Calibration — Type Z
Preset the Z Unit front-panel
PO ITION
A VOLT /CM
B VOLT /CM
Both AC-DC switches
Mode
VAR. ATTEN.
COMPARI ON VOLT
AGE Polarity
COMPARI ON VOLT
AGE Range
Check to see that the COMPARI ON VOLTAGE Helidial
reads exactly zero when the control is rotated fully counter
clockwise.
Rotate the control fully clockwise.
should read about one half a minor division over 10.000
or a reading of 10.005. If the Helidial does not read ex
actly as described, loosen the set screw with 0.050" alien
wrench and set the Helidial for correct readings. Tighten
set screw.
Remove the left side panel from the oscilloscope. Con
nect the plug-in cable extension between the Z Unit and
the oscilloscope vertical plug-in connector. Connect the os
cilloscope power cord to the nominal voltage source for
which it is wired and turn on the power.
Do
not touch
exist on transistor cases. If a transistor needs to
be replaced, turn off the power.
CAL BRAT ON PROCEDURE
1. Check Voltage Measurements.
Use the VOM to make the following measurements:
(a) Measure the voltage drops across the 105-volt zener
diodes, D7621 and D8621 (see Fig. 5-1). The drops must be
Q8672
D8679
Fig. 5-1. Location of components and test points where some of
the step 1 voltage measurements are made.
5-2
htt ://manoman.sqhill.com
controls as follows:
Midrange
.05
.05
DC
TE T
Fully Clockwise
0
100 V
WARNING
transistors!
Dangerous
potentials
BASE OF
D8621
JUNCT ONS OF
R7645 AND
R3645
P N 4,
V7618
D7621
The dial
NTERCONNECT NG
PLUG
Fig. 5-2. Another view of the Z Unit showing location of remain
ing components and test points described in step 1.
within the range of 95-115 volts and within 3 volts of each
other.
(b) Connect the VOM between pin 7 of V7613 (see Fig.
5-2) and ground. Place the COMPARI ON VOLTAGE Po
larity switch to the + position.
approximately 100 volts. Place the COMPARI ON VOLT
AGE Polarity switch to the — position. The meter should
now be reading approximately — 100 volts.
PARI ON VOLTAGE Polarity switch back to 0.
(c) Check the voltage readings listed in Table 5-1. Those
listed from + 12.2 volts and on are obtained by connecting
the VOM between the given location and ground.
READ
0.3 to 0.6 V
0.3 to 0.6 V
5.7 to 6.7 V
5.7 to 6.7 V
+ 12.2 V
+ 65 to +70 V
+ 75 V
-98 V
-138 V
+210 V
P N 7,
F LAMENT AND R8691
The meter should read
TABLE 5-1
LOCATION
Emitter-to-base
drop
across
Q7672.
Emitter-to-base
drop
across
Q8672.
Drop across RT-6 zener diode
D7675.
Drop across RT-6 zener diode
D8679.
Junction
of V8623
filament
lead (blue and orange tracer)
and R8691.
Terminals 1 and 3, intercon
necting plug.
Terminal
15,
interconnecting
plug-
Pin 4 of V7618.
Base of Q7618.
Junction of R7645 and R8645.
et the COM
FIG. NO.
5-2
5-1
5-2
5-1
5-2
5-2
5-2
5-1
5-1
5-1