BM5VET User’s Manual Table of Contents Table of Contents ....................... 1 1. BASIC ....................6 1.1 CE Standard Information ................... 7 1.2 Read before Use ......................8 How to Contact Us ............................ 8 Warranty Period ............................9 Warning, Caution, Note ..........................9 General Precaution on Environment .......................
Page 4
BM5VET User’s Manual 5. ECG ....................79 5.1 Introduction ....................... 80 Colors and Standards of Cables ....................... 80 Position of ECG Connector and Measuring Cable .................. 80 Skin Preparation and Lead Contact ......................81 Choosing an ECG lead for Arrhythmia Monitoring ................82 Information on the ECG waveform ......................
1. BASIC 1.1 CE Standard Information 1.2 Read before Use Warranty Period Warning, Caution, Note General Precaution on Environment General Precaution on Electric Safety Equipment Connection, Maintenance & Washing Equipment Connection Product Components Product Outline Principal Characteristics of Product Product Configuration and Option Product Product Body Configuration Function and Key External Function...
BM5VET User’s Manual 1.1 CE Standard Information Electromechanical safety standards met: - EN 60601-1: 1990 + A1:1993 + A2: 1995 + A13:1996 Medical Electrical Equipment, Part 1, General Requirements for Safety. - IEC/EN 60601-1-2:2001 Electromagnetic compatibility -Requirements and tests. - EN 1060-1:1995 Non-invasive sphygmomanometers - Part 1: General requirements...
※ If you need the supply circuit diagram, component list, description and calibration instruction etc. you can contact us we will provide you with it. The information in this manual only applies to BM5Vet monitor software version 1.06. Due to continuing product innovation, specifications in this manual are subject to change without notice.
“Consumer’s protection law” noticed by Korea Fair Trade Commission. We provide a 4-year warranty period.(Two years in Europe) We will repair or replace any part of the BM5VET found to be defective in usual operating circumstance for free to you.
BM5VET User’s Manual General Precaution on Environment - Do not keep or operate the equipment in the environment listed below. Avoid placing in an area exposed to moist. Avoid exposure to direct Do not touch the equipment sunlight with wet hand.
Page 12
BM5VET User’s Manual CAUTIONS Before Installation Compatibility is critical to safe and effective use of this device. Please contact your local sales or service representative prior to installation to verify equipment compatibility. Defibrillator Precaution Patient signal inputs labeled with the CF and BF symbols with paddles are protected against damage resulting from defibrillation voltages.
Page 13
BM5VET User’s Manual Electrocute Precautions To prevent skin burns, apply electrocute electrodes as far as possible from all other electrodes, a distance of at 15 cm/6 in. is recommended. Magnetic and electrical fields are capable of interfering with the proper performance of the device.
Page 14
MPSO as it will increase the chance of the single protective earth conductor interruption. Negligence BIONET does not assume responsibility for damage to the equipment caused by improperly vented cabinets, improper or faulty power, or insufficient wall strength to support equipment mounted on such walls.
Page 15
BM5VET User’s Manual NOTES Power Requirements Before connecting the device to the power line, check that the voltage and frequency. Ratings of the power line are the same as those indicated on the unit’s label. If this is not the case, do not connect the system to the power line until you adjust the unit to match the power source.
1. Be sure that AC power supply line is appropriate to use. (AC100 - 240V) 2. Be sure that the power source is the one supplied from Bionet. (DC18V, 2.5A, MW160KA1803) 3. Be sure that the entire connection cable of the system is properly and firmly fixed.
IEC 61000-3-3 Manufacturer’s declaration - electromagnetic immunity The BM5Vet is intended for use in the electromagnetic environment specified below. The customer or the user of the BM5Vet should assure that it is used in such an environment Immunity test IEC 60601...
Page 18
Note: Uт is the a.c. mains voltage prior to application of the test level. The BM5Vet is intended for use in the electromagnetic environment specified below. The customer or the user of the BM5Vet should assure that it is used in such an environment Immunity test...
Page 19
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the BM5Vet . The BM5Vet is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The user of the BM5Vet can help prevent electromagnetic interference by maintaining a...
Page 20
Guidance and manufacturer’s declaration - electromagnetic immunity The BM5Vet is intended for use in the electromagnetic environment specified below. The customer or the user of the BM5Vet should assure that it is used in such an environment Immunity test IEC 60601...
Page 21
BM5VET User’s Manual a- Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephone s and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be pr edicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF trans mitters, an electromagnetic site survey should be considered.
BIONET or its representatives. Maintenance and Washing Equipment Connection Using various methods can clean BM5VET and its accessories. Please follow the methods mentioned below to avoid unnecessary damage or contamination to the Equipment. We do not repair with free of charge regardless of warranty period if it is contaminated or damaged with using dangerous material not designated for washing.
Page 23
BM5VET User’s Manual Note The Equipment needs safety inspection once a year. Please refer to user’s guide or service manual for the examine objects. Please check carefully both frame and sensor, after cleaning the Equipment, Do not use the equipment that is worn out or damaged.
Page 24
Warning Check the electrodes of batteries before changing them. · Operate BM5VET with internal electric power supply when unsure of external ground connection or installation occur. · Remove the 1st Battery when not using equipment for a while without any damage.
BM5VET User’s Manual 1.3 Product Components Overview of the Product BM5VET monitor is a product used for monitoring biological information of Canine and Feline. Main functions of the product include displaying information such as ECG, respiration, SpO , NIBP, carbon dioxide(CO ), and temperature on its LCD screen and monitoring parameter, and alarming.
Warning In order to avoid electrical shock, do not open the cover. Disassembling of the equipment should be done only by the service personnel authorized by BIONET Warning Users must pay attention on connection any auxiliary device via LAN port or nurse calling.
Page 27
BM5VET User’s Manual Features of Main Body FRONT BACK ALARM LAMP RJ45 LAN Port VGA Output Protective Ground Terminal DC Power Input Port Wireless LAN POWER Rev. 2.53 2. ANIMAL/DATA MANAGEMENT 26...
Page 28
BM5VET User’s Manual Right Side CONNECTOR NIBP CONNECTOR SpO2 CONNECTOR TEMPERATURE CONNECTOR EtCO2 CONNECTOR CONNECTOR Left Side Rev. 2.53 2. ANIMAL/DATA MANAGEMENT 27...
Page 30
BM5VET User’s Manual Equipment Symbols ATTENTION : Consult accompanying documents Defibrillator-proof TYPE CF APPLIED PART : Insulated (floating) applied part suitable for intentional external and internal application to the patient including direct cardiac application. "Paddles" outside the box indicate the applied part is defibrillator proof.
