NEO Photoionization Detector
User Manual
D e t e c t i o n
D e t e c t i o n
1. Briefly block the inlet probe with your finger.
2. Observe the rise in pump current value.
3. Set the stall threshold about 50-100 units above the unblocked pump reading, but lower than the
blocked pump reading.
7.5.5 ROLLING GRAPH
If the Rolling Graph function is enabled, the display in basic User Mode shows the
instantaneous
reading together with a real-time plot of readings in the last 2 minutes.
To set Rolling Graph:
OK
Ent
Esc
1. Press
or
to select the desired option.
OK
Ent
Esc
OK
Ent
Esc
2. Press
to continue.
3. Press
to save and exit
OK
Ent
Esc
7.5.7 REAL TIME DATA
Real time data output is available on some NEO models via a USB cable and/or wirelessly via Bluetooth
Low Energy using an Android App. Please contact WatchGas for more information.