Emergency buttons
Two emergency buttons are located at the front of the Centricity™ SP1001. The
emergency buttons determine the speed class, i.e. the sensitivity, which will be
used to digitize the image plate. The sensitivity associated with the emergency
buttons has been set during the configuration of your system. For more
information, contact your local service organization.
The emergency buttons have the following labels:
For digitizing images of the trunk.
For digitizing images of the limbs.
In the interchangeable configuration, the emergency buttons permit you to
digitize emergency image plates without ID data.
In the dedicated configuration, the emergency buttons permit you to digitize
an image plate while you are entering the identification data via the ID Station.
Audio signals
The Centricity™ SP1001 gives status information via beeps. The length of the
beep indicates the response of the system to a key command.
• A short beep means that Centricity™ SP1001 has accepted the key command
and is starting the operation.
• A long beep means that you have pressed a non-active key or that the
Centricity™ SP1001 has rejected the key command.
• An interval beep accompanies an error, locked or warning message. Refer to
'Troubleshooting checklist'