Page 31
BM5VET User’s Manual External Ground PRINTER RS-232, Serial port LAN port AUX CONNECTOR DC INPUT INDICATOR BATTERY OPERATION INDICATOR DC INPUT CONNECTOR Rev. 2.53 2. ANIMAL/DATA MANAGEMENT 30...
Page 32
BM5VET User’s Manual NIBP Temperature Function Power on Power off Respiration Heart Pulse Rev. 2.53 2. ANIMAL/DATA MANAGEMENT 31...
BM5VET User’s Manual 1.4 Function and Key External Function The front panel of this product consists of an LCD screen and five function keys and one trim knob. Alarm Lamp Parameter window Silence Alarm Print Go/Stop NIBP Go/Stop Funtion Trim Knob...
Page 34
BM5VET User’s Manual 3. NIBP Go/Stop : This key is used for starting and stopping NIBP opration. When this key is pressed, the NIBP measurement is started. And if this key is pressed again during the measurement, the NIBP measurement is stopped, 4.
BM5VET User’s Manual 1.5 Standard Power Supply Application DC Power DC Power LED is lighted on when the DC Power is plugged into the inlet at the back of the product. A press of power key makes the machine ready for use.
BM5VET User’s Manual 1.6 Battery Power Supply Application Battery power can be supplied for enabling a portable use or a use during DC power failure. Operation 1. Battery Power LED is lighted on when the machine is in use. 2. The DC/battery power is only sustainable for 1 hour.
Page 37
5 minutes from that warning sign. In case of that warning sign with red and blink at icon box, charge the device immediately with DC power adaptor which is provided from BIONET. -Battery charging time: More than 6 hours...
BM5VET User’s Manual To insert and remove the battery pack. Assembly or replacement, as shown in the figure below. Rev. 2.53 2. ANIMAL/DATA MANAGEMENT 38...
BM5VET User’s Manual 1.7 General Manu Operation Screen Composition Real Time Wave Parameter Window Windows JOHN 10-JAN-07 12:23 PVC (/min)--, ST(mm)--.- ST(mm) mmHg MEDI 09:30 (93) 02:10 %SpO2 SpO2 RR II RR II (94) MAIN USER DISPLAY MENU SERVICE ℃...
BM5VET User’s Manual Menu Selection Turn or press the knob. When the Trim Knob Key is turned, menus are selected in the order indicated above. The above screen shows that the MORE menus is selected. The menus move to the right in the order of MORE MENU →...
Page 42
BM5VET User’s Manual Numerical value sign widow This window displays a measured parameter, function setup, and the boundary of parameter values. Breathing Alarm limit rate value Menu selection by using Trim Knob key As the key is turn to the right, the menu selection moves clockwise. As the key is turn to the left, the menu selection moves counterclockwise.
Page 43
BM5VET User’s Manual When moving the within quadrilateral, the letter reverses, and the numeric value reflects immediately. > MAIN VOLUME : MENU PREV 100% MENU Word feature menu The following figure shows the screen where the word sequence menu is activated within the word sequence correction menu.
Page 44
BM5VET User’s Manual Operation menu The setup value changes without a selection when the menu is moved. ADMIT CHANGE MAIN TYPE : ADMIT ADMIT MENU INFO HEIGHT WEIGHT PREV UNIT: UNIT: MENU MAIN ADMIT CHANGE ADMIT MENU TYPE ADMIT INFO...
BM5VET User’s Manual 2.1 ADMIT CHANGE ANIMAL INFO The CHANGE ADMIT INFO option allows you to change or enter information pertinent to the monitored animal. ANIMAL TYPE: You can select animal type as follow. HEIGHT UNIT: these options change the units of measure for height...
BM5VET User’s Manual CHANGE ANIMAL INFORMATION Hospital ID(11 letters for each), animal name (11 letters for each), sex (male or female), date of birth, weight, height, and animal ID (11 characters) CHANGE ANIMAL MAIN ANIMAL TYPE: MENU INFO HEIGHT WEIGHT...
BM5VET User’s Manual HEIGHT Unit of height is set as Cm / Inch. CHANGE ANIMAL MAIN ANIMAL TYPE: MENU INFO HEIGHT WEIGHT PREV DEFAULT UNIT: UNIT: MENU SETTING CHANGE ANIMAL MAIN ANIMAL TYPE: MENU INFO HEIGHT WEIGHT PREV DEFAULT UNIT:...
BM5VET User’s Manual 2.2 ALARM Alarm is divided into two, alarm for the patient’s condition and for the product’s condition. The patient’s alarm sounds when the diagnostic functions (ASYSTOLE, VTAC/VFIB, and VTAC) are detected. Each alarm sound differs in order and volume according to the levels of HIGH, MEDIUM, LOW and MESSAGE.
BM5VET User’s Manual Alarm for the Product The machine gives alarm sounds for its system with a related message flashing. ≡ ≡ Alarm Text ALARM LIMITS:The machine enables one to see and change the limits of alarm for all parameter functions.
BM5VET User’s Manual ALL LIMITS ALARM ALARM MAIN PRINT: VOLUME: MENU LIMITS NURSE PREV ALARM ALARM CALL: MENU LEVEL REVIEW ALARM LEVEL ON/OFF ALL LIMITS MEDIUM HIGH UNITS HIGH RETURN LIMITS PARA MEDIUM °F TEMP2 96.8 104.0 IBP1-S mmHg SPO2-%...
BM5VET User’s Manual ALARM VOLUME Set the alarm volume to be set at 10 grades. ALARM ALARM MAIN PRINT: VOLUME: MENU LIMITS NURSE PREV ALARM ALARM CALL: MENU LEVEL REVIEW ALARM > MAIN VOLUME: MENU PREV 100% MENU ALARM LEVEL Set the order of priority in each alarm.
BM5VET User’s Manual ARRHYTH LEVEL One can set up priorities when he or she uses the alarm for the diagnostic function. MAIN PARAMETER PARAMETER ARRHYTH MENU LEVEL LEVEL2 LEVEL PREV MENU ARRHYTH ALARM LEVELS RETURN ALARM LEVEL ASYSTOLE HIGH VTAC/VFIB...
BM5VET User’s Manual ALARM REVIEW After an alarm is triggered the alarms and data wave pattern can be reviewed. Set up for priority of each parameter alarm. ALARM ALARM MAIN PRINT: VOLUME: MENU LIMITS NURSE PREV ALARM ALARM CALL: MENU...
Page 56
BM5VET User’s Manual 10-MAR-2007 12:23 ST(mm): PVC (0/min): 0 mmHg mmHg II x2 SMALL 09:30 (93) 02:10 %SpO2 09:30 ALARM REVIEW (93) 2:10 RETURN TIME KIND %SpO2 2007/03/10 10:22:45 HIGH SPO2 2007/03/08 12:25:34 SpO2 RESP 2007/03/06 23:32:10 HIGH 2007/03/05 09:12:36...
