Permobil X850 Corpus 3G User Manual page 68

Hide thumbs Also See for X850 Corpus 3G:
Table of Contents

Advertisement

Seat rotation out and seat lift pinch protection active Error code: LED 3
flashes. If pinch protection is active, the seat lift cannot be lowered.
If the seat rotation is out, the wheelchair cannot be driven.
Remedy: Check if the seat rotation is out or contact your service technician.
Warm motors or hot motors. Error code: LEDs 4–6 flash, depending on the
current maximum speed setting.
Warm: six short beeps at intervals of two minutes when the joystick is
activated.
Hot: Continuous beeps for ten seconds.
Remedy: Turn off the wheelchair and wait for the motors to cool down.
Warm electronics. Error code: LEDs 4–6 flash, depending on the current
maximum speed setting. Six short beeps at intervals of two minutes.
The wheelchair's driving power is reduced. If you drive on a hill, the reduced
driving power can cause the wheelchair to stop or roll slowly downhill. The
ability to brake remains the same and the wheelchair stops when you let go of
the joystick.
Remedy: Turn off the wheelchair.
68 - Troubleshooting

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents