6.7.1
Data-Hold function
 Switch on the balance, keep the HOLD-button pressed until the current setting
„Hx" (H1 – H6) appears.
 Press ON/OFF button repeatedly until the desired setting „H1-H4" is displayed.
 Confirm setting with the HOLD button.
 Suspend the material to be weighed
 The weight value is fixed and displayed depending on the setting (H1 – H4) (see
chap. 5.7), represented by the [] in the upper left.
6.7.2
Animal weighing function
This function is suitable for busy weighing procedures. The result is a mean value
formed by 16 weighing values which is found out within 3 seconds.
 Switch on the balance, keep the HOLD-button pressed until the current setting
„Hx" (H1 – H6) appears.
 Press ON/OFF button repeatedly until the setting „H5" is displayed.
 Confirm setting with the HOLD button.
 Suspend the material to be weighed
 Press the HOLD-button, the display counts reverse from 3 -1.
The calculated mean value is indicated, symbolised by the [] top left.
 Before further measurements press first the ON/OFF/TARE-button.