BM5VET User’s Manual SAVE CONDITION This determines the order in which triggered alarms are saved. SAVE MAIN ALARM CONDITION : MENU LIST HIGH PREV MENU SAVE MAIN ALARM CONDITION : MESSAGE MENU LIST HIGH MEDIUM PREV > HIGH MENU Rev. 2.53...
BM5VET User’s Manual NURSE CALL When an alarm is triggered, this activated the NURSE CALL function. ALARM ALARM MAIN PRINT: VOLUME: MENU LIMITS PREV NURSE ALARM ALARM MENU CALL LEVEL REVIEW NURSE CALL SETUP RETURN CONTENTS NURSE CALL NORMAL MODE...
BM5VET User’s Manual 3.1 SETUP DISPLAY : screen set menu DEMO : Set ON/OFF DEMONSTRATION of equipment. KEY SOUND : Set ON/OFF Key sound of equipment. USER SERVICE : This is the menu to set the connection used to interface with an external computer MAKER SERVICE : This is the basic adjustment menu used to adjust the features of this product.
BM5VET User’s Manual SET PARA Select measurement function to use WAVE MAIN SELECT: MENU PARA DATE & TIME SWEEP PREV SPEED: SOURCE: MENU 25mm/s PARAMETER WINDOW SET RETURN WINDOW ON/OFF SPO2 RESP NIBP TEMP IBP I IBP II EtCO2 Rev. 2.53...
BM5VET User’s Manual WAVE SELECT Select waveform to display in large parameter display. WAVE MAIN SELECT: MENU PARA DATE & TIME SWEEP PREV SPEED: SOURCE: MENU 25mm/s WAVE MAIN SELECT: MENU PARA > SPO2 RESP SWEEP PREV SPEED: MENU 25mm/s...
Page 63
BM5VET User’s Manual * The large parameter display at the selection of RESP 10-JAN-2008 12:23 30/10 150/50 mmHg ℉ SMALL 100/90 %SpO2 36.5 42.0 30.0 (93) 36.5 09:58 42.0 30.0 0:56 PVC (/min):-, ST(mm): 0.0 II X1 (90) II X4 BED-001 Rev.
Page 64
BM5VET User’s Manual The large parameter display at the selection of EtCO2 WAVE MAIN SELECT: MENU PARA DATE & TIME SPO2 SWEEP PREV SPEED: SOURCE: MENU 25mm/s WAVE MAIN SELECT: MENU PARA SPO2 > SPO2 ETCO2 SWEEP PREV SPEED:...
BM5VET User’s Manual SET DATE & TIME It has sub menu to set date and time. WAVE MAIN SELECT: MENU PARA DATE & TIME SWEEP PREV SPEED: SOURCE: MENU 25mm/s SET TIME Set time of equipment. MAIN MENU TIME DATE...
BM5VET User’s Manual SET DATE Set date of equipment MAIN SET TIME MENU DATE PREV MENU MAIN MENU DATE: 06-MAR-2007 PREV MENU HR SOURCE This menu is used to set the source that detects heart and pulse rate. The source can select among ECG and SPO2.
BM5VET User’s Manual SWEEP SPEED Set speed of drawing wave signal pattern in this widow. WAVE MAIN SELECT: MENU PARA DATE & TIME SWEEP PREV SPEED: SOURCE: MENU 25mm/s SWEEP MAIN SPEED: > 6.25 mm/s MENU DATE & TIME 25mm/s 12.5 mm/s...
BM5VET User’s Manual USER SERVICE The user is able to set the communication parameters, power supply filter, and patient’s age. MAIN USER DISPLAY MENU SERVICE PREV DEMO : MAKER SOUND: MENU SERVICE SET BED MAIN UNIT NUMBER MENU NAME : 00A...
BM5VET User’s Manual SET BED NUMBER Set up for patient bed number. Allowable setters are from 0~ 9, A ~ Z . SET BED MAIN UNIT NUMBER MENU NAME : 00A PREV AC FILTER: W-LAN: SYSTEM MENU 50Hz SET BED...
BM5VET User’s Manual SYSTEM System able to change and verify Equipment version information and system information SYSTEM INFO SET RETURN CONTENTS MAIN VER 1.00.BVCDDCB EIA VER 1.01 NBP VER CENTRAL HOST IP 192 . 168 . 030 . 077 DEVICE IP 192 .
BM5VET User’s Manual 4.1 TREND TREND shows saved data graphically displayed with numeric values. Real-time data recording duration is 1 minute. Amount of saving time is for this data will be saving for 128hours. TREND MAIN GRAPHIC TABULAR WINDOW MENU...
BM5VET User’s Manual GRAPHIC TREND Wave Data can be stored and seen according to section. TREND MAIN GRAPHIC TABULAR WINDOW MENU TREND TREND SETUP PREV MENU John 10-JAN-2007 12:23 PVC(/min):--, ST(mm): mmHg GRAPHIC TREND 10-SEP-2007 MEDI HR(66 ST(12.4 S-PR(12 09:30...
BM5VET User’s Manual TIME PERIOD One can set up and store data and time that one can see in a screen. John 10-JAN-2007 12:23 PVC(/min):--, ST(mm): mmHg GRAPHIC TREND 10-SEP-2007 MEDI HR(66 ST(12.4 S-PR(12 09:30 (93) 300BPM 20mm 80RPM 02:10...
Page 75
BM5VET User’s Manual TABULAR TREND One can see the stored data at the time previously set up. TREND MAIN GRAPHIC TABULAR WINDOW MENU TREND TREND SETUP PREV MENU John 10-SEP-2007 13:30 PVC(/min):--, ST(mm): mmHg mmHg MEDI 09:30 TABULAR TREND 10-SEP-2007...
BM5VET User’s Manual TIME INTERVAL One can store data and set up time. John 10-SEP-2007 13:30 PVC(/min):--, ST(mm): mmHg mmHg MEDI 09:30 TABULAR TREND 10-SEP-2007 13:00 (93) 02:10 %SpO2 10-SEP 10-SEP 10-SEP 10-SEP 10-SEP %SpO2 09:30 12:10 12:09 12:08 12:07...
BM5VET User’s Manual TREND WINDOW SETUP Set the trend display window that will show the real time wave window. TREND MAIN GRAPHIC TABULAR WINDOW MENU TREND TREND SETUP PREV MENU 10-JAN-2007 12:23 ST(mm): PVC (0/min): 0 II x2 mmHg mmHg ST/PVC 20.0...
BM5VET User’s Manual TIME PERIOD Set visible time period in a screen. TIME MAIN PERIOD: TREND MENU 30MINS PARA PREV MENU TIME > 30MIN. MAIN PERIOD: 60MIN. MENU 30MINS 90MIN. PREV MENU 12H. Rev. 2.53 4.TREND 77...
