The Pacific Sun Research and Development team is continuously testing, developing, and releasing new products with new features. This document serves to cover the basics of the Kore 7th dosing station and the kHLab module. Details will be made available about additional new products and features as they are released.
Kore 7th Manual / kHLab Edition 1.3. The Kore 7th / kHLab Device Features The kHLab has an ability to measure alkalinity with lab grade accuracy. It can provide KH measurement values with 0.1 dKH accuracy. It’s a minimal confirmed resolution during continuous device function.
NOT exceed about 50-60cm (20-24”) and connection tubing length between the kHLab module and Kore 7th doser should NOT exceed 100cm (59”). The kHLab device can be mounted directly on a wall with the supplied wall mount, or it can operate sitting on a flat surface.
Page 6
Important! 1. Make sure that a water mixing magnetic stirrer bar/pellet is put inside kHLab module test cylindrical chamber. 2. Make sure to connect silicone tubing properly to individual Kore 7th dosing pumps: – Pump nr 1 - supplies (aquarium water) to kHLab water sample test chamber.
Page 7
Kore 7th Manual / kHLab Edition Pumps with connectors facing UP Note: Pumps heads can function with all connectors facing UP or DOWN. Pumps with connectors facing DOWN Page 7...
(small white bar). That bar/pellet is necessary for proper water sample mixing during perform tests. Also, connect the kHLab module to the Kore 7th dosing station by using the dedicated 8-pin plug cable. Page 8...
Kore 7th Manual / kHLab Edition 2.4. Tubing and cables connection Kore 7th and kHLab connection setup examples Page 9...
Page 10
Kore 7th Manual / kHLab Edition Water Sample Connector 1 OUT Test Water Sample outlet to the kHLab module Water - using silicone tubing, connect to the port at the top of the kHLab module. Connector 2 IN Test Water Sample...
Kore 7th Manual / kHLab Edition 2.5. Calibration With Precise Digital Scale and Stand/Holder The Precise Digital Scale and Stand/Holder are required for the Reagent Solution Preparation and Kore 7th calibration process. Calibration station setup Digital Scale Page 11...
Kore 7th Manual / kHLab Edition 3. pH Probe Setup, Maintenance and Preparation Use only dedicated Pacific Sun pH probes for these calibrations. If you use probes from another manufacturer, Pacific Sun cannot guarantee the accuracy that measured results will match actual conditions. The typical lifespan for a pH probe in kHLab is about 16-18 months, but this time can be shortened in the case of long-term exposure to seawater.
Kore 7th Manual / kHLab Edition 3.1. Initial pH Probe Preparation - First Use After Purchase Important! Never expose pH probe to the air for more than 3-4 minutes. The pH probe should be always in the water or in it's transport protection container.
Kore 7th Manual / kHLab Edition 4. Reagent Solution Preparation This device preparation step is very important. If the reagent dilution ratio is not correct, the tested results may differ from the actual alkalinity levels. To obtain accurate results, we strongly suggest using the digital scale for reagent dilution.
Kore 7th Manual / kHLab Edition 5. Kore 7th Wi-Fi Cloud Pacific Sun network setup 5.1. Connecting Kore 7th to your home Wi-Fi network 1. Connect power to Kore 7th. Find your doser on you Wi-Fi network list devices. Click Connect directly to the Kore7th Wi-Fi Network.
Page 16
Next Click: Connect! After this Kore7th will be available under your Home WiFi network usually with IP address format: 192.168.#.### 7. Restart Kore 7th. Power OFF and ON the Kore 7th. Disconnect power and Connect back. On the LCD screen, see and take a note of the IP Address in a similar format.
Kore 7th Manual / kHLab Edition 5.2. Adding Kore 7th to Pacific Sun Cloud 1. Connect to your home Wi-Fi internet network, open Internet Browser, and go to the following address: https://cloud.pacific-sun.eu Click Create an account 2. Create Cloud Pacific Sun account.
Page 18
Kore 7th Manual / kHLab Edition 4. Add a Tank: Type in your preferred Name Click Add tank 5. Under your tank Name: Click Add Device 6. Select Kore 7th from the list of devices: Click Add Device Page 18...
