13. Troubleshooting

Problems
Device is not accurate
Problem with
communication between
the Kore 7th and PS
Cloud service
Solutions
Make sure that pump #1 and pump #2 are correctly
calibrated, and that there is no air (long air gaps) inside the
silicone tubing. Small air bubbles inside the tubing don't
impact measurement accuracy.
Make sure that the magnetic stirring bar/pellet for mixing
tests fluid is inside kHLab cylindrical test chamber per the
instructions.
Make sure that silicone tubing supplying reagent and test
water are submerged below reagent solution and aquarium
water levels.
Make sure that the
chamber cover is submerged below the test water level
when the alkalinity test is in progress.
When replacing dosing pumps PharMed tubing, the device
may initially perform slightly different from before tubing
replacement but should go back to the same accuracy
after a few days without any problem. During that time
next pump calibration is not recommended. New pump
calibration is required when the new tubing has a different
diameter than the previous, original one or has a different
nominal flow, i.e. tubing from a manufacturer other than
Pacific Sun.
1. Reboot your home Wi-Fi router (power OFF/ON)
and after that
reboot your Kore 7th (power OFF/ON).
Wait a few minutes
interface functionality again.
2. If you have multiple Wi-Fi routers,
them has a unique
3. If you have multiple Wi-Fi routers, make sure you are
connecting to the closest one
Page 36
tip under the kHLab module
green
and check Cloud application
name.
with the strongest signal.
make sure each of