BM5VET User’s Manual SET TREND PARA Set parameter for display in a screen. TIME MAIN PERIOD: TREND MENU 30MINS PARA PREV MENU PARAMETER WINDOW SET RETURN ON / OFF SPO2 RESP NIBP TEMP IBP1 IBP2 EtCO2 TREND PRINT Graphic: select the number which selects a graphic trend and press print to prints the selected trend.
5. ECG 5.1 Outline Color and Name for Each Cable Size ECG Connector Location and Measurement Cable 5 Lead Electrode Attached Location 3 Lead Electrode Attached Location Method to Attach Electrode to Baby 5.2 ECG Data Window 5.3 ECG Data Setup TRACE 1 LEAD SELECT ALARM LIMIT ALARM...
BM5VET User’s Manual 5.1 Introduction It calculates the heart rate with 3 or 5 leads ECG signal acquisition and perform the alarm according to the setting value. Colors and Standards of Cables Leadwire Color code Label Color code Label Right Foreleg...
BM5VET User’s Manual Skin Preparation and Lead Contact Sites where leads are attached to the animal must be properly prepared to optimize contact. Dogs and cats have enough electrolyte material on their skin and hair so that merely moistening lead sites with 70% isopropyl alcohol is appropriate.
BM5VET User’s Manual Choosing an ECG lead for Arrhythmia Monitoring It is very important to select a suitable lead for arrhythmia monitoring. Guidelines for non-paced animals: QRS should be tall and narrow(recommended amplitude > 0.5mV) R wave should be above or below the baseline (but not bi-phasic) ...
BM5VET User’s Manual 5.2 ECG Data Window Pace Detector QRS: Detects QRS, and Indicators: flashes when QRS is Detects and displays detected. the patient’s pace maker and flashes. Heart Rate Alarm Limit: Decides the QRS limits, and Heart Rate: gives an alarm if a value is over Displays heart rate per minute.
BM5VET User’s Manual Safety Precautions Warning CABLES — Route all cables away from patient's throat to avoid possible strangulation. CONDUCTIVE CONNECTIONS — Extreme care must be exercised when applying medical electrical equipment. Many parts of the human/machine circuit are conductive, such as the patient, connectors, electrodes, transducers.
Page 88
BM5VET User’s Manual INTERFACING OTHER EQUIPMENT — Devices may only be interconnected with each other or to parts of the system when it has been determined by qualified biomedical engineering personnel that there is no danger to the patient, the operator, or the environment as a result. In...
BM5VET User’s Manual 5.3 ECG Data Setup A setup window appears at lower part of the screen when the Trim Knob Key is pressed in the ECG Parameter Window. Selection is made by pressing the Trim Knob Key, while movement across the menu is performed by turning the key either clock or anticlockwise.
Page 90
BM5VET User’s Manual LEAD 1 SELECT MENU MAIN TRACE I : TRACE II : TRACE III : MENU NONE PREV MENU LEAD MAIN SELECT : > MENU PREV MENU LEAD 2 SELECT MENU MAIN TRACE I : TRACE II :...
Page 91
BM5VET User’s Manual LEAD 3 SELECT MENU MAIN TRACE I : TRACE II : TRACE III : MENU PREV MENU LEAD MAIN SELECT : MENU > PREV MENU 3 traces are displayed at once in case of connecting 5-Leads ECG cable and IBP para OFF 3 difference traces are displayed in case of selecting all 3 traces in LEAD SELECT menu.
Page 92
BM5VET User’s Manual Two different traces are displayed in case of selecting only 2 traces in LEAD SELECT menu. JOHN 10-JAN-07 12:23 PVC (/min):--, ST(mm):--.- mmHg MEDI 09:30 (93) 02:10 %SpO2 SpO2 RR II (94) ℃ 36.7 36.7 BED-001 The two traces of one lead are displayed in case of selecting only one lead in LEAD SELECT menu.
BM5VET User’s Manual ALARM LIMIT Alarm Limit is 0 ~ 350. LEAD MAIN ALARM SELECT : MENU ANALYSIS DISPLAY VOLUME : SETTING MAIN ALARM ALARM MENU LIMIT SOUND PREV MENU ECG ALARM LIMIT RETURN UNITS HIGH Rev. 2.53 5.ECG...
BM5VET User’s Manual QRS VOLUME Move the Key to select a volume rate from OFF, 10% to 100%. MAIN LEAD ALARM MENU SELECT ANALYSIS DISPLAY VOLUME : SETTING > MAIN VOLUME : MENU PREV 100% MENU DISPLAY Set the sweep speed and waveform size.
BM5VET User’s Manual ANALYSIS SETTING Analysis setting is divided to 3 menus. ECG FILTER:One may select from three frequency types for WAVE FILTER. MONITOR 0.5Hz ~ 40Hz MODERATE 0.5Hz ~ 25Hz MAXIMUM 5Hz ~ 25Hz DIAGONOSIS 0.05Hz ~ 120Hz MAIN...
Page 100
BM5VET User’s Manual PACE:Sets up ON/OFF to indicate that the patient has PACE. The PACE menu option enables/disables the pacemaker detection program. PACE- MAIN ARRHYTHM : FILTER : MAKER : MENU MONITOR PREV PVC SETTING ST SETTING MENU Be aware of the following when monitoring a patient with a pacemaker.
Page 101
BM5VET User’s Manual ARRHYTH: Sets up ON/OFF to indicate detection of diagnosis (Asys, VTAC/VFIB and VTAC). OFF: Do not perform arrhythmia diagnosis. LETHAL: Performs the detection of Asys, VTAC/VFIB, and VTAC at the selected lead FULL: Performs the detection of all 13 arrhythmia.
Page 102
BM5VET User’s Manual ACC VENT Over 14 years : Accelerated ventricular occurs when six or more ventricular beats are detected with an average heart rate for the ventricular beat between 50 and 100 beats per minute. 0-2 years : Occurs when six or more ventricular beats are detected with an average heart rate for the ventricular beat between 60 and 160 beats per minute.
Page 103
BM5VET User’s Manual COUPLET : Occurs when two ventricular beats are detected and have non-ventricular beats before and after the couplet. The coupling interval must be less than 600 milliseconds. IRREGULAR : Occurs when six consecutive normal R-to-R intervals vary by 100 milliseconds or more.
Page 104
BM5VET User’s Manual ST SETTING : ST signal and setting related ST menu. PACE- MAIN ARRHYTHM : MAKER : FILTER : MENU MONITOR PREV PVC SETTING ST SETTING MENU ST ANALYSIS: ON/OFF ST analysis signal. MAIN MEASUREMENT ANALYSIS : MENU...
Page 105
BM5VET User’s Manual ST ALARM LIMIT: ST alarm limit range setting MAIN MEASUREMENT ANALYSIS : MENU CONDITION ALARM LIMIT TEMPLETE PREV SELECT: MENU ALARM LEVEL ST ALARM LIMIT RETURN UNITS HIGH -4.0 Rev. 2.53 5.ECG 104...
Page 106
BM5VET User’s Manual ST ALARM LEVEL: ALARM LEVEL setting MAIN MEASUREMENT ANALYSIS : MENU CONDITION ALARM LIMIT TEMPLETE PREV SELECT: MENU ALARM LEVEL ST ALARM LEVEL > RETURN ST ALARM LEVEL MEDIUM Rev. 2.53 5.ECG 105...
Page 107
BM5VET User’s Manual TEMPLETE SELECT: Select a Representative Lead of ST LEVEL. The trace of the selected LEAD shows up at ST Window of POPUP TREND WINDOW MAIN MEASUREMENT ANALYSIS : MENU CONDITION ALARM LIMIT TEMPLETE PREV SELECT: MENU ALARM LEVEL...
Page 108
BM5VET User’s Manual PVC ALARM LIMIT: Set alarm indicate to PVC MAIN ANALYSIS : MENU ALARM LIMIT PREV MENU ALARM LEVEL PVC ALARM LIMIT RETURN UNITS HIGH /min Rev. 2.53 5.ECG 107...
Page 109
BM5VET User’s Manual PVC ALARM LEVEL: Set PVC ALARM LEVEL MAIN ANALYSIS : MENU ALARM LIMIT PREV MENU ALARM LEVEL PVC ALARM LEVEL > RETURN PVC ALARM LEVEL MEDIUM Warning Display Hart Beat Equipment Signal Hart Beat equipment signal displays when the PACE mode is. the signal appears series form.
Page 110
CAUTION FDA POSTMARKET SAFETY ALERT The United States FDA Center for Device and Radiological Health issued a safety bulletin October 14, 1998. this bulletin states “that minute ventilation rate-adaptive implantable pacemakers can occasionally interact with certain cardiac monitoring and diagnostic programmed rate.”...
Page 111
BM5VET User’s Manual WARNINGS VENTRICULAR ARRHYTHMISAS The arrhythmia analysis program is intended to detect ventricular arrhythmias. It is not designed to detect a trial or supra ventricular arrhythmias. Occasionally it may incorrect identify the presence or absence of an arrhythmia. Therefore, a physician must analyze the arrhythmia information in conjunction with other clinical findings.
Page 112
BM5VET User’s Manual Problem : False ventricular calls. Solution : Check ECG signal from patient: (the chest lead may exhibit polarity changes which may occasionally cause an inaccurate call.) 1. Check/adjust lead placement. 2. Check/perform skin preparation. 3. Check/replace electrodes. (if chest lead is a problem, move the chest lead to another chest position or leg position.)
BM5VET User’s Manual 6.1 Outline SPO2 monitoring is a noninvasive technique used to measure the amount of oxygenated hemoglobin and pulse rate by measuring the absorption of selected wavelengths of light. The light generated in the probe passes through the tissue and is converted into an electrical signal by the photodetector in the probe.
Page 115
When transflectance SpO2 probe is used, the preferred sensor application site for animals is in the rectal part or under the tail. WARNING Use only Bionet's reusable multisite SpO probe or transflectance SpO probe. Use of other oxygen transducers may cause improper performance.
BM5VET User’s Manual Signal and Data Validity It is extremely important to determine that the probe is attached to the patient correctly and the data is verifiable. To make this determination, three indications from the monitor are of assistance—signal strength bar, quality of the SPO2 waveform, and the stability of the SPO2 values. It is critical to observe all three indications simultaneously when ascertaining signal and data validity.
Page 118
BM5VET User’s Manual Stability of SPO2 Values The stability of the displayed SPO2 values can also be used as an indication of signal validity. Although stability is a relative term, with a small amount of practice one can get a good feeling for changes that are artifactual or physiological and the speed of each.
BM5VET User’s Manual 6.3 SpO2 Data Setup ALARM : Menu in which SpO alarm are set up. RATE VOLUME:Menu in which RATE VOLUME is set up RATE MAIN ALARM VOLUME: MENU RATE VOLUME Move the KEY to select the volume from OFF to 100%.
BM5VET User’s Manual ALARM LIMIT Number setting of alarm value of %SpO 0 ~ 100 2 is 1. Move the ▶ mark to select from RETURN, SpO or SpO -R, and press. 2. After pressing at SpO , move the cursor right or left to LOW, and press.
BM5VET User’s Manual ALARM SOUND Warning sound or message displays configuration menu when an alarm is triggered. ALARM MAIN ALARM SOUND: MENU LIMIT PREV MENU ALARM MAIN ALARM SOUND: MENU LIMIT PREV MENU LEAD FAULT Condition When using a reusable finger probe, there is a system alarm to alert you when the probe is off the Monitor.
BM5VET User’s Manual SPO2 Messages Below is a list of system status alarm messages which may be displayed in the SPO2 parameter window during monitoring. CHECK PROBE Reusable finger probe is off the patient. Check the probe. The factory default for this alarm is MESSAGE ALARM.
7.1 Outline Respiration via ECG Lead II electrode makes the skin area of the chest enlarged, causing changes in the resistance of skin. Through this it calculates respiration value per minute and performs the alarm function according to limit value. Respiration Connector and Measuring Cable Rev.
BM5VET User’s Manual 7.3 Respiration Data Setup ALARM: Respiration alarm setting menu RESP SIZE: A menu to setup Wave Display SWEEP SPEED: A menu to setup Wave Display of speed APNEA DETECT: A menu to setup APNEA alarm display SWEEP...
BM5VET User’s Manual RESPIRATION Set wave pattern size X2~ X10. SWEEP RESP MAIN ALARM SPEED : SIZE : MENU 12.5mm/s APNEA LEAD DETECT : SELECT: RESP MAIN > X 2 ALARM SIZE : MENU APNEA DETECT : APNEA DETECT Deciding function of activating Apnea Alarm...
BM5VET User’s Manual LEAD SELECT This is for changing the reference LEAD for respiration LEAD I or LEAD III can be selected. SWEEP RESP MAIN ALARM SPEED : SIZE : MENU 25mm/s APNEA LEAD DETECT : SELECT: SWEEP RESP MAIN...
BM5VET User’s Manual ALARM LIMIT Alarm Limit of Respiration Numeric Value is 5 ~ 150bpm Alarm Limit of RESPIRATION APNEA Numeric Value is 3 ~ 30sec. ALARM MAIN ALARM SOUND : MENU LIMIT PREV MENU 1. Move the ▶ mark to select RETURN, RESP or RESP-A, and press.
