Sensor connector into the receptacle of the host monitor as shown in Figure 1. 2. Make sure the arrows on the connector are at the top of the connector and line up the two keys of the connector with the receptacle and insert.
Page 113
BM3 User’s Manual Connecting the LoFlo Sample Kit 1. The sample cell of the sampling kit must be inserted into the sample cell receptacle of the LoFlo CO Module as shown in Figure 1. A “click” will be heard when the sample cell is properly inserted.
BM3 User’s Manual Display ① ② A W R R E tC O 2 F iC O 2 5 0 / 3 0 S ③ ④ ⑤ EtCO2 CO2 concentration alarm upper and lower limit value display ① Apnea alarm set time in seconds ②...
Page 115
BM3 User’s Manual ETCO2-CO2 INVALID ETCO2-OVER RANGE ETCO2-ZERO REQUIRED ETCO2-SYSTEM FAULT ETCO2-TEMP UNSTABLE A-2. DISPLAY OPTION EtCO2 Parameter Wave Display Setup Menu A-2-1. SWEEP SPEED Waveform sweep speed setup 6.25mm/s, 12.5mm/s, 25mm/s A-2-2. SCALE Display waveform scale setup. 40mmHg (5.3 vol%) The selectable value is the maximum 50mmHg (6.6 vol%)
Page 116
BM3 User’s Manual A-3. APNEA DETECT APNEA detection menu ON/OFF A-4. MODULE INFORMATION A-4-1. SENSOR PN PNXXXXXX The sensor part number A-4-2. OEM ID The id is a 7bit identifier which is set 0X01 at the factory to a unique value for each OEM.
Page 117
BM3 User’s Manual the measurement. the type of gas that is mixed with the HELIUM breathing gas measuring A-5-3. SLEEP MODE Sleep mode is used to save power NORMAL MODE when the host monitor is in standby TURNOFF MODE mode. There are two sleep modes POWER SAVING available for the Capnostat.
Page 118
BM3 User’s Manual (the default). Only change to nitrogen ) when performing a zero on 100% gas; this is provided for use in a laboratory environment. B-1. ZEROING This function is used to initiate a Capnostat Zero. A zero is used to correct for differences in airway adapter types.
BM3 User’s Manual Note For best result, connect the CAPNOSTAT 5 CO2 Sensor to an adapter and wait 2minutes before performing the Adapter Zero procedure. Status Message Following is a list of some of the message that may appear on the monitor when monitoring CO2.
Page 120
BM3 User’s Manual * ZERO REQUIRED - Cause : Zero Required , Zero Error - Solution : To clear, check airway adapter and clean if necessary. If this does not correct the error, perform an adapter zero. If you must adapter zero more than once, a possible hardware error may exist.
Page 121
BM3 User’s Manual Note In the following monitoring conditions, the measured values may be inaccurate. Read the measured values carefully. 1. When using this in an environment of using nitrous oxide gas of high concentration 2. When using this in an environment where abrupt temperature change takes place 3.
BM3 User’s Manual 13. Temperature Overview This function is used to indicate the changes of resistance generated by the changes of temperature in numbers. The function involves the process of transferring the changes into electric signals. Temperature Connector and Measuring Cable...
BM3 User’s Manual Note Temperature probe is correctly positioned and fixed to do not disconnect on the patient. Temperature cable is attached to the monitor. Display Temperature alarm limit display ① Temperature value display ② Temperature unit display ③ Note...
BM3 User’s Manual Warning To measure the ambient temperature, connect the probe to your ankle or wrist. If the patient is sweating or moving heavily, fix the pads with surgical tape. Temperature settings A. Temp menu MENU Description Available Settings A-1.
BM3 User’s Manual 14. Printer Overview The monitor in order to print out monitoring data, including trends and alarm data. Recordings of waveforms are either timed or continuous and print at a recording speed of 25mm/s. All recordings are identified by the patient’s name, ID as well as the date and time of the recording request.
This is configuration of printed time in normal printing. Delay (5sec) If the print out is not stopped in manual by PRINTER KEY, BM3 print out for setup time after starting print out with PRINTER KEY. REAL TIME: Prints the data from the point where the PRINTER key was pressed.
Page 127
BM3 User’s Manual DELAY: Prints data before 5seconds when PRINTER key is pressed A-1-7. Time Interval Continue, 10sec, Set the time for printing the printout on normal printout. If you do not stop 20sec, 30sec manually after pressing the PRINTER KEY, the output will be output only for the following period of time.
BM3 User’s Manual Thermal Paper Storage To avoid print quality degradation or attenuation of printouts, follow these precautions: Note These precautions apply to both unused paper as well as paper that has already been run through the printer. • Store in cool, dark locations. Temperature must be below 27°C (80°F). Relative humidity must be between 40% and 65%.
