Calibration — Type Z
connectors. Make certain the VAR. ATTEN. control is set
fully clockwise.
et the GAIN ADJU T control (R8639) to
obtain exactly 4 cm of vertical deflection. Place the Mode
switch to — B ONLY. Check to see that there are exactly 4
cm of vertical deflection.
8. Operational Checks
Perform the following Z Unit operational checks:
(a) Disconnect the calibrator signal from the input con
nectors.
Check to see that the baseline does not shift
more than 2 mm after switching the Mode switch from
A ONLY to — B ONLY.
Reconnect the calibrator signal
to the input connectors.
(b) With the Mode switch at — B ONLY, check the opera
tion of the B AC-DC switch by setting it to the AC posi
tion. Observe the signal to see that it shifts to its average
voltage level. The shifting of the signal indicates that r he
AC-DC switch and the blocking capacitor function properly.
et the Mode switch to A ONLY and check the A AC-DC
switch for the same waveform shift indication. Place both
AC-DC switches back to the DC position.
(c) Check operation of the VAR. ATTEN. control. When
the control is rotated fully counterclockwise, the vertical de
flection of the signal should not exceed 1.6 cm. Rotate the
control back to the fully clockwise position.
9. Turret Attenuation Ratios
With the Mode switch at A ONLY, check for 2 cm of
vertical deflection at each position of the A VOLT /CM
switch by following the information in Table 5-2.
TABLE 5-2
VOLT /CM
witch
etting
.05
.1
.25
.5
1
2.5
5
10
25
After checking the A VOLT /CM turret attenuator, set
the Mode switch to — B ONLY and check the B VOLT /CM
attenuator in the same manner.
10.
Differential Balance
et the Mode switch to the A-B position and both
VOLT /CM switches to .05.
a 100-volt output signal.
controls for a stable display. Ignoring the spikes on the
waveform, adjust the DIFF. BAL. control for minimum square
wave amplitude.
The maximum allowable amplitude is
0.5 mm. When the control is correctly adjusted and the Z
Unit balances properly, the waveform appears as a straight
line with positive and negative spikes.
5-4
Calibrator Output
Volts
.1
.2
.5
1
2
5
10
20
50
et the calibrator switch lor
et the oscilloscope triggering
An ac mismatch between two tubes or components can
cause the straight line segments of the waveform to tilt
diagonally to form a zigzag pattern. The amplitude of the
tilt at its highest point should not exceed 1 mm. To check
the amplitude, free run the sweep at 20 millisec/cm and
observe the width of the trace. It should not exceed 1.5 mm
(amplitude of tilt and square wave together). After com
pleting this check, repeat step 5. Then go on to step 11.
11. Characteristic Unbalance
Disconnect the signal from the Z Unit inputs, and set
the Mode switch to the TE T position.
Time/cm switch to .5 millisec/cm.
trace, move the COMPARI ON VOLTAGE Polarity switch
to the +, 0, and — positions several times. The overall
baseline shift should not exceed 1 cm. Leave the COM
PARI ON VOLTAGE Polarity switch in the 0 position at
the conclusion of this step.
12. nput-Capacitance Standardization
Loosen the set screws in both VOLT /CM knobs with a
0.05" alien wrench and remove the knobs. This allows ac
cess to the attenuator adjustments through slots in the front
panel.
Each position of the VOLT /CM control has a
HUNT and a
ERIE
adjustment. The only exception is the
.05 position which has only an input-capacitance trimmer
accessible through the
ERIE
Connect the 52-ohm cable to the Type 105
Generator, or equivalent.
Connect the other end of the
cable to the 5-to- 1 " L" pad. Connect the other end of the
"L" pad through the 24- μμ f capacitance standardizer to
the A input connector on the Z Unit.
to A ONLY and both VOLT /CM switches to .05.
To make this adjustment, set the square-wave generator
to produce about 3.5 cm of deflection at a frequency of
1 kc.
et the oscilloscope front-panel controls for " double
triggering" as follows:
Triggering Mode
Trigger lope
Trigger Level
tability
Time/cm
Rotate the Variable Time/cm control slowly counterclock
wise until you obtain a stable display of overlapping square
waves as shown in Fig. 5-5.
Adjust the input-capacitance trimmer in the A VOLT /CM
attenuator for best flat top on the displayed waveform
(see Fig. 5-5), disregarding any fast-rise leading edge spikes.
et the Mode switch to — B ONLY. Disconnect the signal
from the A input connector and connect it to the B input.
Adjust the input-capacitance trimmer in the B VOLT /CM
attenuator for best flat bottom, disregarding the spikes.
13. Output Voltage Divider Compensation
Increase the square-wave generator frequency to 10 kc.
et the oscilloscope Time/cm switch to 50 μ sec and re
adjust the Variable Time/cm control to obtain a stable
display of several overlapping waveforms. Adjust the out
put voltage divider compensations C7655 and C8655 for
et the oscilloscope
While observing the
side of the slot.
quare-Wave
et the Mode switch
AC Fast
+ Internal
0
Fully clockwise
.5 Millisec