BM5VET User’s Manual ALARM SOUND Warning sound or message displays activation setting when Respiration ALRAM occurs. ALARM MAIN ALARM SOUND : MENU LIMIT PREV MENU ALARM MAIN ALARM SOUND : MENU LIMIT PREV MENU Rev. 2.53 7.RESPIRATION...
BM5VET User’s Manual 8.1 Outline This function is to measure minimum, Maximum and average blood pressure by using Oscillometric method Position of NIBP Connecter and cuff NIBP Connector LARGE CUFF Rev. 2.53 8. NIBP 132...
Page 134
BM5VET User’s Manual CAT CUFF Placement DOG CUFF Placement Note As the value of NIBP can vary according to the age and sex of a patient, the user needs to set up right data in Parameter Menu before measurement. WARNING Noninvasive blood pressure monitoring is not recommended for patients with hypotension, hypertension, arrhythmias or extremely high or low heart rate.
Page 135
BM5VET User’s Manual Note As the value of NIBP can vary according to the age and sex of a animal, the user needs to set up right data in parameter Menu before measurement. Tubes between the cuff and the monitor are not kinked or blocked.
BM5VET User’s Manual 8.2 NIBP Data Window Alarm Limit:Indicates alarm limit of blood pressure. Systolic pressure: Indicates the maximum limit of blood pressure mmHg display LARGE Diastolic blood 09:30 pressure:Indicates 1 HR (93) the minimum limit of 0:54 Measurement time...
BM5VET User’s Manual 8.3 NIBP Data Setup ALARM: A menu to set the Alarm CUFF SIZE:A menu to select cuff size UNIT SELECT: A menu to select the pressure unit INTERVAL :A menu to set Interval time when measures the blood pressure periodically...
BM5VET User’s Manual ALARM LIMIT Alarm setting Numeric Value of Systolic, Diastolic, and mean pressure is 10 ~ 360mmHg. 1. Move the ▶ mark to select one from RETURN, NIBP-S, NIBP-M, or NIBP-D, and press. 2. Press the key at NIBP-S, and move to LOW, and press again.(The user gets the same result regardless of the LOW-HIGH, or HIGH-LOW order.)
BM5VET User’s Manual ALARM SOUND The menu which decide activate of warning sign and message display when the respiration alarm is ALARM MAIN ALARM SOUND: MENU LIMIT PREV MENU CUFF SIZE The user can select a CUF between ADULT and NEONATAL.
BM5VET User’s Manual UNIT SELECT It is a function to set blood pressure measurement unit. The blood pressure measurement unit provides mmHg and kPa. NIBP CUFF MAIN STAT: ALARM SIZE: MENU MEDIUM UNIT INFLATION: INTERVAL: SELECT: 170mmHg mmHg NIBP CUFF...
Page 141
BM5VET User’s Manual INFLATION It is a function for pressurization pressure. HORSE/DOG/PUPPY/CAT : Numeric value is 80, 90, 100, 110, ~ 230, and 240. Numeric value is 60, 70, 80, 90, 100, 110, and 120. NIBP CUFF MAIN STAT: ALARM...
BM5VET User’s Manual 9.1 Description IBP has an alarming function based on the maximum & minimum alarming values configured by measuring the systolic, diastolic and mean blood pressure values with signal processing of electric signals which are transformed from changes in impedance components according to the changes of blood flow in vessels.
Page 144
BM5VET User’s Manual IBP ACCESSARY MEDEX Kit is used for IBP MONITORING KIT. LogiCal Disposable Pressure Transducers Cartridges and Monitoring kit Rev. 2.53 9. IBP...
9.2 IBP Data Window Different data windows are displayed on the screen according to the measuring positions. Measuring Position: Position of blood pressure measurement Systolic Blood Pressure: Indicating maximum blood pressure value Diastolic Blood Pressure: Indicating minimum blood pressure value Mean Blood Pressure: (94) (94)
9.3 IBP Data Setting Labels for measuring positions are described on each menu. CHANGE NAME: Menu to set measuring position SCALE: Menu to set size of measurement waveform on screen. LIMITS: Menu to set alarming range. SETTING: Menu for processing various pressure signals. ZERO: Menu to set zero-point of Transducer.
Page 148
List & Description of IBP Measurement Parameter Label Parameter Window, Scales Menu Window or Alarm Limits Pop-up Menu will appear according to the Labels. IBP displays the measuring positions based on 10 labels shown in the below table. The below table shows the names for each label and the descriptions to be displayed on the Parameter Window.
Page 149
BM5VET User’s Manual SCALE (Setting size of measurement waveform) You can set the pressure range for measurement waveform on this menu. The selectable values mean the maximum blood pressure range value that can be shown in a waveform. CHANGE MAIN...
Page 150
BM5VET User’s Manual Alarming Limits for ART Alarming limits vary according to measuring positions. The settable alarming range for systolic pressure, diastolic pressure and mean pressure is - 50 ~ 350mmHg. IBP ALARM LIMIT RETURN UNITS HIGH mmHg IBP1-S mmHg...
Page 151
BM5VET User’s Manual The below table shows the settable values of standard alarm limits and scales of parameters for label setting. Horse Puppy/Cat Parameter High Scale High Scale ART-S ART-D ART-M ART-PR FEM-S FEM-D FEM-M FEM-PR UAP-S UAP-D UAP-M UAP-PR...
Page 152
BM5VET User’s Manual IBP SETTING (Setting Various Functions) Other menus are to be applied for special functions to process pressure signals in various ways. CHANGE MAIN ALARM NAME SCALE: MENU LIMIT :CVP UNIT SETTINGS ZERO SELECT: :CVP mmHg Setting three labels of ART, FEM and UAP displaying PULSE-RATE among labels, the functions of PULSE-RATE DISPLAY and DISCONNECT ALARM will be added.
Page 153
BM5VET User’s Manual BP FILTER: It filters waveforms by selecting three frequency bands. 0Hz ~ 40Hz 12Hz 0Hz ~ 12Hz Generally recommended for monitoring 20Hz 0Hz ~ 20Hz Used for processing waveform components of higher frequency. Pressure value can be increased with this filter.
Page 154
BM5VET User’s Manual A function to adjust an error value based on the other index manometer. How to Adjust 1. Select a menu by pressing the knob switch key. 2. Measure blood pressure along with another index manometer. 3. Compare the measured values of ‘mmHg’ for both manometers.
Page 155
BM5VET User’s Manual Troubleshootings for a case the measured value is different from the expected value Description Action to Take In case there are air bubbles in tubes Remove the air bubbles In case an extension tube is connected Remove the extension tube...