BM3 User’s Manual when properly imaged and stored for about 3-5 years. If your retention requirements exceed these guidelines, we recommend you consider alternate image storage techniques. Paper Change Open the window of the printer. Insert the paper roll offered with the product into the printing unit.
Examine all system cables, the power plug for damage. Make sure that the prongs of the plug do not move in the adaptor. If damaged, replace it with an appropriate Bionet power cord and adaptor. Inspect the parameter cable and ensure that it makes good connection with the ...
BM3 User’s Manual Monitor. Make sure that there are no breaks in the insulation. Apply the transducer or electrodes to the patient, and with the monitor switched on, flex the Patient cables near each end to make sure that there are no intermittent faults...
Page 132
BM3 User’s Manual Parameter Module Tests Performance assurance for all measurements At least once every two years, or if you suspect not listed below. the measurement values are incorrect. Parameter Module Maintenance NBP calibration At least once every two years, or as specified by local laws.
BM3 User’s Manual Noise in ECG Gel is dry - Electrodes does not stick well to skin Apply ECG gel or water to the area of The patient's skin contact then use the electrode is extremely dry Relpace the electrodes...
BM3 User’s Manual SpO2 malfunction Connectors of the equipment’s are in bad condition? The extension cables Replace extension cables Are disconnected The finger probe is Replace the finger probe in bad condition? Repair the ECG B/D Temperature malfunction - If the temperature cannot be measured, check the connection with the equipment...
BM3 User’s Manual NIBP malfunction - Connector connection status, confirmation that the hose is normally connected Are leaks from the hose Replace the horse of cuff connector of cuff? Repair the NIBP B/D Abnormality in NIBP measurements Make sure the patient stay...
BM3 User’s Manual EtCO2 malfunction The extension cables Replace extension cables Are disconnected The module is Replace the module in bad condition? Repair the adaptor Failure in battery recharge (the battery does not fully recharge in 6 hours or more)
BM3 User’s Manual Power failure Replace the adapter The adapter connector is in bad condition The output voltage of the adpateris lower Replace the adapter than 18V Repair the power B/D Data storage failure Execute the "admit" function "Admit" has been selected in the menu...
BM3 User’s Manual Periodic noises All alarm has been turned off Turn functions not in use off Display lead fault (Mode display set para) Message appears Replace the printer The printer function abnormally Repair the accompanying Equipment Print failure Shut the cover tight...
We recommend the following cleaning solution and procedures. To avoid contamination and unnecessary damage to the equipment, follow the instructions below. Bionet does not claim the right to the following chemical efficacy, disinfectant method, the ability of the drug to inhibit bacterial infection, environmental impact, safe handling or precautions related to use.
Page 140
BM3 User’s Manual liquid stick to the edge of the screen. Patient’s Cable Clean the patient cables with a gauze pad moistened with a soap solution. To disinfect patient cables, wipe the cables with a gauze moistened with diluted alcohol ...
Page 141
BM3 User’s Manual Wipe the sensor surface and sensor window with a damp cloth. Do not attempt to wet the sensor or disinfect it with hot water. Allow to dry completely with a lint-free cloth. Make sure the sensor window is clean and dry before use.
Page 142
BM3 User’s Manual Clean the exterior of the equipment at least once a month using a soft cloth moistened with lukewarm water or alcohol. Do not use lockers, thinners, ethylene, or oxidizers that could damage the equipment. Make sure that the cables and accessories are free from dust and dirt, then wipe them with a soft cloth moistened with 40 °...
The device is to be used by trained health care professionals. The monitor is intended for use in health care facilities; the BM3 Monitor is additionally intended for use in transport situations within the hospital setting.
Page 144
For example, a Bluetooth-compliant device using the 2.4 GHz frequency band may interfere with the wireless communication of the patient monitor. For more information on wireless deployment, please contact your Bionet representative. Low amplitude signals such as EEG and ECG are particularly sensitive to interference ...
BM3 User’s Manual Manufacturer’s declaration - electromagnetic emission The BM3 system is intended for use in the electromagnetic environment specified below. The customer or the user of BM3 system should assure that it is used in such an environment Emission test Compliance...
Manufacturer’s declaration - electromagnetic immunity The BM3 system is intended for use in the electromagnetic environment specified below. The customer or the user of the BM3 system should assure that it is used in such an environment Immunity test IEC 60601...
Page 147
0.5cycle rcial or hospital environmen Interruptions an for 0.5cycle t. If the user of the BM3 system requires continued o 40% Uт (60% dip in Uт ) Voltage variation peration during power main 40% Uт...
Page 148
BM3 User’s Manual The BM3 system is intended for use in the electromagnetic environment specified below. The customer or the user of the BM3 system should assure that it is used in such an environment Immunity test IEC 60601 Compliance level...
Page 149
BM3 User’s Manual Radiated RF 3 V/m 3 V/m Recommended separation distance 80.0 MHz to 2.5 G 80.0 MHz to 2.5 G IEC 61000-4-3 Where P is the maximum output power r ating of the transmitter in watts (W) acco...
