Page 1
LUDLUM MODEL 2242 SURVEY METER November 2010 Serial Number 219764 and Succeeding Serial Numbers...
Page 2
LUDLUM MODEL 2242 SURVEY METER November 2010 Serial Number 219764 and Succeeding Serial Numbers LUDLUM MEASUREMENTS, INC. 501 OAK STREET, P.O. BOX 810 SWEETWATER, TEXAS 79556 325-235-5494, FAX: 325-235-4672...
Page 3
RETURN OF GOODS TO MANUFACTURER If equipment needs to be returned to Ludlum Measurements, Inc. for repair or calibration, please send to the address below. All shipments should include documentation containing return shipping address, customer name, telephone number, description of service requested, and all other necessary information.
Recalibration Batteries Setup and Calibration Entering or Changing Switch Board Parameters The Function Switch Function Switch Position Descriptions and Variables Calibration General Detector Setup Information Exposure Rate Calibration Detector Overload (OVL) Calibration Loading Default Parameters November 2010 Ludlum Measurements, Inc.
Page 5
MODEL 2242 Technical Manual Troubleshooting Technical Principle of Operation Recycling Parts List Model 2242 Survey Meter 10-1 Main Board, Drawing 408 × 223 10-1 Calibration Board, Drawing 408 × 229 10-3 Display Board, Drawing 408 × 79 10-4 Switch Board, Drawing 408 × 46 10-4 Wiring Diagram, Drawing 408 ×...
Technical Manual Section 1 Section Introduction he Model 2242 is a portable microprocessor-based digital Ratemeter designed for detection and measurement of ionizing radiation. The data is presented on a 4-digit Liquid Crystal Display (LCD). A 3- position switch labeled “...
Page 7
MODEL 2242 Technical Manual Section 1 The instrument operates on two ˝D˝ cell flashlight batteries. The unit body is made of cast-and-drawn aluminum and is covered in beige powder- coating to aid in surface decontamination. Page 1-2 November 2010 Ludlum Measurements, Inc.
(batteries, cables, etc.) and ensure that all of the items listed on the packing list are in the carton. Check individual item serial numbers and ensure calibration certificates match. The Model 2242 serial number is located on the front panel below the battery compartment.
(with the exception of a single discrimination parameter). The Model 2242 incorporates a detachable switch board used to program detector and operating parameters into non-volatile memory (retains the data even after the power is removed). The switch board may be removed so that the operating parameters cannot be altered.
Page 10
MODEL 2242 Technical Manual Section 2 An LCD provides the readout for the ratemeter data with the programmed units and multipliers, as well as ALERT ALARM annunciators, OVERLOAD OVERFLOW "low battery" icon, and scaler counting mode indication ( – not used). The four 0.5" digits are used for COUNTING the ratemeter data.
MODEL 2242 Technical Manual Section 3 Section Specifications : Two standard "D" size batteries; current drain approx. 35 mA POWER (backlight off); minimum battery voltage 2.2 ± 0.1 Vdc : Instrument calibration change less than 3% to BATTERY DEPENDENCE battery endpoint...
Page 12
MODEL 2242 Technical Manual Section 3 : Separate visual and audible adjustable alarm points for ALERT/ALARM both the mR/hr and R/hr ranges : Adjustable from 2-100 mV; negative pulse INPUT SENSITIVITY response. Normally set at approximately 90 mV for the internal detectors : Externally adjustable from 400 to 2500 volts;...
RESET and/or alert condition. Depressing the scaler count switch located in the end of the Model 2242 handle resets the scaler alarm only. Remove the front panel CAL cover to access the following controls: : A multi-turn potentiometer (approximately 20...
To access the internal circuit boards, unlatch the latches at each end of the Model 2242. Carefully separate the top chassis from the bottom cover (referred to as a ˝can˝). The can has the audio speaker (unimorph) with a 2- conductor cable attached to the main board.
16 setup parameters. (Refer to schematics and component layout drawing near the end of the manual.) All setup parameters are stored in the non-volatile EEPROM, which will retain data even after the Model 2242 batteries are removed. After parameters are entered the switch board can be...
Page 16
MODEL 2242 Technical Manual Section 4 - Indicates that the detector is being exposed to radiation OVERLOAD intensities greater than the detector maximum operating limit. The overload alarm point is set by adjusting the control located underneath the cover. - Indicates that the batteries have decreased to the "low battery"...
Pollution Degree 1 (as defined by IEC 664) Cleaning Instructions and Precautions The Model 2242 Scaler/Ratemeter may be cleaned externally with a damp cloth, using only water as the wetting agent. Do not immerse the instrument in any liquid. Observe the following precautions when cleaning: 1.
The Model 2242 instrument may be cleaned with a damp cloth (using only water as the wetting agent). Do not immerse instrument in any liquid. Observe the following precautions when cleaning: 1.
