Customer Service: 904 208-2291 Toll Free 877 814-2237 EC|REP RMS UK, Ltd. 28 Trinity Road Nailsea, Somerset BS48 4NU United Kingdom TEL: 01275 858891 ® LIT034 CUDA SURGICAL XLS-180 Rev. O (English) Date of Revision: 05/14/14 Page 1 of 73...
Page 2
DISINFECTING LIGHT SOURCE MAINTENANCE, SERVICING & REPAIR, WARRANTY LAMP REPLACEMENT WARRANTY END OF PRODUCT LIFE TROUBLESHOOTING CHART OF MEDICAL DEVICE SYMBOLS CHART OF ELECTRICAL SYMBOLS USED ® LIT034 CUDA SURGICAL XLS-180 Rev. O (English) Date of Revision: 05/14/14 Page 2 of 73...
1. INTRODUCTION Congratulations on the purchase of your new XLS-180 Xenon Illuminator! This user-friendly Xenon Illuminator is a high efficiency light source utilizing state-of-the-art superior illumination technology. It offers a variety of features such as: 5600º K daylight brightness for perfect color definition •...
User is responsible for determining if interruption of light output will create an unacceptable risk. If this determination is made arrangements should be taken to reduce the risk. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (English) Date of Revision: 05/14/14...
Relative Humidity 0 to 95% Air Pressure 700 to 1060 hPa Dimensions 355(14)W x 130 (5)H x 250 (10)D mm Weight 9.5 lbs./4.3 kg ® LIT034 CUDA SURGICAL XLS-180 Rev. O (English) Date of Revision: 05/14/14 Page 5 of 73...
Accepts light source end fitting of light guide cable Intensity Control Mechanically controls the light output Power switch Turns the light source on and off; illuminates green when turned on. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (English) Date of Revision: 05/14/14 Page 6 of 73...
Allows adequate ventilation and cooling of the light source Fan 12V DC, airflow for cooling of unit Hour meter reset Resets hour meter to zero hours button Connector For potential equalization ® LIT034 CUDA SURGICAL XLS-180 Rev. O (English) Date of Revision: 05/14/14 Page 7 of 73...
Connect the light cable to endoscope and then plug the light cable endfitting into the appropriate port of the turret on the front panel. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (English) Date of Revision: 05/14/14 Page 8 of 73...
Such disinfectant agents usually arrive in the form of sprays or damp cloths. Follow the instructions given by the manufacturer of the disinfectant solution. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (English) Date of Revision: 05/14/14 Page 9 of 73...
Close the lamp compartment door. Re-connect the power cord and turn the light source on according to 6.1. Reset hour meter to zero by depressing hour meter reset button on rear panel. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (English) Date of Revision: 05/14/14 Page 10 of 73...
In the US a list of recyclers in your area can be found at: http:/www.eiae.org/. Please contact customer service to issue a return authorization to return product to manufacturer at the end of product life. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (English) Date of Revision: 05/14/14 Page 11 of 73...
Type BF Protective Earth (Ground) Product labeling Product label containing model number, serial number, electrical ratings, manufacturer name, Regulatory marks, and FDA “Rx Only” symbol. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (English) Date of Revision: 05/14/14 Page 12 of 73...
Page 13
Customer Service: 904 208-2291 Toll Free 877 814-2237 EC|REP RMS UK, Ltd. 28 Trinity Road Nailsea, Somerset BS48 4NU United Kingdom TEL: 01275 858891 ® LIT034 CUDA SURGICAL XLS-180 Rev. O (French) Date de révision: 05/14/14 Page 13 of 73...
Page 14
ENTRETIEN, MAINTENANCE ET RÉPARATIONS, GARANTIE REMPLACEMENT DE LA LAMPE GARANTIE FIN DE VIE DU PRODUCT DÉPANNAGE TABLEAU DES SYMBOLES DE DISPOSITIFS MEDICAUX UTILISES SYMBOLES ELECTRIQUES ® LIT034 CUDA SURGICAL XLS-180 Rev. O (French) Date de révision: 05/14/14 Page 14 of 73...
Page 15
1. INTRODUCTION Nous vous félicitons d’avoir acheté le nouvel illuminateur au xénon XLS-180. Cet illuminateur au xénon est simple à utiliser, il produit une source lumineuse d’une grande efficacité grâce à une technologie d’avant-garde et supérieure dans le domaine de l’illumination.
Page 16
à 12 pouces (30,5 cm). Si l'on utilise moins de 12 pouces du patient, l'intensité lumineuse doit être rejetée. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (French) Date de révision: 05/14/14...
-20 ° à +60 ° C (-4 ° à 140 ° F) Pression atmosphérique 0 à 95% 700 à 1060 hPa Dimensions 355(14)L x 130 (5)H x 250 (10)P mm Poids 9,5 lb/4,3 kg ® LIT034 CUDA SURGICAL XLS-180 Rev. O (French) Date de révision: 05/14/14 Page 17 of 73...
Page 18
Contrôle mécanique de la sortie lumineuse Interrupteur de tension Allume et éteint la source lumineuse ; le voyant est vert lorsque l’appareil est allumé. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (French) Date de révision: 05/14/14 Page 18 of 73...
Ventilateur 12 VDC, écoulement d’air pour le refroidissement de l’appareil Réglage du compteur Remet le compteur horaire à zéro horaire ’ Connecteur Pour l égalisation potentielle ® LIT034 CUDA SURGICAL XLS-180 Rev. O (French) Date de révision: 05/14/14 Page 19 of 73...
Brancher le câble de transfert de la lumière à l’endoscope puis brancher l’extrémité du câble de transfert de la lumière au réceptacle de la tourelle située sur le panneau avant. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (French) Date de révision: 05/14/14...
Page 21
électriques. Les produits désinfectants de ce type sont généralement fournis sous forme de bombes aérosol ou de chiffons humides. Observer les conditions d’utilisation indiquées par le fabricant du produit désinfectant. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (French) Date de révision: 05/14/14 Page 21 of 73...
Rebrancher le cordon d’alimentation et allumer la source lumineuse (cf. 6.1). Pour remettre le compteur horaire à zéro, appuyer sur le bouton de remise à zéro situé sur le panneau arrière ® LIT034 CUDA SURGICAL XLS-180 Rev. O (French) Date de révision: 05/14/14 Page 22 of 73...
S'il vous plaît le service clients de contact pour distribuer une autorisation de retour pour retourner le produit pour fabriquer à la fin de vie de produit. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (French) Date de révision: 05/14/14 Page 23 of 73...
électriques nominales, le nom du fabricant, les marques de réglementation, et le symbole « Rx Only » de la FDA (Usage limité aux praticiens agréés). ® LIT034 CUDA SURGICAL XLS-180 Rev. O (French) Date de révision: 05/14/14...
Page 25
Jacksonville, FL 32216 Customer Service: 904 208-2291 Toll Free 877 814-2237 EC|REP RMS UK, Ltd. 28 Trinity Road Nailsea, Somerset BS48 4NU United Kingdom TEL: 01275 858891 ® LIT034 CUDA SURGICAL XLS-180 Rev. O (German) Datum Auflage: 05/14/14 Page 25 of 73...
Page 26
WARTUNG, INSPEKTION UND REPARATUR UND GARANTIE ERSETZEN VON LAMPEN BESCHRÄNKTE GARANTIE ENDE DER BETRIEBSZEIT FEHLERSUCHE UND –BEHEBUNG TABELLE MIT SYMBOLEN, DIE AUF MEDIZINISCHEN GERÄTEN VERWENDET WERDEN ELEKTRISCHE SYMBOLE ® LIT034 CUDA SURGICAL XLS-180 Rev. O (German) Datum Auflage: 05/14/14 Page 26 of 73...
Page 27
1. EINFÜHRUNG Wir gratulieren Ihnen zum Erwerb ihres neuen XLS-180 Xenon Lichtgenerators. Dieser leicht bedienbare Xenon Lichtgenerator is eine hochleistungsfähige Lichtquelle, welche auf modernster hochentwickelter Lichttechnik beruht. Zu seinen vielfältigen Eigenschaften gehören: 5600° K Tageslichthelligkeit für perfekte Farbtrennschärfe • Geräuscharmer Betrieb •...
Page 28
Anschluss an die Lichtquelle sauber ist. Warnung Wenn er hohe Intensität Lichtquellen am vollen Ausgang verwendet, ist der empfohlene Abstand vom Scheinwerfermodul zum Patienten nicht kleiner als 12 Zoll (30.5 ® LIT034 CUDA SURGICAL XLS-180 Rev. O (German) Datum Auflage: 05/14/14 Page 28 of 73...
-20° C bis +60° C Luftfeuchtigkeit 0 bis 95% Luftdruck 700 bis 1060 hPa Abmessungen 355 (B) x 130 (H) x 250 (T) mm Gewicht 4,3 kg ® LIT034 CUDA SURGICAL XLS-180 Rev. O (German) Datum Auflage: 05/14/14 Page 29 of 73...
Zum Anschluss der entsprechenden Verbindung des Lichtleiterkabels an die Lichtquelle Intensitätsregelung Zur mechanischen Regelung des Lichtausgangs Anschaltknopf Zum Ein- und Ausschalten der Lichtquelle; leuchtet grün bei Einschaltung ® LIT034 CUDA SURGICAL XLS-180 Rev. O (German) Datum Auflage: 05/14/14 Page 30 of 73...
Page 31
Zur Belüftung und Kühlung der Lichtquelle Gebläse 12 V Wechselstrom Gebläse zur Kühlung des Geräts Betriebsstundenzähler- Zur Rücksetzung des Betriebsstundenzählers auf Null Rücksetztaste Verbinder Für potenziellen Ausgleich ® LIT034 CUDA SURGICAL XLS-180 Rev. O (German) Datum Auflage: 05/14/14 Page 31 of 73...
Radio- oder Fernsehgeräten in Wohnbereichen oder klinischen Einrichtungen verwendet wird. Zur Vermeidung elektromagnetischer VORSICHT Störungen darf das Gerät NICHT in der Nähe von HF-Geräten in Betrieb genommen werden. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (German) Datum Auflage: 05/14/14 Page 32 of 73...
Reinigungsmittel dürfen auf keinen Fall an den Reglerausgängen eindringen. Reinigungsmittel oder Flüssigkeiten dürfen nicht in die Ausgänge der Regeleinheit oder die Belüftungsschlitze seitlich, rückwärts oder unten an der Einheit gelangen. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (German) Datum Auflage: 05/14/14 Page 33 of 73...
Sie den Hauptschalter aus, bevor Sie eine neue Lampe einsetzen. VORSICHT Der Gasdichtungsrand der Lampe darf auf keinen Fall berührt werden, da dies zum Versagen der Lampe führen. VORSICHT ® LIT034 CUDA SURGICAL XLS-180 Rev. O (German) Datum Auflage: 05/14/14 Page 34 of 73...
Werk (Fracht bezahlt) übersenden. Ihr Produkt wird einer Inspektion unterzogen, worauf Sie einen Reparatur-Kostenangebot zu Ihrer Zustimmung erhalten. • In den USA: 877-814-2237 (gebührenfrei) • Kundendienst 904-208-2291 ® LIT034 CUDA SURGICAL XLS-180 Rev. O (German) Datum Auflage: 05/14/14 Page 35 of 73...
Achtung, in Begleitdokumenten nachlesen Achtung: Die nationale Gesetzgebung verbietet den Verkauf dieses Geräts von oder im Auftrag eines lizenzierten Angehörigen eines Heilberufs In Gebrauchsanweisung nachlesen CE-Kennzeichnung ® LIT034 CUDA SURGICAL XLS-180 Rev. O (German) Datum Auflage: 05/14/14 Page 36 of 73...
ELEKTRISCHE SYMBOLE Wechselstrom (AC) Potenzialgleichheit Einschalten Ausschalten Typ BF Schutzerde (Masse) Produktbeschriftung Produktetikett mit Modellnummer, Seriennummer, elektrischen Werten, Herstellername, Sicherheitsmarken und dem FDA "Rx Only" Symbol. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (German) Datum Auflage: 05/14/14 Page 37 of 73...
Page 38
Customer Service: 904 208-2291 Toll Free 877 814-2237 EC|REP RMS UK, Ltd. 28 Trinity Road Nailsea, Somerset BS48 4NU United Kingdom TEL: 01275 858891 ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Italian) Data di Revisione: 05/14/14 Page 38 of 73...
Page 39
MANUTENZIONE, ASSISTENZA TECNICA E RIPARAZIONI, GARANZIA SOSTITUZIONE DELLA LAMPADA GARANZIA FINE DELLA VITA DEL PRODOTTO SOLUZIONE DEI PROBLEMI TABELLA DEI SIMBOLI USATI SUI DISPOSITIVI MEDICI SIMBOLI ELETTRICI ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Italian) Data di Revisione: 05/14/14 Page 39 of 73...
Page 40
1. INTRODUZIONE Congratulazioni per l’acquisto del nuovo Illuminatore Xenon XLS-180! Si tratta di un dispositivo per illuminazione Xenon di semplice utilizzo, ad elevata efficienza, che utilizza una tecnologia di illuminazione all’avanguardia. Esso presenta tutta una serie di caratteristiche quali: Temperatura colore a 5600 ° K (luminosità della luce diurna) per una perfetta •...
