5. If your network analyzer is an N8362A/B or N8364A/B PNA, you will need to perform the
Phase-Lock IF Gain Adjustment on the analyzer after it has been connected to the test set
as a system. This routine adjusts the R Channel receivers ALC gain to ensure phase lock
over the entire frequency range. Refer to Phase-Lock IF Gain Adjustment in the
network analyzer's online help system for details. Use the following steps to perform this
adjustment:
a. On the network analyzer, from the System menu, click Service, then Adjustments,
then IF Gain Adjustment.
If you are unable to find these selections on your N8362A/N8364A PNA, your analyzer
firmware is a revision prior to 3.0. Follow the instructions in
page 212
to complete this adjustment.
b. Select the test set options installed.
c. No connections to the test ports are required.
d. Click Begin Adj. The adjustment takes about a minute to complete.
The advanced screen is for factory personnel only.
This adjustment must be performed before using the network analyzer each
NOTE
time the system is assembled or disassembled.
6. If it hasn't already been done, install and configure the software. See
Start the Balanced Measurement Software" on page
Chapter 3
Using the Multiport Test System
How to Set Up the Multiport Test System Hardware
"IF Gain Adjustment" on
39.
"How to Install and
19