Page 19
Kore 7th Manual / kHLab Edition 7. Under your Client ID: Click the icon GREEN 8. Connect Kore 7th to your Wi-Fi: Step Click Next 9. Configure device and Ready: Step Use your Kore 7th IP address Click Configure! Note: You can see the Kore7th...
Page 20
Kore 7th Manual / kHLab Edition 10. Kore 7th added to the PS Cloud, and ready for the calibration process, configuration and operation. Page 20...
Kore 7th Manual / kHLab Edition 6. kHLab Pumps Calibration and Accuracy Check The kHLab calibration process is based on the Pacific Sun Cloud solution and requires to follow a few automatic steps. Water Sample Step 1 Fill Tube #1...
Kore 7th Manual / kHLab Edition 6.1. Calibration With Precise Digital Scale and Stand/Holder Under your Kore 7th kHLab device go to System/Config Click Start KH-Lab Wizard Page 22...
Kore 7th Manual / kHLab Edition 6.2. Channel #1 Calibration Process - Water Sample Channel #1 is dedicated to Samples. To calibrate pump#1 in this channel follow Water the steps below. Refer to the following pictures. Step 1 - Fill Tube #1...
Kore 7th Manual / kHLab Edition 6.3. Channel #3 Calibration Process - Waste Water Channel #3 is dedicated to Water. To calibrate pump#3 in this channel follow Waste the steps below. Refer to the following pictures. Step 3 - Fill Tube #3...
Kore 7th Manual / kHLab Edition 6.4. Channel #2 Calibration Process - Reagent Channel #2 is dedicated to the solution. To calibrate pump#2 in this channel Reagent follow the steps below. Refer to the following pictures. Step 5 - Fill Tube #2...
7. Alkalinity Test Intervals The Kore 7th dosing station can perform Alkalinity tests in different time intervals: one, two, four, eight, twelve, and twenty-four times per day (x1, x2,x4, x8, x12, x24).
(keeping the defined KH value in case of the alkalinity dropping below the pre-set “Minimum KH” limit value). The following tables show recipes with supplements from other manufacturers which will work correctly with the Kore 7th/kHLab device. Recipe based on NaHCO₃/Na₂CO₃ RO/DI water NaHCO₃...
Download 1. Connect your Laptop/PC to the Kore 7th dosing station via USB port. 2. Windows Operating System should be able to discover and install the new Kore 7th USB device and show the communication COM port number. NOTE: If for the first time, Windows will not be able...
Page 31
Option 2: Verify the USB Serial Port COM number from “Devices and Printers”. 4. Right Click on the Kore 7th Doser Device and select “Properties”. 5. The USB Serial Port COM number will be visible under Kore 7th Doser Properties.
Page 32
For the Mac OS version download from: http://www.pacific-sun.eu/pliki/ PSunFlash%202.zip 7. In the “Settings” tab, choose the port COM number the Kore 7th is installed on the computer and click Apply Note: For the Mac OS version, select the device you want to update from the list.
Page 33
Kore 7th Manual / kHLab Edition 9. Select the previously downloaded Firmware file and click Open 10. Click Update 11. Firmware upgrade will be in progress with massage “Please wait……..” Page 33...
Page 34
Kore 7th Manual / kHLab Edition 12. Now, press the RESET push button on the Kore 7th dosing station back panel. 13. The Firmware writing process will start and the number in bytes will be increasing. 14. The Firmware upgrade status will be indicated with the green progress bar.
Page 35
Kore 7th Manual / kHLab Edition 15. Final Firmware Upload steps: 1. At the end of the Firmware writing process program will show “Firmware updated”. 2. Disconnect the USB cable from your laptop/PC and Kore 7th dosing station. 3. Disconnect Kore 7th from...
Kore 7th Manual / kHLab Edition 13. Troubleshooting Problems Solutions Make sure that pump #1 and pump #2 are correctly Device is not accurate calibrated, and that there is no air (long air gaps) inside the silicone tubing. Small air bubbles inside the tubing don’t impact measurement accuracy.
Page 37
Kore 7th Manual / kHLab Edition If the water mixing magnetic stirrer bar/pellet is not spinning Spinning problem with then in the Cloud Application, under the "KH Lab Settings", the magnetic stirrer bar/ increase the “Stirrer speed” from the default 100 to e.g. 150 pellet or higher, and test it again.
Page 38