  • Page 1 PERMA PURE LLC Baldwin™-Series Model 3300/3300P Flow Control Drawer User’s Manual PERMA PURE LLC 8 Executive Drive ▪ P.O. Box 2105 ▪ Toms River, NJ 08754 (732) 244-0010 ▪ (800) 337-3762 ▪ fax (732) 244-8140 ▪ www.permapure.com ▪ info@permapure.com...
  • Page 2: Table Of Contents

    PERMA PURE LLC Model 3300 User’s Manual Doc. #532: Revision: 000 Page 2 of 14 Contents Introduction ........................3 General description ......................4 Physical description ......................5 Specifications ........................5 Connections........................5 Important Safety Warnings ....................6 Operation ......................... 7 Pressure Sensor Board Option ..................
  • Page 3: Introduction

    Model 3300 User’s Manual PERMA PURE LLC Doc. #532: Revision: 000 Page 3 of 14 Unpacking Perma Pure has made every effort to ship you a high quality product that has been thoroughly inspected and tested. It has been carefully packed to ensure that it arrives at your facility in good condition.
  • Page 4: General Description

    Model 3300 User’s Manual PERMA PURE LLC Doc. #532: Revision: 000 Page 4 of 14 General description The Baldwin™-Series Model 3300 Flow Control Drawer allows easy control of sample and calibration gases. With the Model 3300, you may choose whether to control sample gases, cal gases, and stack probe blowback, manually or by PLC.
  • Page 5: Physical Description

    Model 3300 User’s Manual PERMA PURE LLC Doc. #532: Revision: 000 Page 5 of 14 Physical description • Dimensions: 5U (8.75”), 19” Rack Mount, 18” depth • Weight: 24 lbs (10.9 kg) Specifications Input cal-gas channels (maximum of 14) Sample 5-15 psig Vacuum 0-30 in.
  • Page 6: Important Safety Warnings

    Model 3300 User’s Manual PERMA PURE LLC Doc. #532: Revision: 000 Page 6 of 14 Important Safety Warnings Please be sure to review the following basic safety procedures. These procedures represent the MINIMUM requirements to operate the equipment safely. It is the ultimate responsibility of the operator to ensure proper safety practices are utilized at the point of operation.
  • Page 7: Operation

    Model 3300 User’s Manual PERMA PURE LLC Doc. #532: Revision: 000 Page 7 of 14 Operation A sample conditioning system supplies a clean, dry sample to the back panel of the Model 3300 Flow Control Drawer. The stack probe filter and sample pump performance are indicated by the Flow Control Drawer Vacuum/Pressure Gauges.
  • Page 8: Local Mode Operation

    Model 3300 User’s Manual PERMA PURE LLC Doc. #532: Revision: 000 Page 8 of 14 Local Mode Operation To enter the Local mode, toggle the Remote/Local switch to the right. While in the Local mode, the operator can perform manual calibrations and blowback. The flow control drawer is in Local Control when the Local Switch is selected, the red LED is on and the green Sample LED is on as well.
  • Page 9: Computer Control Terminals Summary

    Model 3300 User’s Manual PERMA PURE LLC Doc. #532: Revision: 000 Page 9 of 14 Computer Control Terminals Summary Terminal 2 = System Calibration Terminal 1 = Cal Gas 1 Terminal 4 = Direct Calibration Terminal 3 = Cal Gas 2 Terminal 6 = Blowback Terminal...
  • Page 10: Computer Control Truth Table

    Model 3300 User’s Manual PERMA PURE LLC Doc. #532: Revision: 000 Page 10 of 14 COMPUTER CONTROL TRUTH TABLE Terminal # | 2 | 5 | 7 | 9 | 11 | 13 | 15 | FUNCTION System/Calibration 1 System/Calibration 2 System/Calibration 3 | * | System/Calibration 4...
  • Page 11: Troubleshooting And Maintenance Guide

    Model 3300 User’s Manual PERMA PURE LLC Doc. #532: Revision: 000 Page 11 of 14 Troubleshooting and Maintenance Guide Symptom Diagnostic No LED display Not plugged in Plug in instrument Power switch is off Turn on power Fuse blown Replace fuse Power supply defect Replace power supply 12 VDC red-black...
  • Page 12: Replacement Parts

    Model 3300 User’s Manual PERMA PURE LLC Doc. #532: Revision: 000 Page 12 of 14 Replacement parts 4D-3300 Part No. Description 2FTS-021 Cal-Gas connector 3DCB-006 Circuit board: switch board, FCD 2FIV-003 Flow meter: 0.1 - 10 LPM w/valve 2FIN-002 Flow meter: 0.1 - 10 LPM no valve 2FIV-005 Flow meter: 0.2 - 2.5 LPM w/valve 1PSD-032...
  • Page 13: Appendix A: Flow Control Drawer Drawings

    Model 3300 User’s Manual PERMA PURE LLC Doc. #532: Revision: 000 Page 13 of 14 Appendix A: Flow Control Drawer Drawings Panel drawing Flow schematic COPYRIGHT 2008, ALL RIGHTS RESERVED, Original Instructions...
  • Page 14: Appendix B: Warranty And Disclaimer

    Model 3300 User’s Manual PERMA PURE LLC Doc. #532: Revision: 000 Page 14 of 14 Appendix B: Warranty and Disclaimer Perma Pure LLC Perma Pure (Seller) warrants that product supplied hereunder shall, at the time of delivery to Buyer, conform to the published specifications of Seller and be free from defects in material and workmanship under normal use and service.
  • Page 15 PPENDIX RESSURE WITCH ODIFICATION The pressure switch option provides: • Vacuum pressure switch 1.6-27 inHg Vacuum • Sample pressure sensor 2-48 psig (50 psig Max) • Calibration gas pressure sensor 2-38.3 psig (50 psig Max) Outputs 1. The standard output from all switches is Form C contact, 1 Amp 2.