Appendix D: Warming System Response by Temperature

Fluid
Heater
Temp
30 °C
Active
31 °C
Active
32 °C
Active
33 °C
Active
34 °C
Active
35 °C
Active
36 °C
Active
37 °C
Active
38 °C
Active
39°C
Active
40 °C
Active
41 °C
Off
42 °C
Off
43 °C
Off
44 °C
Off
45 °C
Off
46 °C
Off
47 °C
Off
48 °C
Off
49 °C
Off
50 °C
Off
enFlow Service Manual Rev. F 04/12
4400_0060
Temperature LED on
Warmer
Blue Flashing
Blue Flashing
Blue Flashing
Blue
Blue
Green
Green
Green
Green
Green
Green
Green
Green
Yellow
Yellow
Red Flashing High Priority
Alarm after 20 seconds
Red Flashing High Priority
Alarm after 16 seconds
Red Flashing High Priority
Alarm after 12 seconds
Red Flashing High Priority
Alarm after 8 seconds
Red Flashing High Priority
Alarm after 4 seconds
Red Flashing High Priority
Alarm (immediately)
Display on Controller
30°C Blue Flashing
31°C Blue Flashing
32°C Blue Flashing
33°C Blue
34°C Blue
35°C Green
36°C Green
37°C Green
38°C Green
39°C Green
40°C Green
41°C Green
42°C Green
43°C Yellow
44°C Yellow
45°C after 20 seconds
Red Flashing "Over
Temp" message
46°C after 16 seconds
Red Flashing "Over
Temp" message
47°C after 12 seconds
Red Flashing "Over
Temp" message
48°C after 8 seconds
Red Flashing "Over
Temp" message
49°C after 4 seconds
Red Flashing "Over
Temp" message
Red Flashing "Over
Temp" message
(immediately)
Page 26 of 36
Audible
Alarm
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No
After 20
seconds
After 16
seconds
After 12
seconds
After 8
seconds
After 4
seconds
Immediately