Page 156
BM5VET User’s Manual Troubleshootings for a case that blood pressure value is not displayed on screen Description Action to Take In case of ‘out of measurement range’ Check the measurement conditions. situation In case blood pressure transducer is Replace the damaged transducer with new...
10.1 Introduction ETCO2(End-Tidal CO2) is a device to see the concentration of end-tidal carbon dioxide, which uses a method of measurement based on the non-dispersed IR absorption of CO2 using IR ray by sampling a certain part of respiration through pipe during respiration. EtCO2 connector position and accessory (Sidestream, Respironics) EtCO2 Connector LoFlo sidestream CO2 sensor and connector...
Page 159
BM5VET User’s Manual EtCO2 accessories for sidestream applications EtCO2 monitoring accessory uses the accessories for LoFlo™ sidestream module of Respironics Company. The airway adapters for sidestream intubated applications 3473ADU-00 Airway Adapter Dog/Horse Kit w/ Weight: 4.5 grams Deadspace – adds approximately 7...
Page 160
BM5VET User’s Manual Connecting the LoFlo Sample Kit 1. The sample cell of the sampling kit must be inserted into the sample cell receptacle of the LoFlo Module as shown in Figure 1. A “click” will be heard when the sample cell is properly inserted.
Page 161
BM5VET User’s Manual EtCO2 connector position and accessory (Mainstream, Respironics) EtCO2 Connector CAPNOSTAT 5 mainstream CO2 sensor and connector Mainstream sensor Mainstream sensor connector EtCO2 accessories for mainstream applications EtCO2 monitoring accessory uses the accessories for CapnoStat 5 microstream sensor of Respironics Company.
Page 162
BM5VET User’s Manual The airway adapters for mainstream intubated applications 6063-00 Single-Patient Use Horse and Dog Airway Adapter 6312-00 Single-Patient Use Cat and Puppy Airway Adapter 7007-00 Reusable Horse and Dog Airway Adapter 7053-00 Reusable Cat and Puppy Airway Adapter Rev.
Page 163
BM5VET User’s Manual Connecting the CAPNOSTAT® 5 CO2 Sensor to the Host System 1. Insert the CAPNOSTAT 5 CO Sensor connector into the receptacle of the host monitor as shown in Figure 1. 2. Make sure the arrows on the connector are at the top of the connector and line up the two keys of the connector with the receptacle and insert.
BM5VET User’s Manual 10.2 EtCO2 Parameter Window EtCO2 awRR Upper/lower limit value of FiCO2 alarm EtCO2 S: Display of apnea setting time in second unit Upper/lower limit value of alarm: Display of alarm setting range value for concentration of CO...
BM5VET User’s Manual 10.3 EtCO2 Parameter Setting Menu ALARM : A menu to set the alarm limit and sound. SWEEP SPEED: EtCO2 sweep speed is changeable to 6.25, 12.5, 25mm/s. WAVEFORM SCALE: A menu to set the screen scale of measured waveform.
Page 166
BM5VET User’s Manual EtCO2 ALARM LIMIT RETURN UNITS HIGH mmHg EtCO2 mmHg FiCO2 AWRR APNEA The following table shows standard alarm limit of parameter and setting value of scale when setting the label. Horse Puppy/Cat Parameter High Scale High Scale...
BM5VET User’s Manual EtCO2 SWEEP SPEED EtCO2 speed is 6.5mm/s. Speed is changeable to 6.25, 12.5, 25mm/s. SWEEP WAVEFORM MAIN ALARM SPEED: SCALE: MENU 6.25mm/s 40mmHg APNEA DETECT: SETTINGS ZERO SWEEP MAIN ALARM SPEED: MENU > 6.25mm/s 6.25mm/s 12.5mm/s APNEA...
BM5VET User’s Manual ZERO When a new sensor is connected, the message “ZERO REQUEMENT” is displayed. This is required anytime a new sensor is connected to the monitor. SWEEP WAVEFORM MAIN ALARM SPEED: SCALE: MENU 6.25mm/s 40mmHg APNEA DETECT: SETTINGS...
Page 169
BM5VET User’s Manual MODULE SETUP RETURN CONTENTS BAROMETRIC PRESSURE 760 mmHg 36.5 ℃ GAS TEMPERATURE NO BREATH DETECT TIMEOUT 0 SEC O2 COMPENSATION 21 % ANESTHETIC AGENT 0.0 % BALANCE GAS ROOM AIR CURRENT ETCO2 TIME PERIOD 0000-00-00-00 CURRENT CO2 UNIT...
Page 170
BM5VET User’s Manual SLEEP MODE: Sleep mode is used to save power when the host monitor is in standby mode. There are two sleep modes available for the Capnostat. Using Sleep Mode 1 maintains the heaters so the Capnostat is able to run immediately after exiting the sleep mode.
BM5VET User’s Manual MODULE INFO MAIN MODULE MODULE EtCO2 MENU SETUP INFO RESET PREV MENU MODULE INFORMATION SET RETURN CONTENTS SENSOR PN 1022054 OEM ID 0X01 SENSOR SN SN6412 HW REVISION NUM TOTAL USE TIME 148830 MIN. LAST ZERO TIME 0 MIN.
BM5VET User’s Manual EtCO2 RESET This command is used to cause a system watchdog reset in the sensor. When this command is issued, the system enters an infinite loop and a watchdog timer resets the system one second later. MAIN...
Page 173
Warning If defibrillation is performed while doing CO2 monitoring, remove the CO2 FilterLine from patient Getting in touch with sensor cable without removing the FilterLine can result in serious electrical burn, shock, or injury due to electric discharge energy. Note In the following monitoring conditions, the measured values may be inaccurate.
11. TEMPERATURE 11.1 Outline Temperature Connector and Measuring Cable 11.2 Temperature Data Window 11.3 Temperature Data Setup ALARM LIMIT UNIT SELECT Rev. 2.53 11.TEMPERATURE...
11.1 Outline This function is used to indicate the changes of resistance generated by the changes of temperature in numbers. The function involves the process of transferring the changes into electric signals. Temperature Connector and Measuring Cable Temperature Connector Temperature Measuring Cable Note Temperature probe is correctly positioned and fixed to do not disconnect on the patient.
BM5VET User’s Manual 11.2 Temperature Data Window Unit: Displays temperature unit. Temperature: Displays temperature. Alarm limit on the least low temperature : Indicates temperature limits Note The minimum measuring time required to obtain accurate readings at the specific body site is at least 3 minutes.
BM5VET User’s Manual 11.3 Temperature Data Setup ALARM: Temperature measurement alarm set UNIT: Temperature measurement unit set UNIT MAIN ALARM SELECT: MENU °C ALARM Alarm menu provide ALARM LIMIT and ALARM SOUND. ALARM MAIN ALARM SOUND : MENU LIMIT PREV MENU Rev.