Page 41
è meno di 12 pollici (30,5 cm). Se usando meno di 12 pollici dal paziente, l'intensità leggera deve essere rifiutata. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Italian) Data di Revisione: 05/14/14...
Umidità relativa 0% - 95% Pressione dell’aria da 700 a 1060 hPa Dimensioni 355 L x 130 H x 250 (P) mm Peso 4,3 kg ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Italian) Data di Revisione: 05/14/14 Page 42 of 73...
Permette la connessione del cavo a fibre ottiche Controllo intensità luce Controlla meccanicamente l’uscita della luce Interruttore Accende e spegne la sorgente luminosa; luce verde quando accesa ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Italian) Data di Revisione: 05/14/14 Page 43 of 73...
Ventola Ventola da 12 V CC per il raffreddamento dell’unità Pulsante per I’azzeramento Azzera il temporizzatore del temporizzatore Connettore Per l’ equalizzazione potenziale ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Italian) Data di Revisione: 05/14/14 Page 44 of 73...
Page 45
5.2 COLLEGAMENTO DEL CAVO A FIBRE OTTICHE Collegare il cavo a fibre ottiche all’endoscopio e, successivamente, al dispositivo, utilizzando il relativo attacco della torretta, ubicato sul pannello anteriore. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Italian) Data di Revisione: 05/14/14 Page 45 of 73...
Utilizzare disinfettanti comunemente usati per le superfici delle apparecchiature medicali. Generalmente si tratta di prodotti spray o di panni imbevuti di disinfettante. Attenersi alle istruzioni fornite dal produttore del disinfettante stesso. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Italian) Data di Revisione: 05/14/14 Page 46 of 73...
Richiudere lo sportello dell’alloggiamento. Ricollegare il cavo di alimentazione e accendere l’unità come indicato al punto 6.1. Azzerare il temporizzatore, premendo il pulsante per l’azzeramento, ubicato sul pannello posteriore. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Italian) Data di Revisione: 05/14/14 Page 47 of 73...
Etichettatura del prodotto Etichetta del prodotto che contiene il numero di modello, numero seriale, classe elettrica, nome del produttore, segni regolatori e simbolo FDA “Solo Rx”. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Italian) Data di Revisione: 05/14/14 Page 49 of 73...
Page 50
Customer Service: 904 208-2291 Toll Free 877 814-2237 EC|REP RMS UK, Ltd. 28 Trinity Road Nailsea, Somerset BS48 4NU United Kingdom TEL: 01275 858891 ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Spanish) Fecha de revision: 05/14/14 Page 50 of 73...
Page 51
MANTENIMIENTO, SERVICIO Y REPARACIÓN SUSTITUCIÓN DE LA BOMBILLA GARANTÍA LIMITADA FINAL DE LA VIDA PRODUCTIVA RESOLUCIÓN DE PROBLEMAS TABLA DE SÍMBOLOS USADOS EN DISPOSITIVOS MÉDICOS SÍMBOLOS ELÉCTRICOS ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Spanish) Fecha de revision: 05/14/14 Page 51 of 73...
Page 52
1. INTRODUCCIÓN Enhorabuena por la compra de su nueva Fuente de Luz de Xenón XLS-180! Esta fuente de luz de Xenón, de gran eficiencia y fácil uso, utiliza la tecnología de iluminación más moderna. Ello le permite ofrecer una amplia variedad de prestaciones, tales como: Brillante luz de día de 5600º...
Page 53
El usuario es responsable de determinar si la interrupción de la iluminación provocará un riesgo inaceptable. Si lo determina, deben tomarse medidas para reducir este riesgo. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Spanish) Fecha de revision: 05/14/14 Page 53 of 73...
0 a 95% Presión atmosférica 700 a 1060 hPa Dimensiones 355(14) ancho x 130 (5) alto x 250 (8) profundo mm Peso 9,5 lb./4,3 Kg. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Spanish) Fecha de revision: 05/14/14 Page 54 of 73...
Controla mecánicamente la intensidad de la luz. Interruptor de red Enciende y apaga la fuente de luz; se ilumina en verde cuando está encendida. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Spanish) Fecha de revision: 05/14/14 Page 55 of 73...
Ventilador de 12 V CC, para la refrigeración del equipo. Botón de reinicio del Repone el contador a cero contador de horas Conector Para iqualamiento potencial ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Spanish) Fecha de revision: 05/14/14 Page 56 of 73...
