Wellness AquaGlo Plus N1581 Owner's Manual page 18

Abrasion-free liquid facial rejuvenation
Table of Contents

Advertisement

3. Ensure you can see solution in the tubes (solution bottle to machine).
4. Turn machine off and unplug from wall power.
5. Pull the hand piece apart (see picture above) and check for an obstruction or
breakage at the tip and tubing.
6. Open up the side of the machine with the flow adjustment valve.
a. Ensure there is fluid on both sides of the valve control. If there is fluid
on one side of the valve and none on the other, the valve is clogged
and must be removed and cleaned with warm distilled water or
replaced.
b. Ensure the tubing going into and out of the flow control valve is
connected tightly (push tubing into blue connector).
Make sure hoses are not kinked or
obstructed in any way.
16