BM5VET User’s Manual ALARM LIMIT Setting numeric value is 0.0℃ ~ 50.0℃. 1. Move the ▶ mark to select either RETURN or TEMP, and press. 2. After pressing the cursor at TEMP, move it to LOW, and press. 3. When the color has changed, move the cursor again to select a target value, and press.
BM5VET User’s Manual ALARM SOUND The menu which decide activate of warning sign and message display when the respiration alarm is ALARM MAIN ALARM SOUND : MENU LIMIT PREV MENU ALARM MAIN ALARM SOUND : MENU LIMIT PREV MENU UNIT SELECT Able to select unit with °C, °F.
Page 180
BM5VET User’s Manual Warning To measure the peripheral temperature, attach the probe to the ankle or palm. If the patient sweats heavily or moves violently, fasten the pad with surgical tape. NOTE When the measuring site is exposed directly to air, the temperature may be lower than normal.
12.1 Print Printer and Heat Sensitivity Paper A printer used to print data onto thermal paper. Size of the thermal paper roll: 58mm wide x 38mm in diameter any thermal paper of same size can be used for the printer. Side View of Printer Rev.
BM5VET User’s Manual Function and Setup Menu PRINTER WAVE MAIN SPEED: FORM1: MENU 25mm/s WAVE WAVE PREV FORM3: FORM2: MENU SPO2 RESP 1. Press the PRINT Key for continuous printing. 2. Select Printing Speed 25, 50 mm/s. PRINTER WAVE MAIN...
Page 184
BM5VET User’s Manual WAVE MAIN FORM1: > SPO2 MENU SPO2 RESP EtCO2 PREV IBP1 MENU IBP2 PRINTER WAVE MAIN SPEED: FORM1: MENU 50mm/s LEAD II WAVE WAVE PREV FORM3: FORM2: MENU SPO2 RESP WAVE MAIN FORM2: SPO2 MENU RESP >...
BM5VET User’s Manual 12.2 Paper Change Open the window of the printer. Insert the paper roll offered with the product into the printing unit. Place the roll in a proper way so that the printed paper can roll out upwards.
13. MESSAGE LIST Function Message Details LEAD FAULT Cable is not properly attached patient. CABLE OFF Cable is not properly connected. Cable is not properly connected. LEAD FAULT Patient’s finger is off the probe. CHECK PROBE Detection by the monitor of a pulse has ceased. PULSE SEARCH The SpO2 signal si too low.
15. TROUBLE SHOOTING 15.1 Noise in ECG - Gel is dry - Electrodes does not stick well to skin The patient’s skin Apply ECG gel or water to the area of is extremely dry contact then use the electrode Replace the electrodes The electrode is (Use CE compatible products) in poor condition...
BM5VET User’s Manual 15.2 SpO malfunction Connectors of the equipments are in bad condition? The extension cables are disconnected Replace extension cables The Finger probe is in bad condition? Replace the finger probe Repair the ECG B/D 15.3 Temp malfunction...
BM5VET User’s Manual 15.4 NIBP malfunction Are leaks from the hose connector of cuff Replace the hose of cuff Repair the NIBP B/D 15.5 Abnormality in NIBP measurements The patient moved Make sure the patient stay still while measuring during diagnosis...
BM5VET User’s Manual 15.6 Failure in battery recharge ( the battery does not fully recharge in 6 hours or more) Replace the adapter The output voltage of (the battery will not be changed at the rate lower than 17V) the adapter is 18V...
BM5VET User’s Manual 15.7 Power failure The adapter connector is in bad condition Replace the adapter The output voltage of the adapter is lower than 18V Replace the adapter Repair the power B/D Execute the “admit:” function “Admit” has been...
BM5VET User’s Manual 15.8 Periodic noises All alarm has been turned off Display lead fault Turn functions not in use off message appears (Mode-Display-Set para) The printer function abnormally Replace the printer Repair the accompanying Equipment Rev. 2.53 15. TROUBLE SHOOTING...
BM5VET User’s Manual 15.9 Print failure The paper tray cover is properly shut Shut the cover tight The printing face of paper has been reversed Reverse the paper face Repair the printer and printer B/D Rev. 2.53 15. TROUBLE SHOOTING...
16. SPECIFICATION Ease of use Customization Special Features Monitor Environmental Specifications Power adaptor Monitor Performance Specifications Graphical and Tabular Trends SpO2 Performance Specifications Respirations Performance Specifications NIBP Performance Specifications ECG Performance Specifications Temperature Unit Performance Specifications Accessories included OPTION Rev. 2.53 16.
BM5VET User’s Manual Ease of use · Battery operation · Attached printer · Table and graphic trend Additional Function · Able to use auto mobile power supply · LAN Connection Monitor Environmental Specifications · Operating Temperature: 15°C to 40°C (59°F to 104°F) ·...
101-150mmHg ±10% of reading Respiration rate 0 to 150 breath per minute ±1breath per minute Respiration accuracy Accessories Included: 1. Main body of BM5VET Monitor 1 EA 2. 3-Lead vet ECG Cable 1 EA 3. NIBP extension tube 1 EA 4.
Abbreviations and Symbols Abbreviations and symbols which you may encounter while reading this manual or using the monitor are listed below with their meanings. Abbreviations amps alternating current adult ARRYTHM arrhythmia ASYS asystole Auto, AUTO automatic Auxiliary left foot augmented lead left arm augmented lead right arm augmented lead beats per minute...
Page 208
BM5VET User’s Manual kg, KG kilogram kilopascal liter, left left arm, left atrial pounds liquid crystal display light emitting diode left leg M mean, minute meter MIN, min minute MM, mm millimeters MM/S millimeters per second MMHG, mmHg millimeters of mercury...
Page 209
BM5VET User’s Manual precordial lead volt V-Fib, VFIB ventricular fibrillation VTAC ventricular tachycardia multiplier when used with a number (2X) Symbols & ° degree(s) > greater than < less than – minus number percent ± plus or minus Rev. 2.53...
(2 years in Europe) Date of Purchase Hospital Name : Customer Address : Section Name : Phone : Sales Agency Manufacturer * Thank you for purchasing BM5VET. * The product is manufactured and passed through strict quality control and through inspection. Rev. 2.53...
Page 211
BM5VET User’s Manual International Sales & service Bionet Co., Ltd. 5F, Shinsegae I&C Digital Center 61 Digital-ro 31 gil, Guro-gu, SEOUL 08375, REPUBLIC OF KOREA Tel: +82-2-6300-6410 / Fax: +82-2-6300-7798 / e-mail: sales@ebionet.com Website: www.ebionet.com U.S.A sales & service representative Bionet America, Inc.