Memory and Data Transmission Option

Safety Notices

WHEN NON-MEDICAL DEVICES ARE CONNECTED TO THE
RS-232 INTERFACE ENSURE THAT BOTH UNITS ARE
SECURELY CONNECTED TO THE SAME EARTH POTENTIAL.
WHEN OPERATING THE UNIT ON BATTERY AND
SIMULTANEOUSLY USING NON-MEDICAL DEVICES, THE RS-
232 INTERFACE MUST BE FULLY ISOLATED.
AN EXTERNAL DEVICE MUST ONLY BE CONNECTED USING
THE ORIGINAL SCHLLER INTERFACE CABLE ASSEMBLY.
The memory option allows approximately 45 recordings (dependent
on size and parameters specified when the recording was taken) to
be stored and transmitted over the RS-232 interface. When no more
recordings can be stored the message `MEMORY FULL` is
displayed. Old recordings must be deleted or transmitted before
further recordings can be stored. A number of memory settings can
be made as follows:
Note: At the time of print it is not possible to read or to delete
individual stored recordings.
2 8

Auto Storage and Auto Erase

Memory Setup
Entry Key Sequence
ALT
0
5
With `auto save on`, all auto mode recordings, will be automatically
stored on completion.
With `auto erase on`, all stored recordings are erased after sending
over the RS-232 interface.
Save Mode
0
Auto save off
1
Auto save on
2
Auto erase off
3
Auto erase on