CALIBRATION AND VERIFICATION TEST PROCEDURE
1. INTRODUCTION
This Specification defines the suggested method of calibration and
performance verification for the 3M 701 Megohmmeter. It also defines the
test equipment to be used.
2. TEST EQUIPMENT and DOCUMENTATION
A. A Keithley Model 619 Electrometer or equivalent (input impedance ≥ 1
Ω).
9
x 10
B. A decade box with a range of 1OOKΩ to 10GΩ ±1 % or the following
individual resistors:
100KΩ ±1%
1MΩ ±1%
100MΩ ±1%
100MΩ ±1%
1GΩ ±1%
10GΩ ±1%
C. Figure 1 - Calibration adjustments
Figure 2 - Main PC Board Assembly issue 1
Figure 3 - Main PC Board Assembly - issue 2
Figure 4 - Main PC Board Assembly - issue 3
3. VERIFICATION PROCEDURE
A. Place selector switch in the OFF position and adjust the mechanical
zero to the left end of the scale.
B. Place selector switch in the BATTERY TEST mode. Verify that pointer
is in the green (100V) area on the BATTERY TEST scale.
C. Place the 100KΩ resistor across the input. Switch the unit to
CONTINUITY TEST. Verify that the unit reads within limits (± 2° of
100KΩ mark on the CONTINUITY TEST scale. Short the input; verify
that the unit has full-scale deflection.
D. Remove the short; place the decade box across the input. Verify that
the unit reads within limits for 10GΩ and 1 MΩ in both the 10V and
100V SURFACE TEST modes. (Tolerances for the 1 MΩ / 10 GΩ
FOR 3M 701 MEGOHMMETER

Advertisement

loading

Summary of Contents for 3M 701

  • Page 1 FOR 3M 701 MEGOHMMETER 1. INTRODUCTION This Specification defines the suggested method of calibration and performance verification for the 3M 701 Megohmmeter. It also defines the test equipment to be used. 2. TEST EQUIPMENT and DOCUMENTATION A. A Keithley Model 619 Electrometer or equivalent (input impedance ≥ 1 Ω).
  • Page 2: Calibration Procedures

    readings are shown on the top scale of the meter as small dots on either side of the respective resistance mark on the dial.) E. Remove the decade box. Using the Keithley 619, verify that the output is between 93V and 107V in the 100V SURFACE TEST mode. F.
  • Page 3 701 MEGOHMMETER VERIFICATION / CALIBRATION Figure 1 Calibration Adjustments (R36 is located on opposite edge of printed circuit board)
  • Page 4 Figure 2. Main PC Board Assembly - Model T00456 3M 701 (Issue 1)
  • Page 5 Figure 3. Main PC Board Assembly - Model T00456 3M 701 (Issue 2)
  • Page 6 Figure 4. Main PC Board Assembly - Model T00456 3M 701 (Issue 3)
  • Page 7 The accuracy of the 701 is based on two tolerances (+/- 5% electrical & +/-2 deg. of arc, meter’s needle movement). The sum of these two tolerances are identified by two tiny dots, one on each side of the major intervals (100 kilohms, 1megohm, 10 megohms, 100 megohms, 1gigaohm, and 10 gigaohms) marks located on the top scale.