Check the appropriate regulations to determine required recalibration intervals. Ludlum Measurements offers a full service repair and calibration department. We not only repair and calibrate our own instruments but most other manufacturer’s instruments. Calibration procedures are available upon request for customers who choose to calibrate their own instruments.
A 16-position rotary switch labeled ˝0-9˝ and ˝ ˝. This switch selects a parameter setup mode for the Model 2242. If the board is not installed, the normal operation mode (counting mode) is selected. If the switch board is installed, the selector switch must be set to the ˝0˝ position for normal operation.
Page 21
Section 6 o Once the desired data is entered, depress . The LCD stops ENTER flashing and the new parameter data is displayed. Note: switch allows the Model 2242 to have 2 HIGH sets of operating parameters. UNCTION WITCH OSITION...
Page 22
) for POSITION 6 RESPONSE TIME the current detector setup. If the response is set to ˝0˝ the Model 2242 automatically calculates (for variable mode), the time constant based on the incoming cps. If a variable of 1-199 is entered for , the response time becomes fixed.
550 Vdc and 460 Vdc, and instrument sensitivity ( ) of 90 mV. DISC XPOSURE ALIBRATION To calibrate the Model 2242 to exposure rate after setting the potentiometers, start with the following values for (Dead Time, DISC Switch Position 1) and (Calibration Constant, Switch Position 2).
MODEL 2242 Technical Manual Section 6 For exposure rate calibrations, use a calibrated Cs source, and set the Model 2242 to ˝mR/hr˝. Place the detector at the following points: mR/hr µR/hr 8* adjust CC 8* adjust CC 200* adjust DT...
MODEL 2242 Technical Manual Section 6 Place the detector in an increasing radiation field in which the readout no longer increases. Adjust the control until the OVERLOAD OVERLOAD alarm appears. Position the detector between the upper operating limit and the overload set point and ensure the overload alarm is defeated.
MODEL 2242 Technical Manual Section 7 Section Troubleshooting ccasionally, you may encounter problems with your LMI instrument or detector that may be repaired or resolved in the field, saving turnaround time and expense in returning the instrument to us for repair. Toward that end, LMI electronics technicians offer the following tips for troubleshooting the most common problems.
Page 27
MODEL 2242 Technical Manual Section 7 SYMPTOM POSSIBLE SOLUTION No power, or low Check battery contacts. Clean them battery icon displays with rough sandpaper or use an (continued) engraver to clean the tips. Check for loose or broken wires, especially between the main board and the calibration board.
Page 28
MODEL 2242 Technical Manual Section 7 SYMPTOM POSSIBLE SOLUTION No Audio Ensure the switch is in AUD ON position. Remove the instrument housing and check the connection between the circuit board and the speaker. Plug in the 2-pin connector if necessary.
MODEL 2242 Technical Manual Section 8 Section Technical Principle of Operation Refer to the Main Board Schematic, Drawing 408 × 223 for the following: Detector Input/ Amplifier Negative-going detector pulses are coupled from the detector through C021 to Amplifier U021. R024 and CR021 protect the input of U021 from inadvertent shorts.
Page 30
MODEL 2242 Technical Manual Section 8 U231 provides a low signal when the battery voltage decreases to +2.2 ±0.1Vdc. U121 provides the +2.5Vdc reference for the control DISC references. High Voltage Supply High Voltage is developed by blocking oscillator Q241, T141, and C244 and rectified by voltage multiplier CR041-CR043, C041-C043, and C141.
Page 31
MODEL 2242 Technical Manual Section 8 Audio Click/event, divide-by, and alarm audio pulse frequency is generated by the µP and coupled to Q101. Q101 then inverts the pulses and drives the bottom of T101. Bias voltage is provided by the volume control (R002) to the top of T101.
Page 32
MODEL 2242 Technical Manual Section 8 Backlight Drive Depressing the switch instructs the µP to set the line, pin LIGHT BACKLIGHT 31 on µP, "low" for the predetermined backlight time. (Refer to main board schematic for details.) A "low" condition on pin 31 causes Q212 to conduct sending +3V to P8-3 on display board.
To this end, Ludlum Measurements, Inc. strives to supply the consumer of its goods with information regarding reuse and recycling of the many different types of materials used in its products. With many different agencies, public and private, involved in this pursuit it becomes evident that a myriad of methods can be used in the process of recycling.
Technical Manual Section 10 Section Parts List Reference Description Part Number Model 2242 UNIT Completely Assembled Survey Meter Model 2242 Survey Meter 48-3437 BOARD Completely Assembled Main Board, Main Board 5408-223 Drawing 480 × 223 Y221 MICRO X-TAL-6.144 MHZ 01-5262...
Need help?
Do you have a question about the 2242 and is the answer not in the manual?
Questions and answers