Faults, Diagnosis And Troubleshooting - AQUALUX Dinotec Premium Installation, Use And Maintenance Manual

Salt chlorinator
Table of Contents

Advertisement

Available languages

Available languages

15. FAULTS, DIAGNOSIS AND TROUBLESHOOTING

• The set of operations described in chapter 15 presents some risks. These operations must be carried out by
persons with the required training and authorisation.
Comply with all the safety instructions presented in chapter 2.
→ Any alarm triggered : ● will immediately be displayed on the screen.
→ Any alarm triggered :
→ Any alarm triggered :
→ The "
pH Calibration alarms
→ The alarms "
" to "
A1
Message displayed
/ Fault detected
Alarms
pH Calibration
Alarms
attempts to correct the
AL – pH Regulation
Alarms
A1 – Low Salt
Security
A2 – Low Temp
Alarms
A3 – No Flow
Alarms
A4 – Cell Current
● can be manually cleared by pressing button N.
● can be disabled (except "
by disabling the function of the sensor related to the fault detected
(see the "
Parameters - Sensors
" and "
AL – pH Regulation alarms
" automatically and immediately stop chlorine production.
A4
Cause
pH probe incorrectly
calibrated
Series of unsuccessful
pH
Salt levels
below 2.5 g/L
Insufficient amount of
water in the filtration
heat contactor
Water temperature
below 15 °C
Check that :
Insufficient water flow
through the filtration
Check that :
heat contactor
Check that :
Check that :
Check that :
Cell problem
I
A4 alarms - Cell Current
" menu in chapter 10).
" automatically and immediately stop the pH regulation.
Carry out a calibration of the pH probe (see chapter 12.1).
- Ensure the pH corrector drum is not empty.
- Carry out a manual injection (see the "
menu in chapter 10).
- Check the condition of the ballasted filter and injection connector.
- Check the settings in the "
"
pH Regulation – Corrector
10).
- Carry out a calibration of the pH probe (see chapter 12.1).
- Check the salt levels in the pool using a testing kit.
- Top up the salt, if necessary, so as to achieve a salt content of 5 kg/m
- Check that the pipe at the level of the accessories holder is completely
filled with water.
- If necessary, top up the water in the pool.
See Chapter 13.
Check that : - the flow sensor is properly connected to the electronics unit.
- the flow function is properly activated
(see the "
- the valves on the filtration heat contactor are open.
- the filtration pump is working correctly.
- the filtration heat contactor is not blocked.
- there is enough water in the pool.
- Check that scale deposits have not built up on the cell (to remove scale,
see chapter 14.2).
- Check and adjust, if necessary, the current inversion frequency powering
the cell (see the "
Electrolysis - Inversion
- Check that the electrical connections to the terminals of the cell are
sufficiently tight and not oxidised.
- Check that the cell's power cable is in a good condition.
- Check that the cell's power cable connector is correctly connected to the
electronics unit.
- As a last resort, replace the cell.
"),
Checks and solutions
pH Regulation – Manual Injection
pH Regulation - Setpoint
" and "
Parameters – Volume
" menu in chapter 10).
Parameters - Sensors
" menu in chapter 10).
"
" menus
" (see chapter
3
.
31

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents