7.3.4 MAN-DOWN ALARM
The Man-Down Alarm, when enabled, warns nearby personnel if the instrument remains motionless for
a pre-set time. If the instrument is fitted with the wireless modem option and connected to a network, it
sends alarm notifications to the command centre, enabling a rescue operation.
You can change various properties of the man-down alarm:
You can turn the man-down alarm on or off.
You can set the motionless time after which the instrument warns that it is
going to activate the man-down alarms.
You can set the warning time. During the warning time, the user can prevent
the alarms from activating.
You can set the sensitivity. High sensitivity means small movements are
enough to prevent the man-down alarm from activating.

7.4 DATALOG

The instrument automatically stores the detected concentrations at a pre-set time
interval. In the readout screen, the datalog icon shows that data is logged.
Data is stored in non-volatile memory for download to a PC. The data is retained,
even when the instrument is turned off or the battery is removed. Data is
organised in events, with a new event created each time the instrument is turned
on, a configuration parameter is changed or datalogging is interrupted. Every
event includes User ID, Site ID, serial number, last calibration date and the alarm
limits.
Notes
You cannot turn off the datalog function.
By default, the instrument stores the detected concentration every 60 seconds.
In the datalog submenu, you can adjust the interval and clear all logged data.

7.4.1 CLEAR DATALOG

To erase all data stored in the datalog: