Flow Phase; Figure 20: Freeflow Record Phase - LABORIE NXT Go Owner's Manual

Hide thumbs Also See for NXT Go:
Table of Contents

Advertisement

Click the Zero Volume button to prepare the Urocap NXT and then click the Start
Flow Phase button to start the study and move to the Flow phase.
NOTE: Pause the study by clicking the Pause
by clicking the resume
3.2.2.2

Flow Phase

Instruct the patient to void. The Synergy application will automatically start to
record data when the transducer detects the start of the flow.
NOTE: To shift the position or empty the urine beaker during the Flow Phase, click the
Change Urine Beaker button. A popup window will appear, and the study will pause.
Once the action is complete, click Continue on the popup window to resume.
The flow phase will stop 60 seconds after voiding ends.
A Post Voided Residual window (Figure 21) will pop-up. Select the preferred
method of capturing Post Voided Residual (PVR), meaning the amount of fluid
remaining in the bladder at the completion of the voiding phase. Use of the PVR
functions is at the discretion of the overseeing physician.
a. Calculated PVR selection automatically provides the difference between
infused volume and voided volume.
b. Drainage PVR selection allows the operator to pour drained fluid into the
Urocap NXT to measure PVR.
IMPORTANT: Do not empty the Urocap NXT beaker before adding PVR fluid
drained from patient bladder. If the beaker is filled to capacity utilize Calculated
PVR or Input PVR options.
c. Input PVR selection allows the operator to input the PVR measurement
calculated by a separate measurement device.
NXT Go Owner's Manual, LBL-00145, MAN2030
button.

Figure 20: FreeFlow Record Phase

How to Perform a Study
button and resume the study
33