15. FAULTS, DIAGNOSIS AND TROUBLESHOOTING
• The set of operations described in chapter 15 presents some risks. These operations must be carried out by
persons with the required training and authorisation.
•
Comply with all the safety instructions presented in chapter 2.
→ Any alarm triggered : ● will immediately be displayed on the screen.
→ Any alarm triggered :
→ Any alarm triggered :
→ The "
pH Calibration alarms
→ The alarms "
" to "
A1
Message displayed
/ Fault detected
Alarms
pH Calibration
Alarms
attempts to correct the
AL – pH Regulation
Alarms
A1 – Low Salt
Security
A2 – Low Temp
Alarms
A3 – No Flow
Alarms
A4 – Cell Current
● can be manually cleared by pressing button N.
● can be disabled (except "
by disabling the function of the sensor related to the fault detected
(see the "
Parameters - Sensors
" and "
AL – pH Regulation alarms
" automatically and immediately stop chlorine production.
A4
Cause
pH probe incorrectly
calibrated
Series of unsuccessful
pH
Salt levels
below 2.5 g/L
Insufficient amount of
water in the filtration
heat contactor
Water temperature
below 15 °C
Check that :
Insufficient water flow
through the filtration
Check that :
heat contactor
Check that :
Check that :
Check that :
Cell problem
I
A4 alarms - Cell Current
" menu in chapter 10).
" automatically and immediately stop the pH regulation.
Carry out a calibration of the pH probe (see chapter 12.1).
- Ensure the pH corrector drum is not empty.
- Carry out a manual injection (see the "
menu in chapter 10).
- Check the condition of the ballasted filter and injection connector.
- Check the settings in the "
"
pH Regulation – Corrector
10).
- Carry out a calibration of the pH probe (see chapter 12.1).
- Check the salt levels in the pool using a testing kit.
- Top up the salt, if necessary, so as to achieve a salt content of 5 kg/m
- Check that the pipe at the level of the accessories holder is completely
filled with water.
- If necessary, top up the water in the pool.
See Chapter 13.
Check that : - the flow sensor is properly connected to the electronics unit.
- the flow function is properly activated
(see the "
- the valves on the filtration heat contactor are open.
- the filtration pump is working correctly.
- the filtration heat contactor is not blocked.
- there is enough water in the pool.
- Check that scale deposits have not built up on the cell (to remove scale,
see chapter 14.2).
- Check and adjust, if necessary, the current inversion frequency powering
the cell (see the "
Electrolysis - Inversion
- Check that the electrical connections to the terminals of the cell are
sufficiently tight and not oxidised.
- Check that the cell's power cable is in a good condition.
- Check that the cell's power cable connector is correctly connected to the
electronics unit.
- As a last resort, replace the cell.
"),
Checks and solutions
pH Regulation – Manual Injection
pH Regulation - Setpoint
" and "
Parameters – Volume
" menu in chapter 10).
Parameters - Sensors
" menu in chapter 10).
"
" menus
" (see chapter
3
.
31