Fluke FIBER ONESHOT PRO User Manual page 31

Singlemode fiber troubleshooter
Hide thumbs Also See for FIBER ONESHOT PRO:
Table of Contents

Advertisement

Auto turn off
Memory for settings
and test results
Factory calibration
interval
Reflectance
threshold
Loss threshold
Distance to the first
reflective open
Shortest distance to
an open APC
connection
Maximum distance
Distance accuracy
(0 m to 23,000 m or 
0 ft to 7,549 ft)
Testing speed
Connector
Fiber types tested
Bulkhead quality
Live fiber detection
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
Automatically turns off after 5 minutes if
no keys are pressed. Backlight turns off
first.
Non-volatile memory. Saves the results
from up to 99 tests.
None
User-selectable from -24 dB to -62 dB
User-selectable from 0.5 dB to 6.1 dB
2 m
15 m
23 km (14.3 miles or 75,459 ft)
±1 m for reflective incidents ±1 % of length
±3 m for non-reflective incidents ±1 % of
length
< 10 seconds typical
Removable/cleanable SC adapter, UPC
polish
9/125 μm singlemode
If no fiber is attached or if the connector is
dirty, the troubleshooter displays 0 m or 0
ft.
Detects optical signals at 1310, 1490, and
1550, and 1625 nm and shows ACTIVE LINE
if a signal is there. Looks for a signal every 3
seconds after the first detection. +7 dBm
maximum input power.
TestEquipmentDepot.com
Specifications
27

Advertisement

Table of Contents
loading

Table of Contents