Page 150
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the BM3 system. The BM3 system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The user of the BM3 system can help prevent electromagnetic interference...
Page 151
BM3 User’s Manual Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected b y absorption and reflection from structures, objects, and people.
Guidance and manufacturer’s declaration - electromagnetic immunity The BM3 system is intended for use in the electromagnetic environment specified below. The customer or the user of the BM3 system should assure that it is used in such an environment Immunity test...
Page 153
BM3 User’s Manual a- Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telepho nes and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
BM3 User’s Manual System Specification Physical Dimension (H x W x D) 250 x 238 x 163 mm Weight Approx. 3.1kg Indicator 3 LED Cooling Air flow Interface RJ45 , USB , HDMI Power AC 100-240V (50/60Hz) Adapter 18 V, 2.8 A Power consumption <...
Page 155
BM3 User’s Manual Indicator Categorized alarms (3 priority levels), Visual alarm lamp handle SpO2 pulse pitch tone, Battery status, External power LED Interface DC input connector : 18VDC, 2.8A Defibrillator Sync. Output : - Signal Level : 0 to 5V pulse - Pulse width : 100 ±...
Page 156
BM3 User’s Manual Lead type 3-lead, 5-lead(option) Lead Selection 3-lead : I, II, III 5-lead : I, II, III, aVR, aVL, aVF, V ECG waveforms 3-lead : 1 channel 5-lead : 1 channel Heart Rate Range Adult : 30 – 300 bpm Neonate/Pediatric : 30 –...
Page 157
BM3 User’s Manual Saturation accuracy 70 to 100% ±2 digits 0 to 69% unspecified Pulse rate range 30 to 254 bpm Pulse rate accuracy ±2 bpm NIBP Performance Method Oscillometry with linear deflation Operation Mode Manual/Automatic/Continuous Measurement range Adult Pressure : 20 to 260 mmHg...
Page 158
TB : 23 ~ 45℃ TI : 0 ~ 27℃ Alarm range 23 ~ 45℃ Product Configuration 1. Main body of BM3 Monitor 1 EA 2. 3-Lead patient Cable 3. Disposable electrodes 10 EA 4. NIBP extension horse 5. Reusable Adult NIBP Cuff 6.
Page 159
BM3 User’s Manual Option Product 1. Reusable Temperature Probe (Surface/Skin, TEMPSENS-430) 1 EA 2. Sidestream EtCO2 Module (Respironics) 1 SET 3. Mainstream EtCO2 Module (Respironics) 1 SET 4. Sidestream EtCO2 airway adapter sampling kit 1 EA 5. Mainstream EtCO2 airway adapter 1 EA 6.
BM3 User’s Manual Abbreviations and Symbols Abbreviations and symbols are alphabetized by reference, which can be read while reading the manual or using the equipment. Abbreviations amps alternating current adult ARRYTHM arrhythmia ASYS asystole Auto, AUTO automatic Auxiliary left foot augmented lead...
Page 165
BM3 User’s Manual diastolic direct current DEFIB, Defib defibrillator diastolic electrocardiograph electromagnetic compatibility electromagnetic interference electrosurgical cautery unit Fahrenheit gram heart rate, hour hertz intensive care unit incorporated --------------------------------------------------------------------------------------------------------------------------------------- Rev. 3.0...
Page 166
BM3 User’s Manual kg, KG kilogram kilopascal liter, left left arm, left atrial pounds liquid crystal display light emitting diode left leg M mean, minute meter MIN, minminute MM, mm millimeters MM/S millimeters per second MMHG, mmHg millimeters of mercury...
Page 167
BM3 User’s Manual operating room pediatric premature ventricular complex interval of ventricular depolarization right arm, right atrial RESP respiration right leg respiration rate systolic second SpO2 arterial oxygen saturation from pulse oximetry SYNC, Sync synchronization systolic Temp, TEMP temperature ---------------------------------------------------------------------------------------------------------------------------------------...
Page 168
BM3 User’s Manual precordial lead volt V-Fib, VFIB ventricular fibrillation VTAC ventricular tachycardia multiplier when used with a number (2X) Symbols & ° degree(s) > greater than < less than – minus number percent ± plus orminus --------------------------------------------------------------------------------------------------------------------------------------- Rev. 3.0...
Phone : Sales Agency Manufacturer * Thank you for purchasing BM3 * The product is manufactured and passed through strict quality control and through inspection. * Compensation standard concerning repair, replacement, refund of the product complies with “Consumer’s Protection Law” noticed by Korea Fair Trade Commission.
Need help?
Do you have a question about the BM3 and is the answer not in the manual?
Questions and answers
hello we have a BM 3 vet Bio Net , and we need it to print a darker ekg, it like we are running out of ink