3
Insert the cassette containing the emergency image plate into the cassette slot
as shown below.
Make sure to insert the cassette with the hinge [1] at the top and the locking
mechanism [2] at the bottom.
The image plate will be digitized using the speed class, i.e. the sensitivity,
corresponding to the emergency button as defined during configuration.
When the Centricity™ SP1001 has treated the emergency image plate, it displays
the operator main screen. The
processing station accompanied by default ID data. The image processing station will
give the emergency image priority over the other images in the image processing
queue.
If you do not enter a cassette within 15 seconds after pressing the
emergency button, the Centricity™ SP1001 will quit the emergency button
function and return to the operator main screen.
To change the speed class corresponding to the emergency button, contact
your local service organization.
4
Remove the cassette from the cassette slot.