Page 57
TV, tenga en cuenta que éste puede verse afectado por interferencias radioeléctricas. Para evitar efectos electromagnéticos ATENCIÓN adversos, NO utilice el equipo cerca de dispositivos que produzcan emisiones de RF. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Spanish) Fecha de revision: 05/14/14 Page 57 of 73...
No deje que sustancias limpiadores o líquidos entren en las tomas de control de la unidad o los conductos de ventilación situados a los lados, la parte trasera o inferior de la unidad. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Spanish) Fecha de revision: 05/14/14...
ATENCIÓN No toque a la lámpara y a el paciente simultáneamente al cambiar la lámpara. ATENCIÓN ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Spanish) Fecha de revision: 05/14/14 Page 59 of 73...
Por favor servicio de atención al cliente de contacto para publicar una autorización del regreso para volver el producto a fabricar en el fin de la vida del producto. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Spanish) Fecha de revision: 05/14/14...
“Solo con receta” autorizado por Food and Drug Administration ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Spanish) Fecha de revision: 05/14/14 Page 61 of 73...
Page 62
Customer Service: 904 208-2291 Toll Free 877 814-2237 EC|REP RMS UK, Ltd. 28 Trinity Road Nailsea, Somerset BS48 4NU United Kingdom TEL: 01275 858891 ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Portuguese) Data da Revisão: 05/14/14 Page 62 of 73...
Page 63
MANUTENÇÃO, SERVIÇO E REPARAÇÃO SUBSTITUIÇÃO DA LÂMPADA GARANTIA FIM DA VIDA DO PRODUTO ESOLUÇÃO DE PROBLEMAS TABELA DE SÍMBOLOS DISPOSITIVOS MÉDICOS TABELA DE SÍMBOLOS ELÉCTRICOS USADOS ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Portuguese) Data da Revisão: 05/14/14 Page 63 of 73...
Page 64
1. INTRODUÇÃO Parabéns pela compra da sua nova Fonte de Luz Xénon XLS-180! Esta fonte de luz Xénon é muito eficiente e fácil de usar, utilizando a tecnologia de iluminação mais moderna. Oferece uma variedade de utilizações, tais como: Luminosidade diurna de 5600ºK para uma definição perfeita da cor •...
Page 65
O usuário é responsável por determinar se a interrupção da produção de luz vai criar um risco inaceitável. Se esta determinação é feita, os arranjos devem ser tomadas medidas para reduzir o risco. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Portuguese) Data da Revisão: 05/14/14 Page 65 of 73...
Page 66
0 a 95% Pressão atmosférica 700 a 1060 hPa 355 (14) comprimento x 130 (5) altura x 250 (8) Dimensões profundidade mm Peso 4,3 Kg ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Portuguese) Data da Revisão: 05/14/14 Page 66 of 73...
Page 67
Aceita diferentes terminais de cabos de fibras ópticas Controlo de intensidade Controla mecanicamente a intensidade da luz Interruptor Liga e desliga a fonte de luz; permanece verde quando ligada. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Portuguese) Data da Revisão: 05/14/14 Page 67 of 73...
Ventilador de 12V CC, para a refrigeração do aparelho. Botão de reinicio do Repõe o contador a zero. medidor de horas Conector Para equalização da potência. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Portuguese) Data da Revisão: 05/14/14 Page 68 of 73...
Page 69
Ligue o cabo de fibra óptica ao endoscópio e em seguida ligue o terminal do cabo à torre no painel frontal, no encaixe apropriado. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Portuguese) Data da Revisão: 05/14/14 Page 69 of 73...
Page 70
Tais desinfectantes apresentam-se normalmente em formato de pulverizadores ou em toalhetes húmidos. Siga as instruções do fabricante e da solução desinfectante. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Portuguese) Data da Revisão: 05/14/14 Page 70 of 73...
6.1. Pressione o botão do contador, que se encontra no painel posterior, para reiniciar a contagem. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Portuguese) Data da Revisão: 05/14/14 Page 71 of 73...
Por favor contacte o serviço de clientes, para solicitar ao fornecedor uma declaração de devolução do produto em fim de vida. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Portuguese) Data da Revisão: 05/14/14 Page 72 of 73...
Etiqueta do produto contendo número do modelo com número de série as especificações elétricas nome do fabricante marcos regulatórios e A FDA "Rx" símbolo. ® LIT034 CUDA SURGICAL XLS-180 Rev. O (Portuguese) Data da Revisão: 05/14/14 Page 73 of 73...