Welch Allyn Schiller AT-2plus Manual page 50

6-channel electrocardiograph
Hide thumbs Also See for Schiller AT-2plus:
Table of Contents

Advertisement

Technical Data
ECG Amplifier:
Simultaneous, synchronous registration of
all 9 active electrode signals (= 12 standard
leads)
Sampling frequency:
Digital resolution:
Dynamic range:
Max. electrode potential: ±300 mVDC
Time constant:
Frequency response:
dB)
Input impedance:
>2.5MOhms at 10Hz
Myogram Filter (muscle tremor filter)
25 Hz or 35 Hz, programmable (not active
on averaged waveform). The stored ECGs
can be printed with or without filter.
Line Frequency Filter:
superimposed 50 or 60 Hz sinusoidal
interferences by means of an adaptive digital
filter.
Patient Input:
Fully floating and isolated, defibrillation
protected.
4 0
1000 Hz
5 µV
±9.5 mVAC
3.2 s
0.05 to 150 Hz (-3
Distortion-free suppression of
Safety Standard:
CF according to IEC and complying with
the following
RL 93/42/EEC
EN 60601-1:1990
IEC 601-1
IEC 601-2-25:1993
pr EN 1441:1994
EMC:
CISPR 11: 1985, EN 55011: 1992
IEC 801-2: 1991
IEC 801-3: 1984
IEC 801-4: 1988
IEC 801-5:
Safety Class:
I according to IEC 601-1 (with internal power
supply)
IIa according to RL 93/42/EEC, CE-0124
This device is not designed for outdoor use
(IP 20)